生物组织中单能和双能计算机断层成像阻止功率预测的实验验证

Experimental verification of stopping-power prediction from single- and dual-energy computed tomography in biological tissues

Christian Möhler, Tom Russ, Patrick Wohlfahrt, Alina Elter, Armin Runz, Christian Richter, Steffen Greilich

摘要:

介绍了一种连续测量离子和X射线在生物组织或其他材料中吸收的实验装置。通过使用3D打印样品容器的这种设置,可以测量材料的参考停止功率比(SPR),其不确定性低于0.1%。共制备了65份猪和牛组织样本用于测量,包括13种组织类型的各5份样本,代表总体重的约80%(三种不同的肌肉和脂肪组织:肝脏、肾脏、大脑、心脏、血液、肺和骨骼)。使用单能CT(SECT)的标准化学计量校准以及最先进的双能CT(DECT)方法,预测所有组织的SPR,然后与测量的参考值进行比较。采用SECT方法预测所有组织的SPR,平均误差为(-0.84 $\pm$ 0.12)%,平均绝对误差为(1.27 $\pm$ 0.12)%。相比之下,基于DECT的SPR预测与测量的参考值总体一致,平均误差为(-0.02 $\pm$ 0.15)%,平均绝对误差为(0.10 $\pm$ 0.15)%。因此,在本研究中,DECT降低范围不确定性的潜力可以在生物组织中得到证实。

英文摘要:

An experimental setup for consecutive measurement of ion and x-ray absorption in tissue or other materials is introduced. With this setup using a 3D-printed sample container, the reference stopping-power ratio (SPR) of materials can be measured with an uncertainty of below 0.1%. A total of 65 porcine and bovine tissue samples were prepared for measurement, comprising five samples each of 13 tissue types representing about 80% of the total body mass (three different muscle and fatty tissues, liver, kidney, brain, heart, blood, lung and bone). Using a standard stoichiometric calibration for single-energy CT (SECT) as well as a state-of-the-art dual-energy CT (DECT) approach, SPR was predicted for all tissues and then compared to the measured reference. With the SECT approach, the SPRs of all tissues were predicted with a mean error of (-0.84 $\pm$ 0.12)% and a mean absolute error of (1.27 $\pm$ 0.12)%. In contrast, the DECT-based SPR predictions were overall consistent with the measured reference with a mean error of (-0.02 $\pm$ 0.15)% and a mean absolute error of (0.10 $\pm$ 0.15)%. Thus, in this study, the potential of DECT to decrease range uncertainty could be confirmed in biological tissue.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑