Purpose: Beam range control is the essence of radiotherapy with heavy charged particles. In conventional broad-beam delivery, fine range adjustment is achieved by insertion of range shifting and compensating materials. In dosimetry, solid phantoms are often used for convenience. These materials should ideally be equivalent to water. In this study, we evaluated dosimetric water equivalence of four common plastics, HDPE, PMMA, PET, and POM.
Methods: Using the Bethe formula for energy loss, the Gottschalk formula for multiple scattering, and the Sihver formula for nuclear interactions, we calculated the effective densities of the plastics for these interactions. We experimentally measured variation of the Bragg peak of carbon-ion beams by insertion of HDPE, PMMA, and POM, which were compared with analytical model calculations.
Results: The theoretical calculation resulted in slightly reduced multiple scattering and severely increased nuclear interactions for HDPE, compared to water and the other plastics. The increase in attenuation of carbon ions for 20-cm range shift was experimentally measured to be 8.9\% for HDPE, 2.5\% for PMMA, and 0.0\% for POM while PET was theoretically estimated to be in between PMMA and POM. The agreement between the measurements and the calculations was about 1\% or better.
Conclusions: For carbon-ion beams, POM was dosimetrically indistinguishable from water and the best of the plastics examined in this study. The poorest was HDPE, which would reduce the Bragg peak by 0.45\% per 1-cm range shift, although with marginal superiority for reduced multiple scattering. Between the two clear plastics, PET would be superior to PMMA in dosimetric water equivalence.
Wide Area Surface Dosimetry with Conformal Scintillator Array for External Beam Radiotherapy
Thayer School of Engineering, Dartmouth College(达特茅斯学院泰勒工程学院);Department of Radiation Oncology and Applied Sciences, Geisel School of Medicine, Dartmouth College(达特茅斯学院吉泽尔医学院放射肿瘤与应用科学系)
Characterization of a novel plastic scintillation detector for in vivo electron dosimetry
University Medical Centre Mannheim, Heidelberg University(海德堡大学曼海姆医学中心);DKFZ-Hector Cancer Institute, University Medical Centre Mannheim(德国癌症研究中心-海克特癌症研究所,曼海姆医学中心);Mannheim Institute for Intelligent Systems in Medicine (MIISM), Medical Faculty Mannheim, Heidelberg University(海德堡大学曼海姆医学院曼海姆智能医学系统研究所)
Adapting a 3D scanning water phantom for use in brachytherapy dosimetry
Royal Brisbane & Women’s Hospital(皇家布里斯班妇女医院);Herston Biofabrication Institute(赫斯顿生物制造研究所);University of Queensland(昆士兰大学);Queensland University of Technology(昆士兰科技大学)
A GPU-based Monte Carlo framework for IMRT QA using EPID transit dosimetry
School of Nuclear Science and Technology, University of Science and Technology of China(中国科学技术大学核科学与技术学院);Department of Radiation Oncology, First Affiliated Hospital of Anhui Medical University(安徽医科大学第一附属医院肿瘤放射科);School of Medical Imaging, Bengbu Medical College(蚌埠医学院医学影像学院);Anhui Wisdom Technology Co., Ltd.(安徽智慧科技有限公司)
Experimental and computational dosimetry in an integrated workflow for remote audits in Ir-192 interstitial brachytherapy: Development and pilot implementation
National and Kapodistrian University of Athens(雅典国立卡波季斯特里安大学);Alexandra General Hospital of Athens(雅典亚历山德拉综合医院);Greek Atomic Energy Commission (EEAE)(希腊原子能委员会);Hygeia Hospital(海吉亚医院)
Technical Note: Reconstruction of physical and biological dose distributions of carbon-ion beam through deconvolution of longitudinal dosimeter responses