CENNS-10液氩探测器标定用$^{83\mathrm{m}}$ Kr源的研制

Development of a $^{83\mathrm{m}}$Kr source for the calibration of the CENNS-10 Liquid Argon Detector

COHERENT Collaboration, D. Akimov, P. An, C. Awe, P. S. Barbeau, B. Becker, V. Belov, I. Bernardi, M. A. Blackston, L. Blokland, A. Bolozdynya, B. Cabrera-Palme… 展开作者

COHERENT Collaboration, D. Akimov, P. An, C. Awe, P. S. Barbeau, B. Becker, V. Belov, I. Bernardi, M. A. Blackston, L. Blokland, A. Bolozdynya, B. Cabrera-Palmer, N. Chen, D. Chernyak, E. Conley, J. Daughhetee, M. del Valle Coello, J. A. Detwiler, M. R. Durand, Y. Efremenko, S. R. Elliott, L. Fabris, M. Febbraro, W. Fox, A. Galindo-Uribarri, A. Gallo Rosso, M. P. Green, K. S. Hansen, M. R. Heath, S. Hedges, M. Hughes, T. Johnson, A. Khromov, A. Konovalov, E. Kozlova, A. Kumpan, L. Li, J. T. Librande, J. M. Link, J. Liu, K. Mann, D. M. Markoff, O. McGoldrick, P. E. Mueller, J. Newby, D. S. Parno, S. Pentilla, D. Pershey, D. Radford, R. Rapp, H. Ray, J. Raybern, O. Razuvaeva, D. Reyna, G. C. Rich, D. Rudik, J. Runge, D. J. Salvat, K. Scholberg, A. Shakirov, G. Simakov, W. M. Snow, V. Sosnovtsev, B. Suh, R. Tayloe, K. Tellez-Giron-Flores, R. T. Thornton, I. Tolstukhin, J. Vanderwerp, R. L. Varner, R. Venkataraman, C. J. Virtue, G. Visser, C. Wiseman, T. Wongjirad, J. Yang, Y. -R. Yen, J. Yoo, C. -H. Yu, J. Zettlemoyer

摘要:

我们报告的准备和校准测量与$^{83\mathrm{m}}$氪源的CENNS-10液氩探测器。在$^{83}$ Rb源的衰变中产生的$^{83\mathrm{m}}$ Kr原子通过注入Ar循环回路被引入到检测器中。然后观察到探测器体积中$^{83\mathrm{m}}$ Kr衰变中9.4 keV和32.1 keV转换电子产生的闪烁光。该校准源可以表征CENNS-10探测器的低能响应,并适用于其他低能阈值探测器。在总$^{83\mathrm{m}}$ Kr衰变能量为41.5 keV时,探测器的能量分辨率测量为9 $\%$。我们进行了分析,以分别使用9.4 keV和32.1 keV的两个转换电子来校准检测器

英文摘要:

We report on the preparation of and calibration measurements with a $^{83\mathrm{m}}$Kr source for the CENNS-10 liquid argon detector. $^{83\mathrm{m}}$Kr atoms generated in the decay of a $^{83}$Rb source were introduced into the detector via injection into the Ar circulation loop. Scintillation light arising from the 9.4 keV and 32.1 keV conversion electrons in the decay of $^{83\mathrm{m}}$Kr in the detector volume were then observed. This calibration source allows the characterization of the low-energy response of the CENNS-10 detector and is applicable to other low-energy-threshold detectors. The energy resolution of the detector was measured to be 9$\%$ at the total $^{83\mathrm{m}}$Kr decay energy of 41.5 keV. We performed an analysis to separately calibrate the detector using the two conversion electrons at 9.4 keV and 32.1 keV

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑