一种双同位素正电子发射断层成像的图像重建方法

Image reconstruction method for dual-isotope positron emission tomography

Tomonori Fukuchi, Mika Shigeta, Hiromitsu Haba, Daiki Mori, Takuya Yokokita, Yukiko Komori, Seiichi Yamamoto, Yasuyoshi Watanabe

摘要:

我们开发了一种用于多同位素成像的正电子发射断层成像(PET)系统。我们的PET系统,命名为多同位素PET(MI-PET),可以区分不同的示踪核素使用符合测量的即时伽马射线,这是发射后,正电子发射。在具有纯正电子发射体和瞬发-γ发射体的MI-PET成像中,由于瞬发γ射线检测的不稳定性,由MI-PET拍摄的纯正电子发射体的图像被正电子-γ发射体叠加。因此,为了使孤立的纯正电子发射体的图像,我们开发了基于特定于MI-PET的数据相减的图像重建方法。我们测试了两种方法,重建图像之间的减法和正弦图数据之间的减法。在这两种方法中,除了探测器灵敏度归一化之外,还需要对瞬发γ射线灵敏度的位置依赖性进行归一化。对于这些归一化,我们使用正电子-伽马发射体Sc-44 m和Na-22的圆柱体进行归一化扫描。利用Sc-44 m的放射性衰变(半衰期58.6小时)进行的长周期测量阐明了瞬发γ射线与PET事件和纯PET事件之间的采集比根据物体活动而变化。因此,我们开发了一种校正方法,该方法涉及依赖于活动的减法参数,即,计数率。根据双示踪剂体模图像的分析,在具有灵敏度校正的正弦图数据中的数据减法对于纯正电子发射体给出了比来自图像减法的那些更高质量的隔离图像。此外,从双同位素(F-18-FDG和Sc-44 m)小鼠成像,我们得出结论,我们开发的方法可以用于实际的活体成像。

英文摘要:

We developed a positron emission tomography (PET) system for multiple-isotope imaging. Our PET system, named multiple-isotope PET (MI-PET), can distinguish between different tracer nuclides using coincidence measurement of prompt gamma-rays, which are emitted after positron emission. In MI-PET imaging with a pure positron emitter and prompt-gamma emitter, because of the imperfectness of prompt gamma-ray detection, an image for a pure positron emitter taken by MI-PET is superposed by a positron-γ emitter. Therefore, in order to make isolated images of the pure positron emitter, we developed image reconstruction methods based on data subtraction specific to MI-PET. We tested two methods, subtraction between reconstructed images and subtraction between sinogram data. In both methods, normalization for position dependence of the prompt γ-ray sensitivity is required in addition to detector sensitivity normalization. For these normalizations, we performed normalization scans using cylindrical phantoms of the positron-gamma emitters Sc-44m and Na-22. A long period measurement using the activity decay of Sc-44m (Half-life 58.6 hours) elucidated that the acquisition ratio between the prompt gamma-rays coincided with PET event and pure PET event changes on the basis of object activities. Therefore, we developed a correction method that involves subtraction parameters dependent on the activities, i.e., the counting rate. From analysis of dual-tracer phantom images, data subtraction in the sinogram data with sensitivity correction gives a higher quality of isolated images for the pure positron emitter than those from image subtractions. Furthermore, from dual-isotope (F-18-FDG and Sc-44m) mouse imaging, we concluded that our developed method can be used for practical imaging of a living organism.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑