用于X射线和真空紫外光探测的大面积雪崩光电二极管特性研究

Characterization of large area avalanche photodiodes in X-ray and VUV-light detection

L. M. P. Fernandes, F. D. Amaro, A. Antognini, J. M. R. Cardoso, C. A. N. Conde, O. Huot, P. E. Knowles, F. Kottmann, J. A. M. Lopes, L. Ludhova, C. M. B. Monte… 展开作者

L. M. P. Fernandes, F. D. Amaro, A. Antognini, J. M. R. Cardoso, C. A. N. Conde, O. Huot, P. E. Knowles, F. Kottmann, J. A. M. Lopes, L. Ludhova, C. M. B. Monteiro, F. Mulhauser, R. Pohl, J. M. F. dos Santos, L. A. Schaller, D. Taqqu, J. F. C. A. Veloso

摘要:

本文综述了大面积雪崩光电二极管(LAAPD)在X射线和真空紫外(VUV)光探测中的应用。由Advanced Photonix Inc.制造的四种不同LAAPD的操作特性,在室温下研究了有效面积为80和200 mm^2的X射线探测器。最好的能量分辨率被认为是在10-18%的范围内为5.9 keV的X射线。LAAPD结构紧凑,操作简单,计数率高(高达约10^5/s),被证明是有用的几个应用,如低能量X射线检测,在那里他们可以达到更好的性能比正比计数器。由于X射线在光测量中用作参考,因此研究了5.9 keV X射线与光脉冲之间的增益非线性。X射线和VUV光之间的增益比随着增益而减小,对于在氩(~128 nm)和氙(~172 nm)中产生的VUV光,对于增益200,分别达到10%和6%的变化,而对于可见光(~635 nm),变化低于1%。研究了温度对LAAPD性能的影响。对于最高增益,观察到每摄氏度约-5%的相对增益变化。多余的噪声因子被发现是独立的温度,从50到300的增益为1.8和2.3之间。主要由于暗电流,能量分辨率对于降低温度更好。LAAPD在高达5 T的强磁场下进行了测试,当用于X射线和可见光检测时不敏感,而对于VUV光,在5 T下观察到显著的幅度减小。

英文摘要:

The present manuscript summarizes novel studies on the application of large area avalanche photodiodes (LAAPDs) to the detection of X-rays and vacuum ultraviolet (VUV) light. The operational characteristics of four different LAAPDs manufactured by Advanced Photonix Inc., with active areas of 80 and 200 mm^2 were investigated for X-ray detection at room temperature. The best energy resolution was found to be in the 10-18% range for 5.9 keV X-rays. The LAAPD, being compact, simple to operate and with high counting rate capability (up to about 10^5/s), proved to be useful in several applications, such as low-energy X-ray detection, where they can reach better performance than proportional counters. Since X-rays are used as reference in light measurements, the gain non-linearity between 5.9 keV X-rays and light pulses was investigated. The gain ratio between X-rays and VUV light decreases with gain, reaching 10 and 6% variations for VUV light produced in argon (~128 nm) and xenon (~172 nm), respectively, for a gain 200, while for visible light (~635 nm) the variation is lower than 1%. The effect of temperature on the LAAPD performance was investigated. Relative gain variations of about -5% per Celsius degree were observed for the highest gains. The excess noise factor was found to be independent on temperature, being between 1.8 and 2.3 for gains from 50 to 300. The energy resolution is better for decreasing temperatures due mainly to the dark current. LAAPDs were tested under intense magnetic fields up to 5 T, being insensitive when used in X-ray and visible-light detection, while for VUV light a significant amplitude reduction was observed at 5 T.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑