用于实时远程心脏健康评估的单导联ECG参数设备端计算:一项真实世界验证研究

On-device Computation of Single-lead ECG Parameters for Real-time Remote Cardiac Health Assessment: A Real-world Validation Study

College of Integrative Chinese and Western Medicine, Anhui University of Chinese Medicine(安徽中医药大学中西医结合学院);HeartVoice Medical Technology(心音医疗科技);The First Affiliated Hospital of Anhui Medical University(安徽医科大学第一附属医院);National Institute of Health Data Science, Peking University(北京大学健康医疗大数据国家研究院);The First Affiliated Hospital of University of Science and Technology of China(中国科学技术大学附属第一医院);University of Science and Technology of China(中国科学技术大学);Key Laboratory of Xinan Medicine of Ministry of Education, Anhui University of Chinese Medicine(安徽中医药大学教育部新安医学重点实验室);Peking University Health Science Center(北京大学医学部);Institute for Artificial Intelligence, Peking University(北京大学人工智能研究院);Peking University First Hospital(北京大学第一医院)

Sumei Fan, Deyun Zhang, Yue Wang, Shijia Geng, Kun Lu, Meng Sang, Weilun Xu, Haixue Wang, Qinghao Zhao, Chuandong Cheng, Peng Wang, Shenda Hong

AI总结:

本研究验证了设备端算法FeatureDB在单导联ECG参数计算上的表现,发现其准确度达医生级且异常检测性能优于开源算法、媲美商业系统,支持可扩展的远程心脏医疗。

摘要:

准确、连续的院外心电图(ECG)参数测量对于实时心脏健康监测和远程医疗至关重要。单导联ECG参数的设备端计算能够在不依赖集中式数据处理的情况下实现及时评估,推动了个性化、无处不在的心脏护理发展——然而,在异构真实世界人群中的全面验证仍然有限。本研究使用两个数据集验证了设备端算法FeatureDB(https://github.com/PKUDigitalHealth/FeatureDB):HeartVoice-ECG-lite(369名参与者,具有由两名医生标注的单导联ECG)和PTB-XL/PTB-XL+(21,354名患者,具有12导联ECG和医生诊断标注)。FeatureDB计算了PR、QT和QTc间期,通过平均绝对误差(MAE)、相关性分析和Bland-Altman分析评估其与医生标注相比的准确性。针对一度房室传导阻滞(AVBI,基于PR)和长QT综合征(LQT,基于QTc)的诊断性能,使用AUC、准确率、敏感性和特异性与商业12导联系统(12SL,Uni-G)及开源算法Deli进行了基准比较。结果显示,其与专家标注高度一致(Pearson相关系数:0.836-0.960),MAE与观察者间变异度相匹配,且偏差极小。AVBI的AUC达到0.787(12SL:0.859;Uni-G:0.812;Deli:0.501);LQT的AUC为0.684(12SL:0.716;Uni-G:0.605;Deli:0.569)——与商业工具相当且优于开源替代方案。FeatureDB通过单导联设备提供医生级参数准确性和商业级异常检测,支持社区和门诊环境中的可扩展远程医疗、去中心化心脏筛查和持续监测。

英文摘要:

Accurate, continuous out-of-hospital electrocardiogram (ECG) parameter measurement is vital for real-time cardiac health monitoring and telemedicine. On-device computation of single-lead ECG parameters enables timely assessment without reliance on centralized data processing, advancing personalized, ubiquitous cardiac care-yet comprehensive validation across heterogeneous real-world populations remains limited. This study validated the on-device algorithm FeatureDB (https://github.com/PKUDigitalHealth/FeatureDB) using two datasets: HeartVoice-ECG-lite (369 participants with single-lead ECGs annotated by two physicians) and PTB-XL/PTB-XL+ (21,354 patients with 12-lead ECGs and physicians' diagnostic annotations). FeatureDB computed PR, QT, and QTc intervals, with accuracy evaluated against physician annotations via mean absolute error (MAE), correlation analysis, and Bland-Altman analysis. Diagnostic performance for first-degree atrioventricular block (AVBI, PR-based) and long QT syndrome (LQT, QTc-based) was benchmarked against commercial 12-lead systems (12SL, Uni-G) and open-source algorithm Deli, using AUC, accuracy, sensitivity, and specificity. Results showed high concordance with expert annotations (Pearson correlations: 0.836-0.960), MAEs matching inter-observer variability, and minimal bias. AVBI AUC reached 0.787 (12SL: 0.859; Uni-G: 0.812; Deli: 0.501); LQT AUC was 0.684 (12SL: 0.716; Uni-G: 0.605; Deli: 0.569)-comparable to commercial tools and superior to open-source alternatives. FeatureDB delivers physician-level parameter accuracy and commercial-grade abnormality detection via single-lead devices, supporting scalable telemedicine, decentralized cardiac screening, and continuous monitoring in community and outpatient settings.