基于 True FISP 读出的 3T 三维单次激发 CEST 成像
3D Single-shot CEST imaging at 3T Based on True FISP Readout
AI总结:
本文提出基于 True FISP/bSSFP 读出的 3T 三维单次激发 CEST 序列,经仿真、模体、健康志愿者及脑肿瘤患者验证,其 SNR、对比度和结构显示均优于 FLASH。
摘要:
为了同时拟合多个池的效应,频谱选择性三维 CEST 成像通常需要采用单次激发读出以节省时间。然而,迄今为止,FLASH 和 EPI 一直是用于该目的的主要脉冲序列。它们存在 SNR 较低或与 B0 场不均匀性相关的图像畸变问题。在本研究中,我们开发了一种采用 true fast imaging with steady-state precession(True FISP,真实稳态进动快速成像)读出(也称为 bSSFP)的三维单次激发 CEST 序列,并通过仿真优化了扫描参数。该 CEST 序列的性能在一台 3T 人体扫描仪上利用蛋清模体、十名健康志愿者和一名脑肿瘤患者进行了验证。随后,在保持预饱和模式、重复时间、回波时间和扫描时间完全相同的情况下,将所提出的采用 True FISP 的 CEST 序列与常用的基于 FLASH 的 CEST 序列进行了比较,重点关注 SNR 和图像对比度。在仿真实验中,True FISP 获得的最大 CEST 信号显著高于 FLASH 序列获得的信号。在蛋清模体中,True FISP 获得的 amide proton transfer(APT,酰胺质子转移)和 nuclear Overhauser enhancement(NOE,核 Overhauser 增强)效应图像的 SNR 分别比 FLASH 序列获得的高 68.3% 和 57.0%。在健康志愿者中,True FISP 序列在 3.5 ppm 处采集的饱和图像与 FLASH 序列采集的图像相比,平均时间 SNR 提高约 84%。与 FLASH 序列相比,True FISP 序列获得的 CEST 图像能够显示正常个体和脑肿瘤患者更细致的脑组织解剖结构。因此,由于该序列固有的高 SNR,True FISP 有潜力用于临床应用中 CEST 对比度的快速、高质量三维图像读出。
英文摘要:
To simultaneously fit multiple-pool effects, spectrally selective 3D CEST imaging typically requires single-shot readouts to save time. However, to date, FLASH and EPI have been the primary pulse sequences used for this purpose. They suffer from low SNR or image distortion related to B0 field inhomogeneity. In this work, we developed a 3D single-shot CEST sequence using true fast imaging with steady-state precession (True FISP) readout, also known as bSSFP, and optimized the scanning parameters through simulations. The performance of the CEST sequence was validated using an egg white phantom, ten healthy volunteers, and a patient with a brain tumor on a 3T human scanner. Subsequently, the proposed CEST sequence using True FISP was compared with the commonly used FLASH-based CEST sequence, focusing on SNR and image contrast, while maintaining identical pre-saturation modes, repetition time, echo time and scan time. In the simulation experiments, the maximum CEST signal obtained from the True FISP was significantly greater than that obtained from the FLASH sequence. In the egg white phantom, the SNRs of amide proton transfer (APT) and nuclear Overhauser enhancement (NOE) effect images obtained from the True FISP were 68.3% and 57.0% higher than those obtained from the FLASH sequence, respectively. In healthy volunteers, saturated images collected with the True FISP sequence at 3.5 ppm showed an approximate 84% increase in mean temporal SNR compared to those collected with the FLASH sequence. Compared to the FLASH sequence, the CEST images obtained from the True FISP sequence could display more detailed brain tissue structures of both normal individuals and the patient with a brain tumor. Therefore, due to the high SNR inherent in the sequence, True FISP has the potential to be used for fast and high-quality 3D image readout of CEST contrasts in clinical applications.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。