用于两相氩探测器的~(109)Cd gamma-ray源的特性

Characterization of a 109Cd gamma-ray source for the two-phase argon detector

A. Bondar, A. Buzulutskov, A. Dolgov, A. Legkodymov, V. Nosov, V. Oleynikov, V. Porosev, E. Shemyakina, A. Sokolov

摘要:

目前,我们实验室正在研制一台用于暗物质搜寻和低能中微子实验的两相氩探测器。为了校准其能量标度,使用了$^{109}$ Cd伽马射线源。在这项工作中,详细的发射光谱的$^{109}$镉源测量使用YAP:Ce闪烁体和高纯锗(HPGe)探测器。结果表明,$^{109}$镉源,配备了一个W衬底和一个铜过滤器,可以提供一套完整的γ射线线,从8到90 keV,为两相探测器的能量校准。这些测量使我们能够成功地量化的振幅谱的形状时,观察到的两相氩探测器照射的$^{109}$镉源。

英文摘要:

At present, a two-phase argon detector is being developed in our laboratory for dark matter search and low-energy neutrino experiments. To calibrate its energy scale a $^{109}$Cd gamma-ray source was used. In this work a detailed emission spectrum of the $^{109}$Cd source was measured using YAP:Ce scintillator and high-purity germanium (HPGe) detectors. It is shown that the $^{109}$Cd source, equipped with a W substrate and a Cu filter, can provide a complete set of gamma-ray lines, ranging from 8 to 90 keV, for the energy calibration of two-phase detectors. These measurements allowed us to successfully quantify the shape of the amplitude spectra observed in the two-phase argon detector when irradiated with the $^{109}$Cd source.