体积超分辨率超声(SR-US)成像的三维运动校正

3-D Motion Correction for Volumetric Super-Resolution Ultrasound (SR-US) Imaging

S. Harput, K. Christensen-Jeffries, J. Brown, J. Zhu, G. Zhang, R. J. Eckersley, C. Dunsby, M-X. Tang

摘要:

在所有医学成像模态中,图像采集期间的运动会导致图像退化。这在2-D超声成像中特别相关,因为平面外运动只能补偿小于换能器的仰角波束宽度的运动。基于定位的超分辨率成像由于需要大量的采集来形成单个超分辨率帧而产生甚至更具挑战性的运动校正任务。在本研究中,我们提出一种延伸的两阶段运动校正方法来进行三维运动校正。使用手持式探头对高容积率超声采集进行运动估计。所提出的方法的能力进行了演示与3-D微血管流动模拟,以补偿手持探头运动。结果表明,两阶段运动校正方法将平均定位误差从136微米降低到18微米。

英文摘要:

Motion during image acquisition can cause image degradation in all medical imaging modalities. This is particularly relevant in 2-D ultrasound imaging, since out-of-plane motion can only be compensated for movements smaller than elevational beamwidth of the transducer. Localization based super-resolution imaging creates even a more challenging motion correction task due to the requirement of a high number of acquisitions to form a single super-resolved frame. In this study, an extension of two-stage motion correction method is proposed for 3-D motion correction. Motion estimation was performed on high volumetric rate ultrasound acquisitions with a handheld probe. The capability of the proposed method was demonstrated with a 3-D microvascular flow simulation to compensate for handheld probe motion. Results showed that two-stage motion correction method reduced the average localization error from 136 to 18 micrometres.