确定低温量热计最佳操作设置的自动化系统

An automated system to define the optimal operating settings of cryogenic calorimeters

Krystal Alfonso, Carlo Bucci, Lucia Canonica, Paolo Carniti, Sergio Di Domizio, Andrea Giachero, Claudio Gotti, Laura Marini, Irene Nutini, Gianluigi Pessina

摘要:

装有NTD热敏电阻的低温宏观量热计已经发展了几十年。最佳偏置电流的选择对于这些探测器的正确操作至关重要,无论是在能量分辨率还是稳定性方面。在本文中,我们提出了一套自动测量和分析程序的表征和优化的NTD晶闸管的工作配置。所提出的程序是为CUORE开发的,CUORE是一个由988个低温宏量热计组成的阵列,配备了NTD热敏电阻,自2017年以来一直在采集数据。这些程序使得能够以可靠的方式表征大量探测器。它们也足够适合用于其他大型低温探测器阵列,如CUPID。

英文摘要:

Cryogenic macro-calorimeters instrumented with NTD thermistors have been developed for several decades. The choice of the optimal bias current is crucial for a proper operation of these detectors, both in terms of energy resolution and stability. In this paper we present a set of automatic measurements and analysis procedures for the characterization and optimization of the working configuration of the NTD thermistors. The presented procedures were developed for CUORE, an array of 988 cryogenic macro-calorimeters instrumented with NTD thermistors that has been taking data since 2017. These procedures made it possible to characterize a large number of detectors in a reliable way. They are suitable enough to be used also in other large arrays of cryogenic detectors, such as CUPID.