用于TEAM显微镜的CMOS有源像素传感器的表征

Characterisation of a CMOS Active Pixel Sensor for use in the TEAM Microscope

Marco Battaglia, Devis Contarato, Peter Denes, Dionisio Doering, Thomas Duden, Brad Krieger, Piero Giubilato, Dario Gnani, Velimir Radmilovic

摘要:

作为TEAM项目的一部分,已经开发了一个100万像素和一个400万像素的单片CMOS有源像素传感器,像素为9.5x9.5微米^2,用于透射电子显微镜的直接成像。我们提出的设计和检测器的充分表征。在电子显微镜中以各种感兴趣的能量用电子束收集的数据用于确定检测器响应。将数据与模拟预测进行比较。用80 keV和300 keV电子测量的线扩展函数分别为(12.1+/-0.7)微米和(7.4+/-0.6)微米,与我们的模拟非常一致。我们测量的检测量子效率为0.78+/-0.04在80 keV和0.74+/-0.03在300 keV。采用基于单电子重构的新成像技术,80 keV和300 keV电子的线扩展函数分别变为(6.7 ± 0.3)微米和(2.4 ± 0.2)微米。像素的辐射耐受性已经过测试,高达5 Mrad,检测器仍然是功能性的,动态范围减少了~ 30%,对应于全阱深度从~39减少到~27初级300 keV电子,由于漏电流增加,但相同的线扩展函数性能。

英文摘要:

A 1M- and a 4M-pixel monolithic CMOS active pixel sensor with 9.5x9.5 micron^2 pixels have been developed for direct imaging in transmission electron microscopy as part of the TEAM project. We present the design and a full characterisation of the detector. Data collected with electron beams at various energies of interest in electron microscopy are used to determine the detector response. Data are compared to predictions of simulation. The line spread function measured with 80 keV and 300 keV electrons is (12.1+/-0.7) micron and (7.4+/-0.6) micron, respectively, in good agreement with our simulation. We measure the detection quantum efficiency to be 0.78+/-0.04 at 80 keV and 0.74+/-0.03 at 300 keV. Using a new imaging technique, based on single electron reconstruction, the line spread function for 80 keV and 300 keV electrons becomes (6.7+/-0.3) micron and (2.4+/-0.2) micron, respectively. The radiation tolerance of the pixels has been tested up to 5 Mrad and the detector is still functional with a decrease of dynamic range by ~30%, corresponding to a reduction in full-well depth from ~39 to ~27 primary 300 keV electrons, due to leakage current increase, but identical line spread function performance.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑