Bimodal intravascular volumetric imaging combining OCT and MPI
摘要:
血管内光学相干断层扫描(IVOCT)是一种基于导管的成像模式,可对血管进行高分辨率成像。它基于轴向空间分辨率在5至10 μm范围内的A扫描的快速顺序采集,即,比传统方法如血管内超声或计算机断层成像血管造影术高一个数量级。但是,仅通过IVOCT测量无法获得导管在患者坐标中的位置和方向。因此,需要建立导管的姿态以正确地重建三维血管形状。磁粒子成像(MPI)是一种三维断层扫描,基于示踪剂和无辐射的图像模式,提供了高的时间分辨率和无限的穿透深度。容积MPI图像是血管造影图像,因此适合作为辅助模态补充IVOCT。我们研究了同时双峰IVOCT MPI成像,目的是基于3D MPI数据估计IVOCT回撤路径。我们提出了一种研究和评估不同形状血管模型的同时IVOCT和MPI图像采集的设置。首先,分析了MPI示踪剂浓度对IVOCT所需光学特性的影响。第二,使用允许同时成像的浓度,依次且同时采集IVOCT和MPI图像数据。第三,根据MPI图像体积建立管腔中心线,并用于估计IVOCT图像重建的导管回撤轨迹。将图像体积与体模的已知形状进行比较。我们能够确定适当的MPI示踪剂浓度为2.5 mmol/L,对IVOCT信号的影响可忽略不计。从MPI估计的回撤轨迹与体模的中心线一致。(...)
英文摘要:
Intravascular optical coherence tomography (IVOCT) is a catheter based image modality allowing for high resolution imaging of vessels. It is based on a fast sequential acquisition of A-scans with an axial spatial resolution in the range of 5 to 10 μm, i.e., one order of magnitude higher than in conventional methods like intravascular ultrasound or computed tomography angiography. However, position and orientation of the catheter in patient coordinates cannot be obtained from the IVOCT measurements alone. Hence, the pose of the catheter needs to be established to correctly reconstruct the three-dimensional vessel shape. Magnetic particle imaging (MPI) is a three-dimensional tomographic, tracer-based and radiation-free image modality providing high temporal resolution with unlimited penetration depth. Volumetric MPI images are angiographic and hence suitable to complement IVOCT as a co-modality. We study simultaneous bimodal IVOCT MPI imaging with the goal of estimating the IVOCT pullback path based on the 3D MPI data. We present a setup to study and evaluate simultaneous IVOCT and MPI image acquisition of differently shaped vessel phantoms. First, the infuence of the MPI tracer concentration on the optical properties required for IVOCT is analyzed. Second, using a concentration allowing for simultaneous imaging, IVOCT and MPI image data is acquired sequentially and simultaneously. Third, the luminal centerline is established from the MPI image volumes and used to estimate the catheter pullback trajectory for IVOCT image reconstruction. The image volumes are compared to the known shape of the phantoms. We were able to identify a suitable MPI tracer concentration of 2.5 mmol/L with negligible influence on the IVOCT signal. The pullback trajectory estimated from MPI agrees well with the centerline of the phantoms. (...)
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。