arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1706.03507physics.ins-detphysics.med-ph
收藏

磁共振成像系统超低噪声放大器设计

Ultra-Low Noise Amplifier Design for Magnetic Resonance Imaging systems

Mustafa A Kabel

摘要:

本文介绍了一种超低噪声放大器的设计和开发,它可能会增加现有磁共振成像(MRI)系统的灵敏度。低噪声放大器的设计制作和特点,包括匹配和输入高功率保护电路。在这里采取的LNA的SNR相比,在室温下操作的估计改善。共源共栅放大器的拓扑结构被选择来研究高性能低噪声放大器的设计和制造。利用频谱分析仪和矢量网络分析仪对共源共栅低噪声放大器的PCB版图进行了测试。在室温下对共源共栅低噪声放大器的PCB版图进行了测试,其工作频率为32 MHz,源阻抗为50 Ω时噪声系数为0.45 dB,增益为11.6 dB,输出回波损耗为21.1 dB,输入回波损耗为0.12 dB,在6 GHz频段内无条件稳定。共源共栅低噪声放大器的信噪比得到了改善,达到了研究的目的。

英文摘要:

This paper demonstrates designing and developing of an Ultra-Low Noise Amplifier which should potentially increase the sensitivity of the existing Magnetic Resonance Imaging (MRI) systems. The Design of the LNA is fabricated and characterized including matching and input high power protection circuits. The estimate improvement of SNR of the LNA in comparison to room temperature operation is taken here. The Cascode amplifier topology is chosen to be investigated for high performance Low Noise amplifier design and for the fabrication. The fabricated PCB layout of the Cascode LNA is tested by using measurement instruments Spectrum Analyser and Vector Network analyzer. The measurements of fabricated PCB layout of the Cascode LNA at room temperature had the following performance, the operation frequency is 32 MHz, the noise figure is 0.45 dB at source impedance 50 Ω, the gain is 11.6 dB, the output return loss is 21.1 dB, and the input return loss 0.12 dB and it is unconditionally stable for up to 6 GHz band. The goal of the research is achieved where the Cascode LNA had improvement of SNR.