在使用β-γ符合的正电子发射断层成像中具有高灵敏度的亚毫米核医学成像

Sub-millimeter nuclear medical imaging with high sensitivity in positron emission tomography using beta-gamma coincidences

C. Lang, D. Habs, K. Parodi, P. G. Thirolf

摘要:

我们提出了一种核医学成像技术,采用从β ^+ -γ符合的三重γ轨迹相交,能够达到亚毫米空间分辨率的3维与减少的要求,重建每个体素的交叉相比,传统的PET重建分析。这种“$γ$ -PET”技术利用特定的β ^+衰变同位素,同时发射出额外的光子。利用正电子湮没和第三光子之间的三重符合,可以将重建的“真实”事件与背景分离。为了表征这种技术,已经进行了蒙特卡罗模拟和图像重建。已发现可实现的空间分辨率达到ca。0.4 mm(FWHM),用于22 Na点源的可视化。只有40个交叉点足以对嵌入在大约1 mm^3的体素体积内的水的散射体积中的点源进行可靠的亚毫米图像重建(“高分辨率模式”)。此外,从400 MBq的^76Br注入活度开始,在2 x 2 x 3 mm^3的较大体素体积(“高灵敏度模式”)的情况下,仅需要大约40个重建交叉点的相同数量。需要这样低数量的重建事件显著地减少了图像重建所需的采集时间(在上述情况下为大约140秒),并且因此可以打开准实时成像的前景。

英文摘要:

We present a nuclear medical imaging technique, employing triple-gamma trajectory intersections from beta^+ - gamma coincidences, able to reach sub-millimeter spatial resolution in 3 dimensions with a reduced requirement of reconstructed intersections per voxel compared to a conventional PET reconstruction analysis. This '$γ$-PET' technique draws on specific beta^+ - decaying isotopes, simultaneously emitting an additional photon. Exploiting the triple coincidence between the positron annihilation and the third photon, it is possible to separate the reconstructed 'true' events from background. In order to characterize this technique, Monte-Carlo simulations and image reconstructions have been performed. The achievable spatial resolution has been found to reach ca. 0.4 mm (FWHM) in each direction for the visualization of a 22Na point source. Only 40 intersections are sufficient for a reliable sub-millimeter image reconstruction of a point source embedded in a scattering volume of water inside a voxel volume of about 1 mm^3 ('high-resolution mode'). Moreover, starting with an injected activity of 400 MBq for ^76Br, the same number of only about 40 reconstructed intersections are needed in case of a larger voxel volume of 2 x 2 x 3~mm^3 ('high-sensitivity mode'). Requiring such a low number of reconstructed events significantly reduces the required acquisition time for image reconstruction (in the above case to about 140 s) and thus may open up the perspective for a quasi real-time imaging.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑