用于磁共振粒子成像阴性对比灌注成像的生理盐水团注
Saline bolus for negative contrast perfusion imaging in magnetic particle imaging
AI总结:
本研究针对MPI灌注成像需反复注射示踪剂增加体内铁剂量的问题,提出生理盐水阴性对比团注方法,体模实验验证其灌注参数与阳性团注一致性好、偏差低,可延长监测时长并降低单次成像铁剂量。
摘要:
研究目的:磁共振粒子成像(Magnetic Particle Imaging, MPI)能够对磁性纳米颗粒的空间分布进行高时间分辨率测量,因此非常适用于灌注成像——这是医学诊断中的重要工具。MPI中的灌注成像通常需要新的示踪剂团注来捕捉关键信号动态,本研究提出一种将成像序列与额外示踪剂注射解耦的方法,且不会因每次成像进一步增加体内给药的铁剂量。\n研究方法:不含任何颗粒的生理盐水团注(阴性对比)在流经过程中会降低长循环示踪剂的稳态浓度,这种测量浓度的下降提供了所需的对比动态,因此长循环示踪剂的存在是获得阴性对比的前提。作为一种基于示踪剂的定量成像方法,在任何含有纳米颗粒的位置,信号与示踪剂浓度呈线性关系,而在无固有信号的周围组织中信号为零。示踪剂注射后,浓度随时间的变化(阳性对比)可用于计算动态诊断参数,如血管和器官中的灌注参数,因此每次获取的灌注图像都需要新的示踪剂团注,且其铁剂量需足够大以在背景之上可见。\n主要结果:基于所提出的阴性团注的时间响应计算灌注参数,并与阳性团注的结果进行比较。体模实验结果表明,阳性和阴性团注的归一化信号具有一致性,计算得到的灌注图偏差较低。\n研究意义:本方法为增加总监测时长提供了可能,同时最大限度降低了每次获取图像所需的铁剂量。
英文摘要:
Objective. Magnetic Particle Imaging (MPI) is capable of high temporal resolution measurements of the spatial distribution of magnetic nanoparticles and therefore well suited for perfusion imaging, which is an important tool in medical diagnosis. Perfusion imaging in MPI usually requires a fresh bolus of tracer material to capture the key signal dynamics. Here, we propose a method to decouple the imaging sequence from the injection of additional tracer material, without further increasing the administered iron dose in the body with each image. Approach. A bolus of physiological saline solution without any particles (negative contrast) diminishes the steady-state concentration of a long-circulating tracer during passage. This depression in the measured concentration contributes to the required contrast dynamics. The presence of a long-circulating tracer is therefore a prerequisite to obtain the negative contrast. As a quantitative tracer based imaging method, the signal is linear in the tracer concentration for any location that contains nanoparticles and zero in the surrounding tissue which does not provide any intrinsic signal. After tracer injection, the concentration over time (positive contrast) can be utilized to calculate dynamic diagnostic parameters like perfusion parameters in vessels and organs. Every acquired perfusion image thus requires a new bolus of tracer with a sufficiently large iron dose to be visible above the background. Main results. Perfusion parameters are calculated based on the time response of the proposed negative bolus and compared to a positive bolus. Results from phantom experiments show that normalized signals from positive and negative boli are concurrent and deviations of calculated perfusion maps are low. Significance. Our method opens up the possibility to increase the total monitoring time, while minimizing the iron dose per acquired image.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。