Performance of Linear Classification Algorithms on alpha/gamma Discrimination for LaBr3:Ce Scintillation Detectors with Various Pulse Digitizer Properties
Jingjun Wen, Jinfu Zhu, Tao Xue, Jirong Cang, Liangjun Wei, Qiyuan Nie, Ming Zeng, Zhi Zeng, Hao Ma, Jianmin Li, Yinong Liu
With the development of high-speed readout electronics, the digital pulse shape discrimination (PSD) methods have attracted the attention of more researchers, especially in the field of high energy physics and neutron detection. How to choose a PSD algorithm and corresponding data acquisition system (DAQ) naturally becomes a critical problem to settle down for the detection system designers. In this paper, the relationship between the classification performance of different PSD algorithms and digitizers' sampling properties (including sampling rate and the effective number of bits) has been researched based on $\mathrm{LaBr_3}$:Ce scintillation detectors. A self-developed integrated digitizer configured with five different ADCs and a WavePro 404HD oscilloscope were deployed to digitize the waveforms from $\mathrm{LaBr_3}$:Ce scintillators. Moreover, three PSD methods, charge comparison method (CCM), least square for classification method (LS) and Fisher's linear discriminant analysis (LDA), based on linear model were applied to discriminate the alpha signals from the intrinsic isotope $^{227}$Ac. With the LS method and a 125 MSPS 14-Bit ADC, the FoM value was 1.424$\pm$0.042, which is similar to the result from LDA but 31\% better than the result of CCM. The discrimination results showed that the performances of LS and LDA are less affected by the sampling rate with respect to the CCM method, which reflects in a better PSD capability. The results of this paper can help the developers of detector systems to make a trade-off among sampling properties, desirable discrimination results and the cost of systems.
Sufficient current pulse information of nuclear radiation detectors can be retained by direct waveform digitization owing to the improvement of digitizer's performance. In many circumstances, reasonable cost and power consumption are on demand while the energy resolution and PSD performance should be ensured simultaneously for detectors. This paper will quantitatively analyse the influence of vertical resolution and sampling rate of digitizers on the energy resolution and PSD performance. The energy resolution and PSD performance can be generally optimized by improving the sampling rate and EN…
The proton light yield of organic scintillators has been measured extensively in recent years using fast waveform digitizers and large discrepancies exist in the values reported by different authors. In this letter, we address principles of digital signal processing that must be considered when conducting scintillator light yield measurements. Digitized waveform pulse height values are only proportional to the amount of scintillation light if the temporal shape of the scintillation pulse is independent of the amount of energy deposited. This is not the case for scintillation pulses resulting f…
The CAEN V1751 is a new generation of Waveform Digitizer recently introduced by CAEN SpA. It features 8 Channels per board, 10 bit, 1 GS/s using Flash ADCs Waveform Digitizers (or 4 channels at 2 GS/s in Dual Edge Sampling mode) with threshold and Auto-Trigger capabilities. This provides a good basis for data acquisition in Dark Matter searches using PMTs to detect scintillation light in liquid argon, as it matches the requirements for measuring the fast scintillation component. The board was tested by operating it in real experimental conditions and by comparing it with a state of the art dig…
Signal digitizers revolutionized the approach to the electronics readout of radiation detectors in Nuclear Physics. These highly specialized pieces of equipment are designed to acquire the signals that are characteristic of the detectors in nuclear physics experiments. The functions of the several modules that were once needed for signal acquisition, can now be substituted by a single digitizer. As suggested by the name, with such readout modules, signals are first digitized (i.e. the signal waveform is sampled and converted to a digital representation) and then either stored or analyzed on-th…
In beam test experiments have been carried out for particle identification using digital pulse shape analysis in a 500~$μ$m thick Neutron Transmutation Doped (nTD) silicon detector with an indigenously developed FPGA based 12 bit resolution, 1 GHz sampling digitizer. The nTD Si detector was used in a low-field injection setup to detect light heavy-ions produced in reactions of $\sim$ 5 MeV/A $^{7}$Li and $^{12}$C beams on different targets. Pulse height, rise time and current maximum have been obtained from the digitized charge output of a high bandwidth charge and current sensitive pre-amplif…
Simulation studies have been carried out to explore the ability to discriminate between single-site and multi-site energy depositions in large scale liquid scintillation detectors. A robust approach has been found that is predicted to lead to a significant statistical separation for a large variety of event classes, providing a powerful tool to discriminate against backgrounds and break important degeneracies in signal extraction. This has particularly relevant implications for liquid scintillator searches for neutrinoless double beta decay ($0νββ$) from $^{130}$Te and $^{136}$Xe, where it is …
A two-phase argon detector has high discrimination power between electron recoil and nuclear recoil events based on the pulse shape discrimination and the ionization/scintillation ratio (S2/S1). This character is very suitable for the dark matter search to establish the low background experiment. However, the basic properties of S2/S1 of argon are not well known, as compared with xenon. We report the evaluation of S2/S1 properties with a two-phase detector at drift-fields of 0.2-3.0 kV/cm. Finally, the discrimination power against electron recoil background of S2/S1 is discussed.
The SNO+ liquid scintillator experiment is under construction in the SNOLAB facility in Canada. The success of this experiment relies upon accurate characterization of the liquid scintillator, linear alkylbenzene (LAB). In this paper, scintillation decay times for alpha and electron excitations in LAB with 2 g/L PPO are presented for both oxygenated and deoxygenated solutions. While deoxygenation is expected to improve pulse shape discrimination in liquid scintillators, it is not commonly demonstrated in the literature. This paper shows that for linear alkylbenzene, deoxygenation improves disc…
A successfull application of Geiger-mode multipixel avalanche diodes (GMAPDs) for pulse-shape discrimination in alpha-beta spectrometry using organic liquid scintillator is described in this paper. Efficient discrimination of alpha and beta components in the emission of radioactive isotopes is achieved for alpha energies above 0.3 MeV. The ultra-compact design of the scintillating detector helps to efficiently suppress cosmic-ray and ambient radiation background. This approach allows construction of hand-held robust devices for monitoring of radioactive contamination in various environmental c…
We have developed a detector, consisting of a cryogenic calorimeter with a scintillating crystal as absorber, and a second calorimeter for the detection of the scintillation light, both operated at 12 mK. Using a CaWO4 crystal with a mass of 6g as scintillating absorber, we have achieved a discrimination of nuclear recoils against electron recoils with a suppression factor of 99.7% at energies above 15 keV. This novel method will be applied for background rejection in the CRESST (Cryogenic Rare Event Search with Superconducting Thermometers) experiment looking for dark matter Weakly Interactin…
This paper describes a FADC Readout system developed for the tap water $α/β$ dose monitoring system which is based on EJ444 phoswich scintillation detector and wavelength shifting fiber readout. The Readout system contains dual sampling channels that can supply sampling rate up to 1 GSPS, 14-Bit vertical resolution and adequate effective number of bits (9.7 Bits at 10 MHz), which is optimum for the discrimination of the minimal difference between alpha/beta signals. Moreover, the system is based on a ZYNQ SoC which provides high data throughput speed, low latency and excellent flexibility. As …
In many physical experiments and applications with photomultiplier tubes (PMTs), high-speed pulse digitalization is used extensively. The current pulse from a PMT has a rising time of several to a hundred nanoseconds. The current pulse shape can be used to obtain the pulse energy and discriminate between different particles. The conventional readout system with a PMT uses the discrete or integrated pulse-shaping circuits and analogue-to-digital converter boards which are based on the NIM or VME system to implement the data acquisition function. In many applications, such as security instrument…
We describe here measurements with a new device, the "dichroicon," a Winston-style light concentrator built out of dichroic reflectors, which could allow large-scale neutrino detectors to sort photons by wavelength with small overall light loss. Photon sorting would benefit large-scale water or ice Cherenkov detectors such as Hyper-Kamiokande or ICECUBE by providing a measure of dispersion, which in turn could allow improved position reconstruction and timing. For scintillator detectors like JUNO, upgrades to SNO+ or KamLAND-ZEN, or to water-based liquid scintillator detectors like Theia, dich…
A novel algorithm for the discrimination of neutron and γ-ray with wavelet transform modulus maximum (WTMM) in an organic scintillation has been investigated. Voltage pulses arising from a BC501A organic liquid scintillation detector in a mixed radiation field have been recorded with a fast digital sampling oscilloscope. The performances of most pulse shape discrimination methods in scintillation detection systems using time-domain features of the pulses are affected intensively by noise. However, the WTMM method using frequency-domain features exhibits a strong insensitivity to noise and can …
In a dedicated test setup at the Kamioka Observatory we studied pulse shape discrimination (PSD) in liquid xenon (LXe) for dark matter searches. PSD in LXe was based on the observation that scintillation light from electron events was emitted over a longer period of time than that of nuclear recoil events, and our method used a simple ratio of early to total scintillation light emission in a single scintillation event. Requiring an efficiency of 50% for nuclear recoil retention we reduced the electron background to 7.7\pm1.1(stat)\pm1.2 0.6(sys)\times10-2 at energies between 4.8 and 7.2 keVee …
The CDEX (China Dark matter Experiment) is currently in the stage to upgrade to approximately 10 kg HPGe (High Purity Germanium) detectors (CDEX-10) in CJPL (China Jinping Underground Laboratory) and aims to detect the WIMP (Weakly Interacting Massive Particles). Six analog signals and one inhibited signal are outputted from preamplifier by one HPGe detector (~1 kg) of CDEX-10. The analog signals are conditioned with two 6-us and one 12-us shaping amplifiers, and three timing amplifiers respectively. The inhibited signal is used for trigger. The analog signals from shaping and timing amplifier…
A digital pulse shape discrimination system based on a programmable module NI-5772 has been established and tested with EJ-301 liquid scintillation detector. The module was operated by means of running programs developed in LabVIEW with the sampling frequency up to 1.6GS/s. Standard gamma sources 22Na, 137Cs and 60Co were used to calibrate the EJ-301 liquid scintillation detector, and the gamma response function has been obtained. Digital algorithms for charge comparison method and zero-crossing method have been developed. The experimental results showed that both digital signal processing (DS…
The photoneutron source (PNS, phase 1), an electron linear accelerator (linac)-based pulsed neutron facility that uses the time-of-flight (TOF) technique, was constructed for the acquisition of nuclear data from the thorium molten salt reactor(TMSR) at the Shanghai Institute of Applied Physics (SINAP). The neutron detector signal, with the information on the pulse arrival time, pulse shape, and pulse height, was recorded by using a waveform digitizer (WFD). By using the pulse height and pulse-shape discrimination (PSD) analysis to identify neutrons and $γ$-rays, the neutron TOF spectrum was ob…
Pulse shape discrimination (PSD) is one of the most distinctive features of liquid scintillators. Since the introduction of the scintillation techniques in the field of particle detection, many studies have been carried out to characterize intrinsic properties of the most common liquid scintillator mixtures in this respect. Several application methods and algorithms able to achieve optimum discrimination performances have been developed. However, the vast majority of these studies have been performed on samples of small dimensions. The Counting Test Facility, prototype of the solar neutrino ex…
Pulse shape discrimination using CsI(Tl) scintillators to perform neutral hadron particle identification is explored with emphasis towards application at high energy electron-positron collider experiments. Through the analysis of the pulse shape differences between scintillation pulses from photon and hadronic energy deposits using neutron and proton data collected at TRIUMF, it is shown that the pulse shape variations observed for hadrons can be modelled using a third scintillation component for CsI(Tl), in addition to the standard fast and slow components. Techniques for computing the hadron…
A Discrete Fourier Transform Method (DFTM) for discrimination between the signal of neutrons and gamma rays in organic scintillation detectors is presented. The method is based on the transformation of signals into the frequency domain using the sine and cosine Fourier transforms in combination with the discrete Fourier transform. The method is largely benefited from considerable differences that usually is available between the zero-frequency components of sine and cosine and the norm of the amplitude of the DFT for neutrons and gamma-ray signals. Moreover, working in frequency domain natural…
Linear alkyl benzene (LAB) has in recent years been used as a solvent for PPO in large-scale scintillation detectors, like Daya Bay and SNO+. The combination has several nice properties, including high light yield, good materials compatibility, and excellent pulse shape discrimination. As charged particles move through the LAB+PPO, both Cherenkov and scintillation light are created. Separating Cherenkov from scintillation light would allow a broad range of physics in future large-scale detectors like THEIA, by allowing direction reconstruction with Cherenkov light while retaining the high ligh…
Xenon-doped liquid argon has been proposed as a good alternative to pure liquid argon in scintillation detectors. In this study, we report on the measurement of the time profile of scintillation light emitted from xenon-doped liquid argon with molar concentrations up to 1600 ppm. A compact setup has been developed for this study, with silicon photomultiplier (SiPM) as the photosensor and $^{210}\mathrm{Po}$ and $^{90}\mathrm{Sr}$ as scintillation sources. An effective model based on the de-excitation processes has been developed to describe the data. The results show that xenon-doped liquid ar…
A VME-based experiment system for n-γ discrimination using the charge comparison method was established. A data acquisition program for controlling the programmable modules and processing data online via VME64X bus was developed through the use of LabVIEW. The two-dimensional (2D) scatter plots of the charge in the slow component vs. the total charge of recorded pulses from Am-Be and Cf neutron sources were presented. The 2D scatter plots of the energy vs. the ratio of the charge in the slow component to the total charge of the pulses were presented at the meantime. The quality of n-γ discrimi…
We developed a 12-liter volume neutron detector filled with the liquid scintillator EJ301 that measures neutrons in an underground laboratory where dark matter and neutrino experiments are located. The detector target is a cylindrical volume coated on the inside with reflective paint (95% reflectivity) that significantly increases the detector's light collection. We demonstrate several calibration techniques using point sources and cosmic-ray muons for energies up to 20 MeV for this large liquid scintillation detector. Neutron-gamma separation using pulse shape discrimination with a few MeV ne…
We demonstrate the use of a convolutional neural network to perform neutron-gamma pulse shape discrimination, where the only inputs to the network are the raw digitised SiPM signals from a dual scintillator detector element made of 6LiF:ZnS(Ag) scintillator and PVT plastic. A realistic labelled dataset was created to train the network by exposing the detector to an AmBe source, and a data-driven method utilsing a separate PMT was used to assign labels to the recorded signals. This approach is compared to the charge integration and continuous wavelet transform methods and is found to provide su…
This paper presents the results of the fast neutron irradiation (E$_n$ > 0.5MeV) of an EJ-276 scintillator performed in the MARIA research reactor with fluence up to 5.3$\times$10$^{15}$ particles/cm$^2$. In our work, four samples with size $ϕ$25.4~mm$\times$5~mm were tested. The changes in the light yield, emission and absorption spectrum and neutron/gamma discrimination using PuBe source before and after irradiation are presented. The figure of merit in neutron/gamma discrimination based on the charge integration method for different neutron fluences and different short gate integration time…
We report on the results of the positional-dependent pulse shape discrimination (PSD) parameter observed within a monolithic CLLB scintillation crystal. CLLB, a relatively novel inorganic scintillation crystal, is capable of PSD between gamma rays, neutrons, and alpha particles. In this work, we observed distinguishable differences in the pulse shapes for gamma-ray-induced events. The CLLB crystals used for this experiment are 5 cm (diameter) by 10 cm (length). By using monoenergetic 2.614 MeV photons from a set of thoriated welding rods and performing collimated scans along the length of the …
We present measurements of the scintillation pulse shape in liquid xenon for nuclear recoils (NR) and electronic recoils (ER) at electric fields of 0 to 0.5 kV/cm for energies $<$ 15 keV and $<$ 70 keV electron-equivalent, respectively. The average pulse shapes are well-described by an effective model with two exponential decay components, where both decay times are fit parameters. We find significant broadening of the pulse for ER due to delayed luminescence from the recombination process. In addition to the effective model, we fit a model describing the recombination luminescence for ER at z…
This paper describes the implementation and performance of CsI(Tl) pulse shape discrimination for the Belle II electromagnetic calorimeter, representing the first application of CsI(Tl) pulse shape discrimination for particle identification at an electron-positron collider. The pulse shape characterization algorithms applied by the Belle II calorimeter are described. Control samples of $γ$, $μ^+$, $π^\pm$, $K^\pm$ and $p/\bar{p}$ are used to demonstrate the significant insight into the secondary particle composition of calorimeter clusters that is provided by CsI(Tl) pulse shape discrimination…
We present preliminary results of measurements of the quenching factor for Na recoils in NaI(Tl) at room temperature, made at a dedicated neutron facility at the University of Sheffield. Measurements have been performed with a 2.45 MeV mono-energetic neutron generator in the energy range from 10 keV to 100 keV nuclear recoil energy. A BC501A liquid scintillator detector was used to tag neutrons. Cuts on pulse-shape discrimination from the BC501A liquid scintillator detector and neutron time-of-flight were performed on pulses recorded by a digitizer with a 2 ns sampling time. Measured quenching…
Proportional scintillation counters (PSCs), both single- and dual-phase, can measure the scintillation (S1) and ionization (S2) channels from particle interactions within the detector volume. The signal obtained from these detectors depends first on the physics of the medium (the initial scintillation and ionization), and second how the physics of the detector manipulates the resulting photons and liberated electrons. In this paper we develop a model of the detector physics that incorporates event topology, detector geometry, electric field configuration, purity, optical properties of componen…
In this study, we evaluate and compare the pulse shape discrimination (PSD) performance of multipixel photon counters (MPPCs, also known as silicon photomultiphers - SiPMs) with that of a typical photomultiplier tube (PMT) when testing using CsI(Tl) scintillators. We use the charge comparison method, whereby we discriminate different types of particles by the ratio of charges integrated within two time-gates (the delayed part and the entire digitized waveform). For a satisfactory PSD performance, a setup should generate many photoelectrons (p.e.) and collect their charges efficiently. The PMT …
Pulse-shape discrimination (PSD) is usually achieved using the different fast and slow decay components of inorganic scintillators, such as BaF2, CsI:Tl, etc. However, LaBr3:Ce is considered to not possess different components at room temperature, but has been proved to have the capability of discriminating γ and α events using fast digitizers. In this paper, ionization-density-dependent transport and rate equations are used to quantitatively model the competing processes in a particle track. With one parameter set, the model reproduces the nonproportionality response of electrons or α particl…
A new type of neutron detector, named Stack Structure Solid organic Scintillator (S$^4$), consisting of multi-layer plastic scintillators with capability to suppress low-energy $γ$ rays under high-counting rate has been constructed and tested. To achieve $\it{n}$-$γ$ discrimination, we exploit the difference in the ranges of the secondary charged particles produced by the interactions of neutrons and $γ$ rays in the scintillator material. The thickness of a plastic scintillator layer was determined based on the results of Monte Carlo simulations using the Geant4 toolkit. With layer thicknesses…
A first comparison has been made between the pulse-shape discrimination characteristics of a novel $^{4}$He-based pressurized scintillation detector and a NE-213 liquid-scintillator reference detector using an Am/Be mixed-field neutron and gamma-ray source and a high-resolution scintillation-pulse digitizer. In particular, the capabilities of the two fast neutron detectors to discriminate between neutrons and gamma-rays were investigated. The NE-213 liquid-scintillator reference cell produced a wide range of scintillation-light yields in response to the gamma-ray field of the source. In stark …
Liquid noble based particle detectors often use the organic wavelength shifter 1,1,4,4- tetraphenyl-1,3-butadiene (TPB) which shifts UV scintillation light to the visible regime, facilitating its detection, but which also can scintillate on its own. Dark matter searches based on this type of detector commonly rely on pulse-shape discrimination (PSD) for background mitigation. Alpha-induced scintillation therefore represents a possible background source in dark matter searches. The timing characteristics of this scintillation determine whether this background can be mitigated through PSD. We ha…
Weakly Interacting Massive Particles (WIMPs) are a leading candidate for dark matter and are expected to produce nuclear recoil (NR) events within liquid xenon time-projection chambers. We present a measurement of the scintillation timing characteristics of liquid xenon in the LUX dark matter detector and develop a pulse shape discriminant to be used for particle identification. To accurately measure the timing characteristics, we develop a template-fitting method to reconstruct the detection times of photons. Analyzing calibration data collected during the 2013-16 LUX WIMP search, we provide …
Scintillation produced in liquid xenon by alpha particles and gamma rays has been studied as a function of applied electric field. For back scattered gamma rays with energy of about 200 keV, the number of scintillation photons was found to decrease by 64+/-2% with increasing field strength. Consequently, the pulse shape discrimination power between alpha particles and gamma rays is found to reduce with increasing field, but remaining non-zero at higher fields.
The scintillation mechanism in NaI:Tl crystals produces different pulse shapes that are dependent on the incoming particle type. The time distribution of scintillation light from nuclear recoil events decays faster than for electron recoil events and this difference can be categorised using various Pulse Shape Discrimination (PSD) techniques. In this study, we measured nuclear and electron recoils in a NaI:Tl crystal, with electron equivalent energies between 2 and 40 keV. We report on a new PSD approach, based on an event-type likelihood; this outperforms the charge-weighted mean-time, which …
Liquid-argon scintillation detectors are used in fundamental physics experiments and are being considered for security applications. Previous studies have suggested that the addition of small amounts of xenon dopant improves performance in light or signal yield, energy resolution, and particle discrimination. In this study, we investigate the detector response for xenon dopant concentrations from 9 +/- 5 ppm to 1100 +/- 500 ppm xenon (by weight) in 6 steps. The 3.14-liter detector uses tetraphenyl butadiene (TPB) wavelength shifter with dual photomultiplier tubes and is operated in single-phas…
We measured the decay time of the scintillation pulses produced by electron and nuclear recoils in CaF2(Eu) by a new fitting method. In the recoil energy region 5-30 keVee, we found differences of the decay time between electron and nuclear recoil events. In the recoil energy region above 20 keVee, we found that the decay time is independent of the recoil energy.
Various pulse shape discrimination methods have been used to solve the neutron-gamma discrimination problem. But most of them are limited to off-line calculation due to the computation amount and FPGA performance. In order to realize real time discriminating neutron and gamma, a new algorithm based on the traditional pulse shape discrimination methods was proposed in this paper. The new algorithm takes into account the physical properties of the pulse signal, which greatly reduces the computation and dead time without losing the precision, and can work on FPGA directly. It has a good performan…
Pulse shape discrimination with pure CsI scintillators is investigated as a method for separating energy deposits by energetic neutrons and photons at particle physics experiments. Using neutron data collected near the European XFEL XS1 beam window the pulse shape discrimination capabilities of pure CsI are studied and compared to CsI(Tl) using near-identical detector setups, which were operated in parallel. The inelastic interactions of 100MeV neutrons are observed to produce a slower scintillation emission in pure CsI relative to energy deposits from cosmic muons. By employing a charge-ratio…
Digital pulse shape discrimination has become readily available to distinguish nuclear recoil and electronic recoil events in scintillation detectors. We evaluate digital implementations of pulse shape discrimination algorithms discussed in the literature, namely the Charge Comparison Method, Pulse-Gradient Analysis, Fourier Series and Standard Event fitting. In addition, we present a novel algorithm based on a Laplace Transform. Instead of reducing the performance of these algorithms to a single Figure of Merit, we evaluate them as function of recoil energy. Specifically, using commercial EJ-…
In this manuscript we report on the scintillation properties and pulse shape discrimination (PSD) performance of new organic glass scintillator. Two cylindrical samples with dimensions of 2x2 inches were tested. Additionally, this two samples were used in stack configuration in order to measure the PSD characteristics of a sample with a size of 2x4 inches. The study covers the measurements of neutron/gamma discrimination capability, emission spectra, photoelectron yield and analysis of the light pulse shapes originating from events related to gamma-rays and fast neutrons. The results were comp…
We describe a technique, applicable to liquid-argon-based dark matter detectors, allowing for discrimination of alpha-decays in detector regions with incomplete light collection from nuclear-recoil-like events. Nuclear recoils and alpha events preferentially excite the liquid argon (LAr) singlet state, which has a decay time of ~6 ns. The wavelength-shifter TPB, which is typically applied to the inside of the active detector volume to make the LAr scintillation photons visible, has a short re-emission time that preserves the LAr scintillation timing. We developed a wavelength-shifting polymeri…
Decays of radioisotopes on inner detector surfaces can pose a major background concern for the direct detection of dark matter. While these backgrounds are conventionally mitigated with position cuts, these cuts reduce the exposure of the detector by decreasing the sensitive mass, and uncertainty in position determination may make it impossible to adequately remove such events in certain detectors. In this paper, we provide a new technique for substantially reducing these surface backgrounds in liquid argon (LAr) detectors, independent of position cuts. These detectors typically use a coating …