用于有限角度(LA)-CBCT重建的先验频率引导扩散模型

Prior Frequency Guided Diffusion Model for Limited Angle (LA)-CBCT Reconstruction

University of Texas Southwestern Medical Center(德克萨斯大学西南医学中心)

Jiacheng Xie, Hua-Chieh Shao, Yunxiang Li, You Zhang

AI总结:

针对有限角度CBCT重建的严重欠采样伪影问题,提出利用患者先验CT高频信息引导的条件扩散模型PFGDM及其两种条件策略,显著提升了超小机架角下的重建质量。

摘要:

锥形束计算机断层扫描(CBCT)被广泛应用于图像引导放射治疗。为了提高成像效率、降低剂量并改善机械间隙,从有限角度采集(LA-CBCT)重建CBCT是非常理想的。然而,LA-CBCT重建受到严重欠采样伪影的影响,使其成为一个高度不适定的逆问题。扩散模型可以通过学习到的数据分布逆转数据加噪过程来生成数据/图像;并且可以作为去噪器/正则化项纳入LA-CBCT重建。在本研究中,我们开发了一种基于扩散模型的框架,即先验频率引导扩散模型(PFGDM),用于鲁棒且保持结构的LA-CBCT重建。PFGDM使用条件扩散模型作为LA-CBCT重建的正则化项,该条件基于从患者特异性先验CT扫描中提取的高频信息,为LA-CBCT重建提供了强大的解剖学先验。具体而言,我们开发了PFGDM的两个变体(PFGDM-A和PFGDM-B),它们具有不同的条件方案。PFGDM-A应用高频CT信息条件直到一个预优化的迭代步骤,之后将其丢弃,以使相似和不同的CT/CBCT解剖结构都能被重建。另一方面,PFGDM-B在每个重建步骤中持续应用先验CT信息条件,同时采用衰减机制,以逐步淘汰来自先验CT扫描的重建引导。PFGDM的两个变体通过包括PSNR和SSIM在内的指标进行了测试,并与当前可用的LA-CBCT重建解决方案进行了比较。PFGDM优于所有传统方法和基于扩散模型的方法。PFGDM可在非常有限的机架角度下重建高质量LA-CBCT,从而实现更快、更灵活且降低剂量的CBCT扫描。

英文摘要:

Cone-beam computed tomography (CBCT) is widely used in image-guided radiotherapy. Reconstructing CBCTs from limited-angle acquisitions (LA-CBCT) is highly desired for improved imaging efficiency, dose reduction, and better mechanical clearance. LA-CBCT reconstruction, however, suffers from severe under-sampling artifacts, making it a highly ill-posed inverse problem. Diffusion models can generate data/images by reversing a data-noising process through learned data distributions; and can be incorporated as a denoiser/regularizer in LA-CBCT reconstruction. In this study, we developed a diffusion model-based framework, prior frequency-guided diffusion model (PFGDM), for robust and structure-preserving LA-CBCT reconstruction. PFGDM uses a conditioned diffusion model as a regularizer for LA-CBCT reconstruction, and the condition is based on high-frequency information extracted from patient-specific prior CT scans which provides a strong anatomical prior for LA-CBCT reconstruction. Specifically, we developed two variants of PFGDM (PFGDM-A and PFGDM-B) with different conditioning schemes. PFGDM-A applies the high-frequency CT information condition until a pre-optimized iteration step, and drops it afterwards to enable both similar and differing CT/CBCT anatomies to be reconstructed. PFGDM-B, on the other hand, continuously applies the prior CT information condition in every reconstruction step, while with a decaying mechanism, to gradually phase out the reconstruction guidance from the prior CT scans. The two variants of PFGDM were tested and compared with current available LA-CBCT reconstruction solutions, via metrics including PSNR and SSIM. PFGDM outperformed all traditional and diffusion model-based methods. PFGDM reconstructs high-quality LA-CBCTs under very-limited gantry angles, allowing faster and more flexible CBCT scans with dose reductions.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑