通过屏气黑血Single Shot快速自旋回波采集提高肝脏T1rho测量精度:健康志愿者的验证研究

Improved liver T1rho measurement precision with a breathhold black blood single shot fast spin echo acquisition: a validation study in healthy volunteers

Yi-Xiang Wang, Min Deng, GladsG Lo, Queenie Chan, Jing Yuan, Weitian Chen

摘要:

目的:探讨基于新型单次屏气黑血Single Shot快速自旋回波采集的肝脏成像序列的可用性和正常肝实质T1 rho值。材料和方法:共有19名健康受试者(10名男性,9名女性;平均年龄:37.4岁;范围:23-54岁)参加了这项研究。11名受试者在同一阶段进行了两次肝脏扫描,以评估扫描-再扫描可重复性。12名受试者在两个时段进行两次肝脏扫描,间隔7-10天,以评估扫描-再扫描的可重复性。MR检查采用3.0T双发射机扫描仪。MR序列允许在10秒内同时采集4个自旋锁定时间(TSL:0 ms、10 ms、30 ms、50 ms)。利用快速自旋回波固有的黑血效应和双反转恢复来实现血液信号抑制。结果:该技术表现出良好的图像质量和最小的伪影。对于肝实质,Bland-Altman图显示扫描-再扫描重复性平均差为0.025 ms(95%一致性限值:-1.163至1.213 ms),组内相关系数(ICC)为0.977。扫描-再扫描再现性平均差为-0.075 ms(95%一致性限值:-3.280至3.310 ms),ICC为0.820,优于先前亮血多呼吸保持梯度回波采集技术的ICC 0.764。肝脏T1 rho值为39.9 +/- 2.4 ms(范围:36.1 - 44.2 ms),低于使用之前的亮血技术获得的值42.8=/-2.1 ms。结论:本研究验证了基于人肝T1 rho成像的单次屏气黑血Single Shot快速自旋回波采集的应用。低的肝实质T1 rho值和高的扫描重复性可提高该技术的敏感性。

英文摘要:

Purpose: To explore the usability and normal T1rho value of liver parenchyma with a novel single breathhold black blood single shot fast spin echo acquisition based liver imaging sequence. Materials and Methods: In total 19 health subjects (10 males, 9 females; mean age: 37.4 yrs; range: 23-54 yrs) participated in the study. 11 subjects had liver scanned twice in the same session to access scan-rescan repeatability. 12 subjects had liver scanned twice in two sessions with 7-10 days' interval to access scan-rescan reproducibility. MR was performed with a 3.0 T scanner with dual transmitter. The MR sequence allows simultaneous acquisition of 4 spin lock times (TSLs: 0ms, 10 ms, 30 ms, 50ms) in 10 second. Inherent black blood effect of fast spin echo and double inversion recovery were utilized to achieve blood signal suppression. Results: The technique demonstrated good image quality and minimal artifacts. For liver parenchyma, Bland-Altman plot showed the scan-rescan repeatability mean difference was 0.025 ms (95% limits of agreement: -1.163 to 1.213 ms), and intraclass correlation coefficient (ICC) was 0.977. The scan-rescan reproducibility mean difference was -0.075 ms (95% limits of agreement: -3.280 to 3.310 ms), and ICC was 0.820 which is better than the ICC of 0.764 of a previous bright blood multi-breath hold gradient echo acquisition technique. The liver T1rho value was 39.9 +/- 2.4 ms (range: 36.1 - 44.2 ms), which is lower than the value of 42.8=/-2.1 ms acquired with the previous bright blood technique. Conclusion: This study validated the application of a single breathhold black blood single shot fast spin echo acquisition based for human liver T1rho imaging. The lower liver parenchyma T1rho value and higher scan rescan reproducibility may improve of the sensitivity of this technique.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑