arXiv 1401.0692physics.med-ph
收藏

CyberKnife小射野肺和骨等效介质的非均匀体模中的百分比深度剂量分布

Percentage depth dose distributions in inhomogeneous phantoms with lung and bone equivalent media for small fields of CyberKnife

Chung Il Lee, Jae Won Shin, Sei-Chul Yoon, Tae Suk Suh, Seung-Woo Hong, Kyung Joo Min, Sang Deok Lee, Su Mi Chung, Jae-Yong Jung

摘要:

本文研究了具有肺和骨等效介质的非均匀体模中的百分深度剂量分布。对于肺等效介质,使用Balsa木材,对于软骨等效介质,使用具有环氧树脂、硬化剂和碳酸钙的复合材料。将聚苯乙烯板与这些材料放在一起用作非均匀体模。在Seoul Uridul医院,通过使用来自6MV射波刀的光子束,使用Gafchromic EBT薄膜进行剂量测量。光子束的圆锥尺寸从5、10到30 mm不等。使用GEANT 4 Monte Carlo代码v9.4.p02作为模拟工具。当Balsa木材插入体模中时,发现使用EBT膜测量的剂量与Balsa木材区域内外没有EBT膜的剂量显著不同,特别是对于小射野尺寸。另一方面,当在体模中插入软骨等效材料时,仅在骨等效材料区域可以看到使用EBT胶片测量的剂量与不使用EBT胶片测量的剂量之间的差异。GEANT4模拟在有和没有EBT膜的情况下进行,以将模拟结果与测量结果进行比较。我们发现,包括EBT膜的模拟同意与测量的误差为$2.2\%$的所有情况下。此外,我们发现没有EBT薄膜的“体模剂量”与高达$29\%$的“薄膜剂量”不同,这表明,对于带有EBT薄膜的非均匀体模的准确剂量估计,需要适当考虑EBT薄膜的存在,特别是对于小射野。

英文摘要:

The percentage depth dose distributions in inhomogeneous phantoms with lung and bone equivalent media are studied. For lung equivalent media a Balsa wood is used, and for a soft bone equivalent media a compound material with epoxy resin, hardener and calcium carbonate is used. Polystyrene slabs put together with these materials are used as an inhomogeneous phantom. Dose measurements are performed with Gafchromic EBT film by using photon beams from 6MV CyberKnife at the Seoul Uridul Hospital. The cone sizes of the photon beams are varied from 5, 10 to 30 mm. As a simulation tool GEANT4 Monte Carlo code v9.4.p02 is used. When the Balsa wood is inserted in the phantom, the dose measured with EBT film is found to be significantly different from the dose without the EBT film in and beyond the Balsa wood region, particularly for small field sizes. On the other hand, when the soft bone equivalent material is inserted in the phantom, discrepancy between the dose measured with EBT film and the dose without EBT film can be seen only in the region of bone equivalent material. GEANT4 simulations are done with and without the EBT film to compare the simulation results with measurements. We find that the simulations including EBT film agree with the measurements for all the cases within an error of $2.2\%$. Also, we find the "doses to phantom" without the EBT film differ from the "doses to film" up to $29\%$, which shows that for accurate dose estimations for inhomogeneous phantoms with EBT film the presence of the EBT film needs to be taken into account properly in particular for small fields.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑