arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2409.04674physics.med-ph
收藏

基于谱间结构相似性正则化的优化图像重建用于有限角度双能锥形束CT

Optimization-Based Image Reconstruction Regularized with Inter-Spectral Structural Similarity for Limited-Angle Dual-Energy Cone-Beam CT

Junbo Peng, Tonghe Wang, Huiqiao Xie, Richard L. J. Qiu, Chih-Wei Chang, Justin Roper, David S. Yu, Xiangyang Tang, Xiaofeng Yang

AI总结:

本研究提出一种无需X射线光谱测量或成对训练数据、基于谱间结构相似性正则化的迭代重建方法,用于有限角度双能锥形束CT,有效减少伪影并实现准确定量成像。

摘要:

背景:有限角度(LA)双能(DE)锥形束CT(CBCT)被认为是无需硬件改造即可在当前CBCT扫描仪上实现快速、低剂量双能成像的潜在解决方案。然而,其临床应用受到有限角度投影图像重建挑战的阻碍。尽管已提出基于优化和基于深度学习的方法用于图像重建,但其应用受限于需要X射线光谱测量或成对数据集进行模型训练。目的:本研究旨在通过开发LA-DECBCT中图像重建的实用解决方案,促进快速、低剂量DECBCT的临床应用。方法:将基于谱间结构相似性的正则化集成到LA-DECBCT的迭代图像重建中。通过强制双能图像之间的相似性,有效减少了重建DECBCT图像中的LA伪影。使用四个物理体模和三个数字体模对所提方法进行了评估,证明了其在定量DECBCT成像中的有效性。结论:所提方法无需X射线光谱测量用于优化,也无需成对数据集用于模型训练,即可实现准确的图像重建,在LA-DECBCT的临床实施中显示出巨大的实用价值。

英文摘要:

Background: Limited-angle (LA) dual-energy (DE) cone-beam CT (CBCT) is considered as a potential solution to achieve fast and low-dose DE imaging on current CBCT scanners without hardware modification. However, its clinical implementations are hindered by the challenging image reconstruction from LA projections. While optimization-based and deep learning-based methods have been proposed for image reconstruction, their utilization is limited by the requirement for X-ray spectra measurement or paired datasets for model training. Purpose: This work aims to facilitate the clinical applications of fast and low-dose DECBCT by developing a practical solution for image reconstruction in LA-DECBCT. Methods: An inter-spectral structural similarity-based regularization was integrated into the iterative image reconstruction in LA-DECBCT. By enforcing the similarity between the DE images, LA artifacts were efficiently reduced in the reconstructed DECBCT images. The proposed method was evaluated using four physical phantoms and three digital phantoms, demonstrating its efficacy in quantitative DECBCT imaging. Conclusions: The proposed method achieves accurate image reconstruction without the need for X-ray spectra measurement for optimization or paired datasets for model training, showing great practical value in clinical implementations of LA-DECBCT.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑