使用X射线相位衬度的CT剂量降低因子(单位:千)

CT dose reduction factors in the thousands using X-ray phase contrast

Marcus J. Kitchen, Genevieve A. Buckley, Timur E. Gureyev, Megan J. Wallace, Nico Andres-Thio, Kentaro Uesugi, Naoto Yagi, Stuart B. Hooper, .

摘要:

相衬X射线成像与常规射线照片相比,可以将弱吸收物体(例如软组织)的可见性提高一个数量级或更多。先前,已经表明,在相同的辐射剂量下,将相位恢复与计算机断层成像(CT)相结合可以将信噪比(SNR)增加到传统CT的两个数量级,而不会损失图像质量。我们的实验表明,随着辐射剂量的降低,SNR的相对改善增加。我们发现,这种增强可以用大于SNR增益平方的剂量减少来交换。在将剂量减少300倍时,相位恢复的SNR仍然比吸收对比度数据大几乎10倍。这揭示了在不损失图像质量的情况下降低数万倍剂量的潜力,这将对医疗和工业成像应用产生深远的影响。

英文摘要:

Phase-contrast X-ray imaging can improve the visibility of weakly absorbing objects (e.g. soft tissues) by an order of magnitude or more compared to conventional radiographs. Previously, it has been shown that combining phase retrieval with computed tomography (CT) can increase the signal-to-noise ratio (SNR) by up to two orders of magnitude over conventional CT at the same radiation dose, without loss of image quality. Our experiments reveal that as radiation dose decreases, the relative improvement in SNR increases. We discovered this enhancement can be traded for a reduction in dose greater than the square of the gain in SNR. Upon reducing the dose 300 fold, the phase-retrieved SNR was still almost 10 times larger than the absorption contrast data. This reveals the potential for dose reduction factors in the tens of thousands without loss in image quality, which would have a profound impact on medical and industrial imaging applications.