利用深度图像先验对PET动态图像降噪的体模适用性测试

Applicability test for reducing noise on PET dynamic images using phantom applying deep image prior

Nobuyuki Kudomi, Yukito Maeda

AI总结:

本文提出在PET动态图像中应用深度图像先验(DIP)进行降噪,并通过体模实验验证该方法可在不显著改变衰减速率的前提下降低重建图像和速率常数图像的噪声。

摘要:

目的:正电子发射断层成像(PET)能够以动态图像中速率常数的形式对病理生理信息进行成像。速率常数图像可能受到动态图像噪声的影响。我们引入了一种名为深度图像先验(DIP)的人工智能技术来降低动态图像中的噪声。方法:我们使用了一个体模,其主圆柱体和子圆柱体中分别填充了18F-F-溶液和11C-flumazenil溶液。该体模由Biograph mCT扫描并获得动态图像。将DIP应用于动态图像涉及的所有切片,同时引入一个指标来选择使噪声程度最小化的最优epoch。随后生成衰减速率图像,并测量图像中的定量准确性和质量。结果:图像上获得的衰减速率与参考值之间没有显著差异。在11C和18F填充区域中,无论是重建图像还是衰减速率图像,基于DIP的图像其变异系数(CV)均小于原始图像。结论:本方法在DIP过程中选择生成图像,对于动态PET图像的降噪以及获得噪声更小的速率常数图像是可行的。

英文摘要:

Objective Positron emission tomography (PET) allows imaging of patho-physiological information as a form of rate constants from a dynamic image. The rate constant image(s) may be affected from noise on the dynamic image. We introduced an artificial intelligence technique of deep image prior (DIP) to reduce noise on dynamic images. Method We utilized a phantom filled with 18F-F- and 11C-flumazenil solutions in the main and sub-cylinders, respectively. The phantom was scanned by a Biograph mCT and dynamic images were obtained. DIP was applied to all slices involved in the dynamic images while introducing an index for choosing an optimal epoch with minimize the degree of noise. Then, decay rate images were generated and quantitative accuracy and quality were measured in the images. Results The obtained decay rates on images were not significantly different from those of the reference values. Coefficient of variances (CV) were smaller using DIP-based images than those from original images for the reconstructed and decay rate images, both in the 11C and 18F filled regions. Conclusion The present method for choosing a generated image during DIP is feasible for noise reduction in dynamic PET images, and for obtaining rate constant images with less noise.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑