一种用于3T下快速、可重复且高保真全脑多池CEST成像的框架
A framework for rapid, reproducible, and high-fidelity whole-brain multi-pool CEST imaging at 3T
AI总结:
本研究开发并验证了一种3T下快速、可重复的全脑多池CEST成像框架,通过优化True FISP序列、神经网络B1校正和AREX指标校正,在约9分钟内实现了高质量、高可重复性的定量代谢成像。
摘要:
目的:开发并验证一种用于3T下快速、准确且可重复的全脑多池化学交换饱和转移(CEST)成像框架,以解决采集时间长和混杂因素的挑战。方法:对单次激发3D真实稳态进动快速成像(True FISP)序列进行优化,用于全脑多池CEST。使用双回波改进的四角度法进行快速B0、B1和T1标测。开发了一种前馈神经网络用于快速B1校正,并以常规多功率方法为基准进行训练。使用表观交换依赖弛豫(AREX)指标来校正T1和磁化转移(MT)效应。该框架在体模和健康人类受试者(N=8)中进行了验证,包括重测可重复性评估。结果:True FISP序列在临床可行的扫描时间(约9分钟)内产生了高质量、伪影和畸变最小的全脑图像。体模研究证实了B1校正的有效性(MT_MTRLD的变异系数[CV]从22.49%降至4.61%)以及基于AREX的混杂因素校正有效性(APT_AREX的CV从33.6%降至6.9%)。神经网络B1校正在体内与常规多功率方法显示出极好的一致性(ICC > 0.97)。在96个脑区中表现出高重测可重复性,超过95%区域的APT_AREX平均CV低于10%。结论:成功开发并验证了一种用于全脑定量多池CEST成像的快速且稳健的框架。通过将高效的采集序列与简化的校正流程相结合,该方法克服了临床转化的关键障碍,能够为广泛的脑部病理提供可靠的代谢成像。
英文摘要:
Purpose: To develop and validate a framework for rapid, accurate, and reproducible whole-brain, multi-pool chemical exchange saturation transfer (CEST) imaging at 3T, addressing challenges of long acquisition times and confounding factors. Methods: A single-shot 3D true fast imaging with steady-state precession (True FISP) sequence was optimized for whole-brain multi-pool CEST. Rapid B0, B1, and T1 mapping was performed using a dual-echo modified four-angle method. A feed-forward neural network was developed for rapid B1 correction, trained against the conventional multi-power method. The apparent exchange-dependent relaxation (AREX) metric was used to correct for T1 and magnetization transfer (MT) effects. The framework was validated in phantoms and healthy human subjects (N=8), including a test-retest reproducibility assessment. Results: The True FISP sequence yielded high-quality, whole-brain images with minimal artifacts and distortion in a clinically feasible scan time (~9 minutes). Phantom studies confirmed the effectiveness of B1 correction (coefficient of variation [CV] for MT_MTRLD decreased from 22.49% to 4.61%) and AREX-based confounder correction (CV for APT_AREX reduced from 33.6% to 6.9%). The neural network B1 correction showed excellent agreement with the conventional multi-power method in vivo (ICC > 0.97). High test-retest reproducibility was demonstrated across 96 brain regions, with the average CV for APT_AREX under 10% for over 95% of regions. Conclusion: A rapid and robust framework for whole-brain quantitative multi-pool CEST imaging was successfully developed and validated. By integrating an efficient acquisition sequence with a streamlined correction pipeline, this approach overcomes key barriers to clinical translation, enabling reliable metabolic imaging for widespread brain pathologies.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。