GPU-based Fast Low-dose Cone Beam CT Reconstruction via Total Variation

Xun Jia, Yifei Lou, John Lewis, Ruijiang Li, Xuejun Gu, Chunhua Men, William Y. Song, Steve B. Jiang

摘要:

锥形束CT(CBCT)已广泛用于图像引导放射治疗(IGRT),以在治疗分次之前获取更新的体积解剖信息,从而实现精确的患者对准目的。然而,在大多数IGRT程序中,来自连续CBCT扫描的过量X射线成像剂量引起了临床关注。通过减少CBCT扫描中的X射线投影数量和/或降低mAs水平,可以有效减少过量的成像剂量。本文的目标是开发一种基于GPU的快速算法,从欠采样和噪声投影数据中重建高质量的CBCT图像,以降低成像剂量。通过最小化由数据保真度项和总变分正则化项组成的能量泛函来重建CBCT。我们开发了一个GPU友好版本的前向-后向分裂算法来解决这个模型。还采用了多重网格技术。我们在数字NCAT体模和头颈部患者病例上测试了我们的CBCT重建算法。还使用物理Catphan体模和头颈Rando体模验证了低mAs下的性能。研究发现,40个X射线投影足以重建质量令人满意的CBCT图像,用于IGRT患者对准目的。体模实验表明,CBCT图像可以成功地重建低至0.1 mAs/投影水平与我们的算法。与目前广泛使用的全扇形头颈部扫描方案(约360个投影,0.4 mAs/投影)相比,估计使用我们的算法已实现总体剂量减少36倍。此外,在NVIDIA Tesla C1060 GPU卡上的重建时间约为130秒,估计比类似的迭代重建方法快约100倍。

英文摘要:

Cone-beam CT (CBCT) has been widely used in image guided radiation therapy (IGRT) to acquire updated volumetric anatomical information before treatment fractions for accurate patient alignment purpose. However, the excessive x-ray imaging dose from serial CBCT scans raises a clinical concern in most IGRT procedures. The excessive imaging dose can be effectively reduced by reducing the number of x-ray projections and/or lowering mAs levels in a CBCT scan. The goal of this work is to develop a fast GPU-based algorithm to reconstruct high quality CBCT images from undersampled and noisy projection data so as to lower the imaging dose. The CBCT is reconstructed by minimizing an energy functional consisting of a data fidelity term and a total variation regularization term. We developed a GPU-friendly version of the forward-backward splitting algorithm to solve this model. A multi-grid technique is also employed. We test our CBCT reconstruction algorithm on a digital NCAT phantom and a head-and-neck patient case. The performance under low mAs is also validated using a physical Catphan phantom and a head-and-neck Rando phantom. It is found that 40 x-ray projections are sufficient to reconstruct CBCT images with satisfactory quality for IGRT patient alignment purpose. Phantom experiments indicated that CBCT images can be successfully reconstructed with our algorithm under as low as 0.1 mAs/projection level. Comparing with currently widely used full-fan head-and-neck scanning protocol of about 360 projections with 0.4 mAs/projection, it is estimated that an overall 36 times dose reduction has been achieved with our algorithm. Moreover, the reconstruction time is about 130 sec on an NVIDIA Tesla C1060 GPU card, which is estimated ~100 times faster than similar iterative reconstruction approaches.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑