Performance of a Chamber for Studying the Liquid Xenon Response to Nuclear Recoils
V. Chepel, F. Neves, V. Solovov, A. Pereira, M. I. Lopes, J. Pinto da Cunha, P. Mendes, A. Lindote, C. P. Silva, R. Ferreira Marques, A. J. P. L. Policarpo
备注:
6 pages, 6 figures; Presented at IEEE Nuclear Science Symposium and Medical Imaging Conference, Rome 2004; Submitted to IEEE Transactions on Nuclear Science
The design and performance of a 1.2 liter liquid xenon chamber equipped with 7 two-inch photomultiplier tubes, with the purpose of studying the scintillation response of xenon to gamma-rays and neutrons, is described. Measurements with gamma-rays indicate a high VUV light collection efficiency resulting in ~5.5 photoelectrons per 1 keV of deposited energy. The energy resolution (FWHM) is 18% and 22%, for 122 keV and 511 keV gamma-rays, respectively. An algorithm for the reconstruction of the scintillation coordinates in (x,y) plane was developed and tested. The position resolution is estimated to be 6.9 mm (sigma) for 122 keV gamma-rays.
The scintillation efficiency of liquid xenon for nuclear recoils has been measured to be nearly constant in the recoil energy range from 140 keV down to 5 keV. The average ratio of the efficiency for recoils to that for gamma-rays is found to be 0.19+-0.02.
Liquid xenon is an important detection medium in direct dark matter experiments, which search for low-energy nuclear recoils produced by the elastic scattering of WIMPs with quarks. The two existing measurements of the relative scintillation efficiency of nuclear recoils below 20 keV lead to inconsistent extrapolations at lower energies. This results in a different energy scale and thus sensitivity reach of liquid xenon dark matter detectors. We report a new measurement of the relative scintillation efficiency below 10 keV performed with a liquid xenon scintillation detector, optimized for max…
Liquid xenon-based direct detection dark matter experiments have recently expanded their searches to include high-energy nuclear recoil events as motivated by effective field theory dark matter and inelastic dark matter interaction models, but few xenon recoil calibrations above 100 keV are currently available. In this work, we measured the scintillation and ionization yields of xenon recoils up to 426 keV. The experiment uses 14.1 MeV neutrons to scatter off xenon in a compact liquid xenon time projection chamber and produce quasi-monoenergetic xenon recoils between 39 keV and 426 keV. We rep…
The feasibility of an ultra-low energy nuclear-recoil measurement in liquid xenon using neutron capture is investigated for a small (sub-kilogram) liquid xenon detector that is optimized for a high scintillation gain, and a pulsed neutron source. The measurement uses the recoil energies imparted to xenon nuclei during the de-excitation process following neutron capture, where promptly emitted $γ$ cascades can provide the nuclei with up to $0.3$ keV$_\text{nr}$ of recoil energy due to conservation of momentum. A successful calibration of scintillation photon and ionization electron yields below…
The ionization yield in the two-phase liquid xenon dark-matter detector has been studied in keV nuclear-recoil energy region. The newly-obtained nuclear quenching as well as the recently-measured average energy required to produce an electron-ion pair are used to calculate the total electric charges produced. To estimate the fraction of the electron charges collected, the Thomas-Imel model is generalized to describing the field dependence for nuclear recoils in liquid xenon. With free parameters fitted to experiment measured 56.5 keV nuclear recoils, the energy dependence of ionization yield f…
XENON10 is an experiment designed to directly detect particle dark matter. It is a dual phase (liquid/gas) xenon time-projection chamber with 3D position imaging. Particle interactions generate a primary scintillation signal (S1) and ionization signal (S2), which are both functions of the deposited recoil energy and the incident particle type. We present a new precision measurement of the relative scintillation yield \leff and the absolute ionization yield Q_y, for nuclear recoils in xenon. A dark matter particle is expected to deposit energy by scattering from a xenon nucleus. Knowledge of \l…
Measuring the scintillation and ionization yields of liquid xenon in response to ultra-low energy nuclear recoil events is necessary to increase the sensitivity of liquid xenon experiments to light dark matter. Neutron capture on xenon can be used to produce nuclear recoil events with energies below $0.3$ keV$_\text{NR}$ via the asymmetric emission of $γ$ rays during nuclear de-excitation. The feasibility of an ultra-low energy nuclear recoil measurement using neutron capture was investigated for the Michigan Xenon (MiX) detector, a small dual-phase xenon time projection chamber that is optimi…
The dependence of the light and charge yield of liquid xenon on the applied electric field and recoil energy is important for dark matter detectors using liquid xenon time projections chambers. Few measurements have been made of this field dependence at recoil energies less than 10 keV. In this paper we present results of such measurements using a specialized detector. Recoil energies are determined via the Compton coincidence technique at four drift fields relevant for liquid xenon dark matter detectors: 0.19, 0.48, 1.02, and 2.32 kV/cm. Mean recoil energies down to 1 keV were measured with u…
Liquid xenon detectors such as XENON10 and XENON100 obtain a significant fraction of their sensitivity to light (<10 GeV) particle dark matter by looking for nuclear recoils of only a few keV, just above the detector threshold. Yet in this energy regime a correct treatment of the detector threshold and resolution remains unclear. The energy dependence of the scintillation yield of liquid xenon for nuclear recoils also bears heavily on detector sensitivity, yet numerous measurements have not succeeded in obtaining concordant results. In this article we show that the ratio of detected ionization…
We perform a theoretical study of the scintillation efficiency in the low-energy region crucial for liquid-xenon dark-matter detectors. We develop a computer program to simulate the cascading process of the recoiling xenon nucleus in liquid xenon and calculate the nuclear quenching effect due to atomic collisions. We use the electronic stopping power extrapolated from the experimental data to the low-energy region, and take into account the effects of electrons escaping from the electron-ion pair recombination using the generalized Thomas-Imel model fitted to scintillation data. Our result agr…
We present a measurement of the extraction efficiency of quasi-free electrons from the liquid into the gas phase in a two-phase xenon time-projection chamber. The measurements span a range of electric fields from 2.4 to 7.1 kV/cm in the liquid xenon, corresponding to 4.5 to 13.1 kV/cm in the gaseous xenon. Extraction efficiency continues to increase at the highest extraction fields, implying that additional charge signal may be attained in two-phase xenon detectors through careful high-voltage engineering of the gate-anode region.
We report the testing of a charcoal-based Kr-83m source for use in calibrating a low background two-phase liquid xenon detector. Kr-83m atoms produced through the decay of Rb-83 are introduced into a xenon detector by flowing xenon gas past the Rb-83 source. 9.4 keV and 32.1 keV transitions from decaying 83Krm nuclei are detected through liquid xenon scintillation and ionization. The characteristics of the Kr-83m source are analyzed and shown to be appropriate for a low background liquid xenon detector. Introduction of Kr-83m allows for quick, periodic calibration of low background noble liqui…
We present new measurements of the scintillation and ionization yields in liquid xenon for low energy electronic (about 3--7 keV$_{ee}$) and nuclear recoils (about 8--20 keV$_{nr}$) at different drift fields from 236 V/cm to 3.93 kV/cm, using a three-dimensional sensitive liquid xenon time projection chamber with high energy and position resolutions. Our measurement of signal responses to nuclear recoils agrees with predictions from the NEST model. However, our measured ionization (scintillation) yields for electronic recoils are consistently higher (lower) than those from the NEST model by ab…
We present measurements of the scintillation pulse shape in liquid xenon for nuclear recoils (NR) and electronic recoils (ER) at electric fields of 0 to 0.5 kV/cm for energies $<$ 15 keV and $<$ 70 keV electron-equivalent, respectively. The average pulse shapes are well-described by an effective model with two exponential decay components, where both decay times are fit parameters. We find significant broadening of the pulse for ER due to delayed luminescence from the recombination process. In addition to the effective model, we fit a model describing the recombination luminescence for ER at z…
Dual phase time projection chamber using liquid xenon as target material is one of most successful detectors for dark matter direct search, and has improved the sensitivities of searching for weakly interacting massive particles by almost five orders of magnitudes in past several decades. However, it still remains a great challenge for dual phase liquid xenon time projection chamber to be used as the detector in next-generation dark matter search experiments ($\sim$ 50 tonne sensitive mass), in terms of reaching sufficiently high field strength for drifting electrons, and sufficiently low back…
The Noble Element Simulation Technique (NEST) is an exhaustive collection of models explaining both the scintillation light and ionization yields of noble elements as a function of particle type (nuclear recoil, electron recoil, alphas), electric field, and incident energy or energy loss dE/dx. It is packaged as C++ code for Geant4 that implements said models, overriding the default model which does not account for certain complexities, such as the reduction in yields for nuclear recoils (NR) compared to electron recoils (ER). We present here improvements to the existing NEST models and update…
Scintillation light produced in liquid xenon (LXe) by alpha particles, electrons and gamma-rays was detected with a large area avalanche photodiode (LAAPD) immersed in the liquid. The alpha scintillation yield was measured as a function of applied electric field. We estimate the quantum efficiency of the LAAPD to be 45%. The best energy resolution from the light measurement at zero electric field is 7.5%(sigma) for 976 keV internal conversion electrons from Bi-207 and 2.6%(sigma) for 5.5 MeV alpha particles from Am-241. The detector used for these measurements was also operated as a gridded io…
We have built and operated a crystalline/vapor xenon TPC, with the goal of improving searches for dark matter. The motivation for this instrument is the fact that beta decays from the radon decay chain to the ground state presently limit the state-of-the-art liquid/vapor xenon experiments. In contrast, a crystalline xenon target has the potential to tag and reject radon-chain backgrounds, due to the time and energy signature of their decays. The present article is the first demonstration of a crystalline/vapor xenon TPC with electroluminescence (gas gain) for the electron signal readout. It al…
Low-energy electronic recoil caused by solar neutrinos in multi-ton xenon detectors is an important subject not only because it is a source of the irreducible background for direct searches of weakly-interacting massive particles (WIMPs), but also because it provides a viable way to measure the solar $pp$ and $^{7}\textrm{Be}$ neutrinos at the precision level of current standard solar model predictions. In this work we perform $\textit{ab initio}$ many-body calculations for the structure, photoionization, and neutrino-ionization of xenon. It is found that the atomic binding effect yields a siz…
We extend the study of the performance of a prototype two-phase liquid xenon WIMP dark matter detector to recoil energies below 20 keV. We demonstrate a new method for obtaining the best estimate of the energies of events using a calibrated sum of charge and light signals and introduce the corresponding discrimination parameter, giving its mean value at 4 kV/cm for electron and nuclear recoils up to 300 and 100 keV, respectively. We show that fluctuations in recombination limit discrimination for most energies, and reveal an improvement in discrimination below 20 keV due to a surprising increa…
The search for dark matter is one of today's most exciting fields. As bigger detectors are being built to increase their sensitivity, background reduction is an ever more challenging issue. To this end, a new type of dark matter detector is proposed, a xenon bubble chamber, which would combine the strengths of liquid xenon TPCs, namely event by event energy resolution, with those of a bubble chamber, namely insensitivity to electronic recoils. In addition, it would be the first time ever that a dark matter detector is active on all three detection channels, ionization and scintillation charact…
Dual-phase liquid xenon (LXe) detectors lead the direct search for particle dark matter. Understanding the signal production process of nuclear recoils in LXe is essential for the interpretation of LXe based dark matter searches. Up to now, only two experiments have simultaneously measured both the light and charge yield at different electric fields, neither of which attempted to evaluate the processes leading to light and charge production. In this letter, results from a neutron calibration of liquid xenon with simultaneous light and charge detection are presented for energies from 3-74 keV, …
Neutrinos from supernova (SN) bursts can give rise to detectable number of nuclear recoil (NR) events through the coherent elastic neutrino-nucleus scattering (CE$ν$NS) process in large scale liquid xenon detectors designed for direct dark matter search, depending on the SN progenitor mass and distance. Here we show that in addition to the direct NR events due to CE$ν$NS process, the SN neutrinos can give rise to additional nuclear recoils due to the elastic scattering of neutrons produced through inelastic interaction of the neutrinos with the xenon nuclei. We find that the contribution of th…
XeBRA is a flexible cryogenic platform designed to perform research and development for liquid xenon detectors searching for rare events. Its extra-large outer cryostat makes it possible to install a wide variety of detector designs. We present the system, including its cryogenic, gas handling, data acquisition and slow control subsystems. Two dual phase time projection chambers with sensitive masses at the 1 kg scale have so far been operated in XeBRA. Using data from these, we determine the field-dependence of the electron drift velocity in liquid xenon. We also measure the relative charge a…
We present measurements of light and charge signals in a dual-phase time projection chamber at electric fields varying from 10 V/cm up to 500 V/cm and at zero field using 511 keV gamma rays from a $^{22}$Na source. We determine the drift velocity, electron lifetime, diffusion constant, and light and charge yields at 511 keV as a function of the electric field. In addition, we fit the scintillation pulse shape to an effective exponential model, showing a decay time of 43.5 ns at low field that decreases to 25 ns at high fields.
Dual-phase xenon TPC detectors are a highly scalable and widely used technology to search for low-energy nuclear recoil signals from WIMP dark matter or coherent nuclear scattering of $\sim$MeV neutrinos. Such experiments expect to measure O(keV) ionization or scintillation signals from such sources. However, at $\sim1\,$keV and below, the signal calibrations in liquid xenon carry large uncertainties that directly impact the assumed sensitivity of existing and future experiments. In this work, we report a new measurement of the ionization yield of nuclear recoil signals in liquid xenon down to…
The idea of a hermetic quartz chamber in a dual-phase xenon time projection chamber (TPC) has the potential to improve the detector sensitivity for direct dark matter searches in the future. A major challenge facing TPC detectors in future dark matter experiments will be the reduction of the internal background such as $^{222}$Rn and the deterioration of the ionization signal due to electronegative impurities. The hermetic quartz chamber can isolate the TPC's sensitive volume from external interference and is thus expected to prevent contamination caused by the radioactive and electronegative …
Liquid Xenon (LXe) is an excellent material for experiments designed to detect dark matter in the form of Weakly Interacting Massive Particles (WIMPs). A low energy detection threshold is essential for a sensitive WIMP search. The understanding of the relative scintillation efficiency (Leff) and ionization yield of low energy nuclear recoils in LXe is limited for energies below 10 keV. In this paper, we present new measurements that extend the energy down to 4 keV, finding that Leff decreases with decreasing energy. We also measure the quenching of scintillation efficiency due to the electric …
The continuous emanation of $^{222}$Rn from detector surfaces causes the dominant background in current liquid xenon time projection chambers (TPCs) searching for dark matter. A significant reduction is required for the next generation of detectors which are aiming to reach the neutrino floor, such as DARWIN. $^{222}$Rn-induced back\-grounds can be reduced using a hermetic TPC, in which the sensitive target volume is mechanically separated from the rest of the detector containing the majority of Rn-emanating surfaces. We present a hermetic TPC that mainly follows the well-established design of…
We calibrate a two-phase xenon detector at 0.27 keV in the charge channel and at 2.8 keV in both the light and charge channels using a $^{37}$Ar source that is directly released into the detector. We map the light and charge yields as a function of electric drift field. For the 2.8 keV peak, we calculate the Thomas-Imel box parameter for recombination and determine its dependence on drift field. For the same peak, we achieve an energy resolution, $E_σ/E_{mean}$, between 9.8% and 10.8% for 0.1 kV/cm to 2 kV/cm electric drift fields.
We report a comprehensive study of the energy response to low-energy recoils in dual-phase xenon-based dark matter experiments. A recombination model is developed to explain the recombination probability as a function of recoil energy at zero field and non-zero field. The role of e-ion recombination is discussed for both parent recombination and volume recombination. We find that the volume recombination under non-zero field is constrained by a plasma effect, which is caused by a high density of charge carriers along the ionization track forming a plasma-like cloud of charge that shields the i…
The dual-phase xenon time projection chamber (TPC) is a leading detector technology in rare event searches for dark matter and neutrino physics. The success of this type of detector technology relies on its capability to detect both primary scintillation and ionization signals from particle interactions in liquid xenon (LXe). The ionization electrons are converted into electroluminescence in the gas xenon (GXe), where a single electron can be amplified by more than 100 times in number of photons in a strong electric field. Maintaining a strong and uniform electric field in the small gas gap in…
The liquid xenon (LXe) time projection chamber (TPC) technology is probing a wide range of dark matter masses from sub-GeV to a few TeV. To further improve its sensitivity to sub-GeV dark matter and its application in reactor neutrino monitoring via coherent elastic neutrino-nucleus scattering (CEvNS), more understanding and suppression of single/few electrons background rate are needed. Here we report on the design and performance of a sealed LXeTPC with a graphene-coated fused silica window as the cathode. The purpose of the sealed TPC is for isolating the liquid xenon target volume from the…
The search for dark matter, the missing mass of the Universe, is one of the most active fields of study within particle physics. The XENON1T experiment recently observed a 3.5$σ$ excess potentially consistent with dark matter, or with solar axions. Here, we will use the Noble Element Simulation Technique (NEST) software to simulate the XENON1T detector, reproducing the excess. We utilize different detector efficiency and energy reconstruction models, but they primarily impact sub-keV energies and cannot explain the XENON1T excess. However, using NEST, we can reproduce their excess in multiple,…
Dual phase Xenon Time Projection Chambers (XeTPCs) are being used by several experiments as a promising technique for direct detection of dark matter. We report on the design and performance of a small 3-D sensitive dual phase XeTPC. The position resolution is 2 mm in the center of detector, limited by the hole size of the mesh at the proportional scintillation region. An energy resolution of 1.6%(σ /E) for 662 keV gamma rays is achieved by combining the ionization and scintillation signals at a drift field of 0.5 kV/cm. This represents the best energy resolution achieved among liquid xenon de…
We present the results of the first experimental study of ionization yield of electron recoils with energies below 100 keV produced in liquid xenon by the isotopes: 37Ar, 83mKr, 241Am, 129Xe, 131Xe. It is confirmed by a direct measurement with 37Ar isotope (2.82 keV) that the ionization yield is growing up with the energy decrease in the energy range below ~ 10 keV accordingly to the NEST predictions. Decay time of scintillation at 2.82 keV is measured to be 25 +/- 3 ns at the electric field of 3.75 kV/cm.
Ionization and scintillation produced by nuclear recoils in gaseous xenon at approximately 14 bar have been simultaneously observed in an electroluminescent time projection chamber. Neutrons from radioisotope $α$-Be neutron sources were used to induce xenon nuclear recoils, and the observed recoil spectra were compared to a detailed Monte Carlo employing estimated ionization and scintillation yields for nuclear recoils. The ability to discriminate between electronic and nuclear recoils using the ratio of ionization to primary scintillation is demonstrated. These results encourage further inves…
We present a systematic study of the time and band-resolved scintillation in xenon-based time projection chambers (TPCs), performed simultaneously for the primary (S1) and secondary (S2) components in a small, purity-controlled, setup. We explore a range of conditions of general academic interest, focusing on those of relevance to contemporary TPCs: pressure range ([1- 10] bar), pressure-reduced electric fields ([0-100] V/cm/bar in the drift region (S1) and up to the proportional scintillation regime in the multiplication region (S2)), wavelength-band ([145-250], [250-400], [400-600] nm), and …
The response of liquid xenon to low-energy electronic recoils is relevant in the search for dark-matter candidates which interact predominantly with atomic electrons in the medium, such as axions or axion-like particles, as opposed to weakly interacting massive particles which are predicted to scatter with atomic nuclei. Recently, liquid-xenon scintillation light has been observed from electronic recoils down to 2.1 keV, but without applied electric fields that are used in most xenon dark matter searches. Applied electric fields can reduce the scintillation yield by hindering the electron-ion …
The viability of using a Bubble Chamber for rare event searches and in particular for the detection of dark matter particle candidates is considered. Techniques leading to the deactivation of inhomogeneous nucleation centers and subsequent enhanced stability in such a detector are described. Results from prototype trials indicate that sensitivity to low-energy nuclear recoils like those expected from Weakly Interacting Massive Particles can be obtained in conditions of near total insensitivity to minimum ionizing backgrounds. An understanding of the response of superheated heavy refrigerants t…
A novel technique has been developed to build vessels for liquid xenon ionization detectors entirely out of ultra-clean fluoropolymer. We describe the advantages in terms of low radioactivity contamination, provide some details of the construction techniques, and show the energy resolution achieved with a prototype all-fluoropolymer ionization detector.
We characterize single- and few-electron backgrounds that are observed in dual-phase liquid xenon time projection chambers at timescales greatly exceeding a maximum drift time after an interaction. These instrumental backgrounds limit a detector's sensitivity to dark matter and cosmogenic neutrinos. Using the ~150g liquid xenon detector at Purdue University, we investigate how these backgrounds, produced after 122keV $^{57}$Co Compton interactions, behave under different detector conditions. We find that the rates of single- and few-electron signals follow power-laws with time after the intera…
We report a systematic determination of the responses of PandaX-II, a dual phase xenon time projection chamber detector, to low energy recoils. The electron recoil (ER) and nuclear recoil (NR) responses are calibrated, respectively, with injected tritiated methane or $^{220}$Rn source, and with $^{241}$Am-Be neutron source, within an energy range from $1-25$ keV (ER) and $4-80$ keV (NR), under the two drift fields of 400 and 317 V/cm. An empirical model is used to fit the light yield and charge yield for both types of recoils. The best fit models can well describe the calibration data. The sys…
Scintillation and ionisation yields for nuclear recoils in liquid xenon above 10 keVnr (nuclear recoil energy) are deduced from data acquired using broadband Am-Be neutron sources. The nuclear recoil data from several exposures to two sources were compared to detailed simulations. Energy-dependent scintillation and ionisation yields giving acceptable fits to the data were derived. Efficiency and resolution effects are treated using a light collection Monte Carlo, measured photomultiplier response profiles and hardware trigger studies. A gradual fall in scintillation yield below ~40 keVnr is fo…
Understanding and quantifying the gamma-induced bubble nucleation background in clean nuclear recoil detection bubble chambers is of utmost importance to bubble chamber based dark matter searches. We present data confirming the hypothesis that large Auger cascades from high-Z elements such as iodine and xenon dramatically increase the response of C$_3$F$_8$ bubble chambers to gamma rays. These tests, performed with a small calibration bubble chamber filled with C$_3$F$_8$+$\mathcal{O}$(10) ppm xenon, show that the probability of bubble nucleation scales with the rate of xenon inner-shell vacan…
Directional detection of nuclear recoils is broadly desirable in nuclear and particle physics. At low recoil energies, this capability may be used to confirm the cosmological origin of a dark matter signal, to penetrate the so-called neutrino floor, or to distinguish between different neutrino sources. Gas Time Projection Chambers (TPCs) can enable directional recoil detection if the readout granularity is sufficiently high, as is the case when micro-pattern gaseous detectors (MPGDs) are utilized. A key challenge in such detectors is identifying and rejecting background electron recoil events …
A 30-g xenon bubble chamber, operated at Northwestern University in June and November 2016, has for the first time observed simultaneous bubble nucleation and scintillation by nuclear recoils in a superheated liquid. This chamber is instrumented with a CCD camera for near-IR bubble imaging, a solar-blind photomultiplier tube to detect 175-nm xenon scintillation light, and a piezoelectric acoustic transducer to detect the ultrasonic emission from a growing bubble. The time of nucleation determined from the acoustic signal is used to correlate specific scintillation pulses with bubble-nucleating…
Bubble chambers using liquid xenon (and liquid argon) have been operated (resp. planned) by the Scintillating Bubble Chamber (SBC) collaboration for GeV-scale dark matter searches and CE$ν$NS from reactors. This will require a robust calibration program of the nucleation efficiency of low-energy nuclear recoils in these target media. Such a program has been carried out by the PICO collaboration, which aims to directly detect dark matter using $\mathrm{C_3 F_8}$ bubble chambers. Neutron calibration data from mono-energetic neutron beam and Am-Be source has been collected and analyzed, leading t…
We report on the response of liquid xenon to low energy electronic recoils below 15 keV from beta decays of tritium at drift fields of 92 V/cm, 154 V/cm and 366 V/cm using the XENON100 detector. A data-to-simulation fitting method based on Markov Chain Monte Carlo is used to extract the photon yields and recombination fluctuations from the experimental data. The photon yields measured at the two lower fields are in agreement with those from literature; additional measurements at a higher field of 366 V/cm are presented. The electronic and nuclear recoil discrimination as well as its dependence…