实用CBCT剂量优化方法的提出与评估

Proposal and Evaluation of a Practical CBCT Dose Optimization Method

S. Gros, J. Bian, J. Jackson, M. Delafuente, H. Kang, W. Small, M. Mahesh

AI总结:

提出一种基于科室常用设备的CBCT剂量优化方法,通过调整投影数量和单帧曝光降低累积剂量,在保持临床可接受图像质量的同时实现有效剂量降低,且无需医学物理专业知识即可实施。

摘要:

目的:提出一种使用放射肿瘤科 readily available 测量设备的 CBCT 剂量优化方法。方法:在 Varian TrueBeam 直线加速器上,使用 0.6cc 空气比释动能(Kair)校准的 Farmer 电离室测量五种默认 CBCT 协议(HEAD、PELVIS、LARGE PELVIS、THORAX、SPOTLIGHT)在等中心的 Kair。通过在 FDK 重建中减少投影数量(减少25%、50%、75%),或将单帧曝光降低25%、50%、70%至75%来修改成像参数,以减少累积曝光。在 CTDI 体模中测量锥形束剂量指数(CBDI)和加权 CBDI(CBDI_w),以与 Kair 建立相关性。使用 Catphan604 定量评估图像质量(HU线性、均匀性、低对比度可见性、高对比度分辨率、空间完整性),并通过拟人化体模进行定性评估。结果:建立了 Kair、CTDI 和 CBDI_w 之间的线性关系。高对比度分辨率和空间完整性不受剂量降低影响(R < 0.5),而低对比度可见性与曝光强相关(R > 0.75)。HU均匀性和线性保持稳定(<36 HU偏差),但 LARGE PELVIS 在≤50%单帧曝光下除外。80%至50%的剂量降低保留了临床可接受的图像质量,伪影或噪声极小。意义:TrueBeam 直线加速器上的 CBCT 协议得以优化。剂量降低。该方法使放射治疗师无需医学物理设备或专业知识即可轻松实施。

英文摘要:

Objective: To propose a CBCT dose optimization method using readily available measurement equipment in radiation oncology departments. Approach: A 0.6cc air kerma (Kair) calibrated Farmer chamber measured Kair at isocenter for five default CBCT protocols (HEAD, PELVIS, LARGE PELVIS, THORAX, SPOTLIGHT) on a Varian TrueBeam linac. Imaging parameters were modified to reduce cumulative exposure by reducing projections (by 25%, 50%, 75%) in FDK reconstruction, or lowering single frame exposure by 25%, 50%, 70 to 75%. Cone Beam Dose Index (CBDI) and weighted CBDI (CBDI_w) were measured in CTDI phantoms to correlate with Kair. Image quality was assessed quantitatively (HU linearity, uniformity, low-contrast visibility, high-contrast resolution, spatial integrity) using Catphan604, and qualitatively via anthropomorphic phantoms. Results: A linear relationship between Kair, CTDI, and CBDI_w was established. High-contrast resolution and spatial integrity were unaffected by dose reduction (R < 0.5), while low-contrast visibility strongly correlated with exposure (R > 0.75). HU uniformity and linearity remained stable (<40 HU deviation), except for LARGE PELVIS at <50% single-frame exposure. 80 to 50% dose reduction retained clinically acceptable image quality with minimal artifacts or noise. Significance: CBCT protocols on TrueBeam linacs were optimized for dose reduction. The method enables easy implementation by radiotherapy physicists without diagnostic physics equipment or expertise.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑