基于超声的便携式体外表征设备
Portable Devices for In-Vitro Characterization Based on Ultrasound
摘要:
我们提出了一个便携式墨盒的芯片实验室平台。主要传感机制使用高频声波。研究了两种不同换能器用于生物测量的可行性。我们使用氧化锌(ZnO)为基础的压电换能器的血液凝固时间测量。该方法是基于从充满血液的微流体通道的声反射的振幅的测量。微流体通道的体积可以低至1微升。设计了一种低成本的一次性玻璃盒来进行测量。墨盒由两个玻璃基板和一个双面胶带层组成,其中定义了通道。其中一个玻璃基板具有基于ZnO的微加工换能器,该换能器在盒的外表面上以400 MHz工作,其中还提供了电连接。换能器与通道对准以在填充通道的流体中产生声脉冲。来自通道顶部的反射通过液体传播,因此它们的振幅和相位受流体性质的影响。在实验中,使用全血而不进行任何样品制备。血液的粘度在凝固期间改变,因此通过监测来自通道顶部的反射的幅度,可以测量凝固时间。该方法只需要1微升的血液,已经用钙和活化部分凝血活酶(aPTT)试剂进行了测试。所提出的方法有可能被用于低成本的便携式凝血时间测量系统的患者自我测试。第二换能器由CMUT(电容式微机械超声换能器)制成。我们介绍这种技术浸入液体作为生物传感器的粘度测量。
英文摘要:
We propose a portable cartridge based lab-on-a-chip platform. The main sensing mechanism uses high frequency acoustic waves. Feasibility of two different transducers has been investigated for biological measurements. We used zincoxide (ZnO) based piezoelectric transducers for blood coagulation time measurements. The method is based on the measurement of amplitude of acoustic reflections from a microfluidic channel which is filled with blood. The volume of the microfluidic channel could be as low as 1 micro-litre. A low cost disposable cartridge made of glass has been designed to perform the measurements. The cartridge is composed of two glass substrates and a double sided tape layer where the channel is defined. One of the glass substrates has a micromachined ZnO based transducer that operates at 400 MHz on the outer surface of the cartridge where electrical connections are also provided. The transducer is aligned with the channel to generate acoustic pulses in the fluid that is filling the channel. The reflections coming from the top of the channel propagate through the liquid therefore their amplitude and phase are affected by the fluid properties. In the experiments whole blood was used without any sample preparation. The viscosity of the blood changes during coagulation therefore by monitoring the amplitude of the reflection coming from the top of the channel one can measure the coagulation time. The method which requires only 1 micro-litre of blood has been tested with calcium and activated partial thromboplastin (aPTT) reagents. The proposed method has a potential to be used in a low cost portable coagulation time measurement system for patient self-testing. The second transducer was made of CMUTs (Capacitive Micromachined Ultrasonic Transducers). We introduce this technology for immersion in liquids as a biological sensor for viscosity measurements.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。