位置灵敏硅探测器能量刻度的新方法

A new method of energy calibration of position-sensitive silicon detector

Mingdao Sun, Tianheng Huang, Zhong Liu, Bing Ding, Huabin Yang, Zhiyuan Zhang, Jianguo Wang, Long Ma, Lin Yu, Yongsheng Wang, Zaiguo Gan, Xiaohong Zhou

摘要:

提出了一种改进的位置灵敏硅探测器能量刻度方法。引入了一种描述位置与能量关系的新函数,取代了传统方法中使用的抛物线函数,并获得了更好的能量分辨率。对8.088 MeV的213 Rn α衰变,用该方法标定的能量分辨率约为87 keV(FWHM),优于传统方法的104 keV(FWHM)。此外,新方法还可以尝试不同的函数,使得对不同性能的探测器进行标定成为可能。

英文摘要:

An improved method of energy calibration of position-sensitive silicon detector is presented. Instead of the parabolic function used in traditional method, a new function describing the relation of position and energy is introduced and achieves better energy resolution. For the 8.088 MeV alpha decay of 213Rn calibrated by this new method, the energy resolution is determined to be about 87 keV (FWHM), which is better than the result of the traditional method, 104 keV (FWHM). In addition, different functions can be tried in the new method, which makes the calibration of various detectors with different performances possible.