使用多反转恢复实时扰相GRE在5T下进行心肌T1定量成像
Myocardial T1 mapping at 5T using multi-inversion recovery real-time spoiled GRE
AI总结:
本文提出基于多反转恢复实时扰相GRE的mIR-rt方法,用于5T心肌T1定量成像,经仿真、体模及16名志愿者验证,其精度优于传统MOLLI序列且可重复性高。
摘要:
目的:开发一种在5T下使用基于Look-Locker的多反转恢复结合实时扰相梯度回波(GRE)采集的准确心肌T1定量成像技术。方法:所提出的T1定量成像技术(mIR-rt)使用实时GRE对反转磁化强度的恢复过程进行采样,并选取舒张期捕获的图像进行T1拟合。采用多次反转恢复来增加样本量,以实现准确拟合。该T1定量成像方法通过Bloch仿真、体模研究以及在5T下对16名健康志愿者进行的在体研究进行了验证。主要结果:在仿真和体模研究中,mIR-rt测得的T1值与参考T1值非常接近,误差小于3%,而传统MOLLI序列会低估T1值。5T下的心肌T1值在心尖、中间和基底分别为1553±52 ms、1531±53 ms和1526±60 ms(均值±标准差)。MOLLI测得的T1值(心尖、中间和基底分别为1350±48 ms、1349±47 ms和1354±45 ms)显著低于mIR-rt的测量值,三个层面的p值均小于0.001。本研究使用的mIR-rt序列方法具有较高的可重复性,尤其是在中间层面,支持其在心肌T1定量成像中的实际应用价值。意义:所提出的方法在5T下进行心肌T1定量成像是可行的,并且比传统MOLLI序列具有更好的准确性。
英文摘要:
Objective: To develop an accurate myocardial T1 mapping technique at 5T using Look-Locker-based multiple inversion-recovery with the real-time spoiled gradient echo (GRE) acquisition. Approach: The proposed T1 mapping technique (mIR-rt) samples the recovery of inverted magnetization using the real-time GRE and the images captured during diastole are selected for T1 fitting. Multiple-inversion recoveries are employed to increase the sample size for accurate fitting. The T1 mapping method was validated using Bloch simulation, phantom studies, and in 16 healthy volunteers at 5T. Main results: In both simulation and phantom studies, the T1 values measured by mIR-rt closely approximate the reference T1 values, with errors less than 3%, while the conventional MOLLI sequence underestimates T1 values. The myocardial T1 values at 5T are 1553 +/- 52 ms, 1531 +/- 53 ms, and 1526 +/- 60 ms (mean +/- standard deviation) at the apex, middle, and base, respectively. The T1 values measured by MOLLI (1350 +/- 48 ms, 1349 +/- 47 ms, and 1354 +/- 45 ms at the apex, middle, and base) were significantly lower than those of mIR-rt with p<0.001 for all three layers. The mIR-rt sequence method used in our study provides high reproducibility, particularly in the middle slices, supporting its practical relevance for myocardial T1 mapping. Significance: The proposed method is feasible for myocardial T1 mapping at 5T and provides better accuracy than the conventional MOLLI sequence.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。