A Cosmic Ray Measurement Facility for ATLAS Muon Chambers
O. Biebel, M. Binder, M. Boutemeur, A. Brandt, J. Dubbert, G. Duckeck, J. Elmsheuser, F. Fiedler, R. Hertenberger, O. Kortner, T. Nunnemann, F. Rauscher, D. Sch… 展开作者
O. Biebel, M. Binder, M. Boutemeur, A. Brandt, J. Dubbert, G. Duckeck, J. Elmsheuser, F. Fiedler, R. Hertenberger, O. Kortner, T. Nunnemann, F. Rauscher, D. Schaile, P. Schieferdecker, A. Staude, W. Stiller, R. Stroehmer, R. Vertesi
备注:
14+1 pages, 11 figures, contributed paper to the EPS2003 conference, Aachen
Monitored Drift Tube (MDT) chambers will constitute the large majority of precision detectors in the Muon Spectrometer of the ATLAS experiment at the Large Hadron Collider at CERN. For commissioning and calibration of MDT chambers, a Cosmic Ray Measurement Facility is in operation at Munich University. The objectives of this facility are to test the chambers and on-chamber electronics, to map the positions of the anode wires within the chambers with the precision needed for standalone muon momentum measurement in ATLAS, and to gain experience in the operation of the chambers and on-line calibration procedures.
Until the start of muon chamber installation in ATLAS, 88 chambers built at the Max Planck Institute for Physics in Munich have to be commissioned and calibrated. With a data taking period of one day individual wire positions can be measured with an accuracy of 8.3 micrometers in the chamber plane and 27 micrometers in the direction perpendicular to that plane.
With half the drift-tube diameter of the Monitored Drift Tube (MDT) chambers of the ATLAS muon spectrometer and otherwise unchanged operating parameters, small-diameter Muon Drift Tube (sMDT) chambers provide an order of magnitude higher rate capability and can be installed in detector regions where MDT chambers do not fit. The chamber assembly time has been reduced by a factor of seven to one working day and the sense wire positioning accuracy improved by a factor of two to better than ten microns. Two sMDT chambers have been installed in ATLAS in 2014 to improve the momentum resolution in th…
The ATLAS muon spectrometer uses drift-tube chambers for precision tracking. The performance of these chambers in the presence of magnetic field and high radiation fluxes is studied in this article using test-beam data recorded in the Gamma Irradiation Facility at CERN. The measurements are compared to detailed predictions provided by the Garfield drift-chamber simulation programme.
For the upgrade of the ATLAS muon spectrometer in March 2014 new muon tracking chambers (sMDT) with drift-tubes of 15 mm diameter, half of the value of the standard ATLAS Monitored Drift-Tubes (MDT) chambers, and 10~$μ$m positioning accuracy of the sense wires have been constructed. The new chambers are designed to be fully compatible with the present ATLAS services but, with respect to the previously installed ATLAS MDT chambers, they are assembled in a more compact geometry and they deploy two additional tube layers that provide redundant rack information. The chambers are composed of 8 laye…
Precision drift tube chambers with a sense wire positioning accuracy of better than 20 microns are under construction for the ATLAS muon spectrometer. 70% of the 88 large chambers for the outermost layer of the central part of the spectrometer have been assembled. Measurements during chamber construction of the positions of the sense wires and of the sensors for the optical alignment monitoring system demonstrate that the requirements for the mechanical precision of the chambers are fulfilled.
Small-diameter Muon Drift Tube (sMDT) chambers have been developed for the ATLAS muon detector upgrade. They possess an improved rate capability and a more compact design with respect to the existing chambers, which allows to equip detector regions uninstrument at present. The chamber assembly methods have been optimized for mass production, while the sense wire positioning accuracy is improved to below ten microns. The chambers will be ready for installation in the winter shutdown 2016/17 of the Large Hadron Collider. The design and construction of the new sMDT chambers for ATLAS will be disc…
The Monitored Drift Tube (MDT) chambers for the muon spectrometer of the AT- LAS detector at the Large Hadron Collider (LHC) consist of 3-4 layers of pressurized drift tubes on either side of a space frame carrying an optical monitoring system to correct for deformations. The full-scale prototype of a large MDT chamber has been constructed with methods suitable for large-scale production. X-ray measurements at CERN showed a positioning accuracy of the sense wires in the chamber of better than the required 20 ?microns (rms). The performance of the chamber was studied in a muon beam at CERN. Cha…
ATLAS muon detector upgrades aim for increased acceptance for muon triggering and precision tracking and for improved rate capability of the muon chambers in the high-background regions of the detector with increasing LHC luminosity. The small-diameter Muon Drift Tube (sMDT) chambers have been developed for these purposes. With half of the drift-tube diameter of the MDT chambers and otherwise unchanged operating parameters, sMDT chambers share the advantages of the MDTs, but have an order of magnitude higher rate capability and can be installed in detector regions where MDT chambers do not fit…
The Monitored Drift Tube (MDT) chambers for the muon spectrometer of the ATLAS detector at the Large Hadron Collider (LHC) consist of 3-4 layers of pressurised drift tubes on either side of a space frame carrying an optical deformation monitoring system. The chambers have to provide a track position resolution of 40 microns with a single-tube resolution of at least 80 microns and a sense wire positioning accu- racy of 20 ?microns (rms). The feasibility was demonstrated with the full-scale prototype of one of the largest MDT chambers with 432 drift tubes of 3.8 m length. For the ATLAS muon spec…
Calibration of the discriminator threshold of the Amplifier-Shaper-Discriminator Chip for the Upgrade of the ATLAS Muon Drift-Tube Chambers for the High-Luminosity LHC
A new small-diameter Monitored Drift Tube (sMDT) chamber has been developed for the muon spectrometer of the ATLAS experiment to handle the higher collision rates expected at the CERN High Luminosity Large Hadron Collider (HL-LHC). This paper presents measurements of the tracking resolution and hit efficiency of two prototype sMDT chambers constructed at the University of Michigan. Using cosmic-ray muons the sMDT tracking resolution of 103.7$\pm8.1$ \textmu m was measured for one chamber and 101.8$\pm$7.8 \textmu m for the other, compared with a design resolution of 106 \textmu m. A further tr…
For the muon spectrometer of the ATLAS detector at the large hadron collider (LHC), large drift chambers consisting of 6 to 8 layers of pressurized drift tubes are used for precision tracking covering an active area of 5000 m2 in the toroidal field of superconducting air core magnets. The chambers have to provide a spatial resolution of 41 microns with Ar:CO2 (93:7) gas mixture at an absolute pressure of 3 bar and gas gain of 2?104. The environment in which the chambers will be operated is characterized by high neutron and background with counting rates of up to 100 per square cm and second. T…
The goals of the ongoing and planned ATLAS muon detector upgrades are to increase the acceptance for precision muon momentum measurement and triggering and to improve the rate capability of the muon chambers in the high-background regions corresponding to the increasing LHC luminosity. Smalldiameter Muon Drift Tube (sMDT) chambers have been developed for these purposes. With half the drift-tube diameter of the current ATLAS Muon Drift Tube (MDT) chambers with 30 mm drift tube diameter and otherwise unchanged operating parameters, the sMDT chambers share all the advantages of the MDTs, but have…
The ATLAS Muon Spectrometer is designed to measure the momentum of muons with a resolution of dp/p = 3% and 10% at 100 GeV and 1 TeV momentum respectively. For this task, the spectrometer employs 355,000 Monitored Drift Tubes (MDTs) arrayed in 1200 Chambers. Calibration (RT) functions convert drift time measurements into tube-centered impact parameters for track segment reconstruction. RT functions depend on MDT environmental parameters and so must be appropriately calibrated for local chamber conditions. We report on the creation and application of a gas monitor system based calibration progr…
The ATLAS experiment at the Large Hadron Collider (LHC) at CERN is currently waiting to record the first collision data in spring 2009. Its muon spectrometer is designed to achieve a momentum resolution of 10% pT(mu) = 1 TeV/c. The spectrometer consists of a system of three superconducting air-core toroid magnets and is instrumented with three layers of Monitored Drift Tube chambers (Cathode Strip Chambers in the extreme forward region) as precision detectors. Resistive Plate Chambers in the barrel and Thin Gap Chambers in the endcap regions provide a fast trigger system. The spectrometer pass…
The ATLAS muon spectrometer consists of several major components: Monitored Drift Tubes (MDTs) for precision measurements in the bending plane of the muons, supplemented by Cathode Strip Chambers (CSC) in the high eta region; Resistive Plate Chambers (RPCs) and Thin Gap Chambers (TGCs) for trigger and second coordinate measurement in the barrel and endcap regions, respectively; an optical alignment system to track the relative positions of all chambers; and, finally, the world's largest air-core magnetic toroid system. We will describe the status and commissioning of the muon system with cosmi…
Highly selective triggers are essential for the physics programme of the ATLAS experiment at HL-LHC where the instantaneous luminosity will be about an order of magnitude larger than the LHC instantaneous luminosity in Run 1. The first level muon trigger rate is dominated by low momentum muons below the nominal trigger threshold due to the moderate momentum resolution of the Resistive Plate and Thin Gap trigger chambers. The resulting high trigger rates at HL-LHC can be sufficiently reduced by using the data of the precision Muon Drift Tube chambers for the trigger decision. This requires the …
The influence of fast neutrons on the occupancy and the single tube resolution of ATLAS muon drift detectors was investigated by exposing a chamber built out of 3 layers of 3 short standard drift tubes to neutron flux-densities of up to 16 kHz/cm2 at a neutron energy of E=11 MeV. Pulse shape capable NE213 scintillaton detectors and a calibrated BF3 neutron detector provided monitoring of the neutron flux-density and energy. The sensitivity of the drift chamber to the neutrons was measured to be 4*10-4 by comparing data sets with and without neutron background. For the investigation of tracks o…
Five small prototype micromegas detectors were positioned in the ATLAS detector during Large Hadron Collider running at $\sqrt{s} = 7$ and $8\, \mathrm{TeV}$. A $9\times 4.5\, \mathrm{cm^2}$ double drift gap detector was placed in front of the electromagnetic calorimeter and four $9\times 10\, \mathrm{cm^2}$ detectors on the ATLAS Small Wheel, the first station of the forward muon spectrometer. The one attached to the calorimeter was exposed to interaction rates of about $70\,\mathrm{kHz}/\mathrm{cm^2}$ at $\mathcal{L}=5\times 10^{33}\,\mathrm{cm^{-2}s^{-1}}$ two orders of magnitude higher tha…
The resolution and efficiency of a precision drift-tube chamber for the ATLAS muon spectrometer with final read-out electronics was tested at the Gamma Irradiation Facility at CERN in a 100 GeV muon beam and at photon irradiation rates of up to 990 Hz/square cm which corresponds to twice the highest background rate expected in ATLAS. A silicon strip detector telescope was used as external reference in the beam. The pulse-height measurement of the read-out electronics was used to perform time-slewing corrections which lead to an improvement of the average drift-tube resolution from 104 microns …
Data sets with high statistics taken at the cosmic ray facility, equipped with 3 ATLAS BOS MDT chambers, in Garching (Munich) have been used to study temperature and pressure effects on gas gain and drifttime. The deformation of a thermally expanded chamber was reconstructed using the internal RasNik alignment monitoring system and the tracks from cosmic data. For these studies a heating system was designed to increase the temperature of the middle chamber by up to 20 Kelvins over room temperature. For comparison the temperature effects on gas properties have been simulated with Garfield. The …
The Phase-II Upgrade of the ATLAS Muon Detector requires new electronics for the readout of the MDT drift tubes. The first processing stage, the Amplifier-Shaper-Discriminator (ASD), determines the performance of the readout for crucial parameters like time resolution, gain uniformity, efficiency and noise rejection. An 8-channel ASD chip, using the IBM 130 nm CMOS 8RF-DM technology, has been designed, produced and tested. The area of the chip is 2.2 x 2.9 square mm size. We present results of detailed measurements as well as a comparision with simulation results of the chip behaviour at three…
Following the Higgs particle discovery, the Large Hadron Collider complex will be upgraded in several phases allowing the luminosity to increase to $7 \times 10^{34}cm^{-2}s^{-1}$. In order to adapt the ATLAS detector to the higher luminosity environment after the upgrade, part of the ATLAS muon end-cap system, the Small Wheel, will be replaced by the New Small Wheel. The New Small Wheel includes two kinds of detectors: small-strip Thin Gap Chambers and Micromegas. Shandong University, part of the ATLAS collaboration, participates in the construction of the ATLAS New Small Wheel by developing,…
Highly selective first-level triggers are essential to exploit the full physics potential of the ATLAS experiment at High-Luminosity LHC (HL-LHC). The concept for a new muon trigger stage using the precision monitored drift tube (MDT) chambers to significantly improve the selectivity of the first-level muon trigger is presented. It is based on fast track reconstruction in all three layers of the existing MDT chambers, made possible by an extension of the first-level trigger latency to six microseconds and a new MDT read-out electronics required for the higher overall trigger rates at the HL-LH…
Large area Micromegas detectors will be employed for the first time in high-energy physics experiments. A total surface of about $\mathbf{150~m^2}$ of the forward regions of the Muon Spectrometer of the ATLAS detector at LHC will be equipped with 8-layer Micromegas modules. Each layer covers more than $\mathbf{2~m^2}$ for a total active area of $\mathbf{1200~m^2}$. Together with the small strip Thin Gap Chambers they will compose the two New Small Wheels, which will replace the innermost stations of the ATLAS endcap muon tracking system in the 2018/19 shutdown. In order to achieve a 15$\mathbf…
A quality assurance and performance qualification laboratory was built at McGill University for the Canadian-made small-strip Thin Gap Chamber (sTGC) muon detectors produced for the 2019-2020 ATLAS experiment muon spectrometer upgrade. The facility uses cosmic rays as a muon source to ionise the quenching gas mixture of pentane and carbon dioxide flowing through the sTGC detector. A gas system was developed and characterised for this purpose, with a simple and efficient gas condenser design utilizing a Peltier thermoelectric cooler (TEC). The gas system was tested to provide the desired 45 vol…
The High-Luminosity LHC will provide the unique opportunity to explore the nature of physics beyond the Standard Model of strong and electroweak interactions. Highly selective first level triggers are essential for the physics programme of the ATLAS experiment at the HL-LHC where the instantaneous luminosity will exceed the LHC Run 1 instantaneous luminosity by almost an order of magnitude. The ATLAS first level muon trigger rate is dominated by low momentum muons, selected due to the moderate momentum resolution of the resistive plate and thin gap trigger chambers. This limitation can be over…
This paper describes the small-diameter monitored drift-tube detector construction at the University of Michigan as a contribution to the ATLAS Muon Spectrometer upgrade for the high-luminosity Large Hadron Collider at CERN. Measurements of the first 30 chambers built at Michigan show that the drift tube wire position accuracy meets the specification of 20 microns. The positions of the platforms for alignment and magnetic field sensors are all installed well within the required precision. The cosmic ray test measurements show single wire tracking resolution of 100 +- 7 microns with an average …
ATLAS is a detector under construction to explore the physics at the Large Hadron Collider at CERN. It has a muon spectrometer with an excellent momentum resolution of 3-10%, provided by three layers of precision monitored-drift-tube chambers in a toroidal magnetic field. A single drift tube measures a track point with a mean resolution close to 100 micron, even at the expected high neutron and gamma background rates. The tubes are positioned within the chamber with an accuracy of 20 microns, achieved by elaborate construction and assembly monitoring procedures.
Small-diameter Drift Tube (sMDT) detectors with 15 mm tube diameter have proven to be excellent candidates for precision muon tracking detectors in experiments at future hadron colliders like HL-LHC and FCC-hh where unprecedentedly high background rate capabilities are required. sMDT chambers are currently being installed in the inner barrel layer of the ATLAS muon spectrometer. The rate capability of the sMDT drift tubes in terms of muon detection efficiency and spatial resolution is limited by the performance of the readout electronics. A new (s)MDT ASD (Amplifier-Shaper-Discriminator) reado…
Thin gap chambers (TGCs) are used for the muon trigger system in the forward region of the LHC experiment ATLAS. A TGC consists of a plane of closely spaced wires maintained at positive high voltage, sandwiched between resistive grounded cathode planes with an anode wire to cathode plane gap distance smaller than the wire-to-wire spacing. The TGCs are expected to provide a trigger signal within 25 ns of the bunch spacing of the LHC accelerator, with an efficiency exceeding 95%, while exposed to an effective photon and neutron background ranging from 30 to 500 Hz/cm2. About 2,500 out of the 3,6…
The Monitored Drift Tubes, as a part of the ATLAS muon spectrometer, are precision drift chambers designed to provide excellent spatial resolution and high tracking efficiency independent of the track angle. Through the life of the LHC and ATLAS experiment, this detector has already demonstrated that they provide precise tracking over large areas. The aim of the ATLAS muon spectrometer upgrade is to increase the muon trigger efficiency, precise muon momentum measurement and to improve the rate capability of the muon system in the high-background regions during the High-Luminosity LHC runs. To …
A full-size prototype of a Micromegas precision tracking chamber for the upgrade of the ATLAS detector at the LHC Collider has been built between October 2015 and April 2016. This paper describes in detail the procedures used in constructing the single modules of the chamber in various INFN laboratories and the final assembly at the Laboratori Nazionali di Frascati (LNF). Results of the chamber exposure to the CERN SPS/H8 beam line in June 2016 are also presented. The performances achieved in the construction and the results of the test beam are compared with the requirements, which are impose…
The outer shell of the ATLAS experiment at the LHC consists of a system of toroidal air-core magnets in order to allow for the precise measurement of the transverse momentum p$_T$ of muons, which in many physics channels are a signature of interesting physics processes. For the precise determination of the muon momentum Monitored Drift Tube chambers (MDT) with high position accuracy are used, while for the fast identification of muon tracks chambers with high time resolution are used, able to select muons above a predefined p$_T$ threshold for use in the first Level of the ATLAS triggering sys…
The innermost station of the present ATLAS forward muon detector needs to be upgraded for the super-LHC. We present a proposal to replace it with a sandwiched detector composed of several layers of small-radius Monitored Drift Tube chambers (sMDT) for precision tracking measurement and two stations of multi-gap Resistive Plate Chambers (mRPC) for triggering purpose. We describe the layout of the upgraded detector and the trigger strategy. Several modifications to the RPCs used in the ATLAS barrel region are needed to satisfy the super-LHC requirements. Various studies with the proposed mRPC ti…
SHIP is a new general purpose fixed target facility, proposed at the CERN SPS accelerator. In five years, $2\times 10^{20}$ protons of 400 GeV/c momentum will be dumped on a Molybdenum target. A detector downstream of the target will allow a search to made for long-lived particles with masses below O(10) GeV/c$^2$ foreseen in several extensions of the Standard Model. Another dedicated detector will allow the study of active neutrino cross-sections and angular distributions. The neutrino detector consists of an emulsion target, based on the Emulsion Cloud Chamber technology fruitfully employed …
The main results of the quality assurance tests performed on the Resistive Plate Chamber used by the ATLAS experiment at LHC as muon trigger chambers are reported and discussed. Since July 2004, about 270 RPC units has been certified at INFN Lecce site and delivered to CERN, for being integrated in the final muon station of the ATLAS barrel region. We show the key RPC characteristics which qualify the performance of this detector technology as muon trigger chamber in the harsh LHC enviroments. These are dark current, chamber efficiency, noise rate, gas volume tomography, and gas leakage.
An extensive system test of the ATLAS muon spectrometer has been performed in the H8 beam line at the CERN SPS during the last four years. This spectrometer will use pressurized Monitored Drift Tube (MDT) chambers and Cathode Strip Chambers (CSC) for precision tracking, Resistive Plate Chambers (RPCs) for triggering in the barrel and Thin Gap Chambers (TGCs) for triggering in the end-cap region. The test set-up emulates one projective tower of the barrel (six MDT chambers and six RPCs) and one end-cap octant (six MDT chambers, A CSC and three TGCs). The barrel and end-cap stands have also been…
Resistive Plate Chambers are used in the ATLAS experiment for the muon trigger and two coordinate measurements in the central region. In preparation for the coming years of LHC running at higher luminosity, besides the New Small Wheel project which is expected to complement the ATLAS muon spectrometer in the endcap regions, a smaller size project, known as BIS78, is being developed with a foreseen installation during the LHC Long Shutdown 2. The BIS78 project will reinforce the fake muon rejection and the selectivity of the muon trigger in the transition region between the ATLAS barrel and end…
After the high-luminosity upgrade of the Large Hadron Collider (LHC) at CERN, the ATLAS muon spectrometer is expected to work at 10 times increased background rates of gammas and neutrons. This is challenging as the momentum resolution of the spectrometer is expected to be 10 %. This requires a single tube resolution of the muon drift tubes of 80 mum. At background rates around 1000 Hz/cm2 space charge effects will lead in the slow and non-linear AR:CO2 = 93:7 gas mixture to a degradation of the drift-tube spatial resolution. This was studied before experimentally for gammas and low energetic …
The instantaneous luminosity of the Large Hadron Collider at CERN will be increased up to a factor of five with respect to the present design value by undergoing an extensive upgrade program over the coming decade. The most important upgrade project for the ATLAS Muon System is the replacement of the present first station in the forward regions with the so-called New Small Wheels (NSWs). The NSWs will be installed during the LHC long shutdown in 2018/19. Small-Strip Thin Gap Chamber (sTGC) detectors are designed to provide fast trigger and high precision muon tracking under the high luminosity…
New front-end electronics including ASICs and FPGA boards are under development for the ATLAS Monitored Drift Tube (MDT) detector to handle the large data rates and harsh environment expected at high-luminosity LHC runs. A mobile Data Acquisition (miniDAQ) system is designed to perform integration tests of these front-end electronics. In addition, it will be used for surface commissioning of 96 small-radius MDT (sMDT) chambers and for integration and commissioning of new front-end electronics on the present ATLAS MDT chambers. Details of the miniDAQ hardware and firmware are described in this …
Small-diameter muon drift tube (sMDT) detectors have been developed for upgrades of the ATLAS muon spectrometer. With a tube diameter of 15 mm, they provide an about an order of magnitude higher rate capability than the present ATLAS muon tracking detectors, the MDT chambers with 30 mm tube diameter. The drift-tube design and the construction methods have been optimised for mass production and allow for complex shapes required for maximising the acceptance. A record sense wire positioning accuracy of 0.005 mm has been achieved with the new design. 14 new sMDT chambers are already operational i…
Thin gap chambers (TGCs) are used for the muon trigger system in the forward region of the LHC experiment ATLAS. The TGCs are expected to provide a trigger signal within 25 ns of the bunch spacing. An extensive system test of the ATLAS muon spectrometer has been performed in the H8 beam line at the CERN SPS during the last few years. A relational database was used for storing the conditions of the tests as well as the configuration of the system. This database has provided the detector control system with the information needed for configuration of the front end electronics. The database is us…
Small-diameter muon drift tube (sMDT) chambers with 15 mm tube diameter are a cost-effective technology for high-precision muon tracking over large areas at high background rates as expected at future high-energy hadron colliders including HL-LHC. The chamber design and construction procedures have been optimized for mass production and provide sense wire positioning accuracy of better than 10 ?m. The rate capability of the sMDT chambers has been extensively tested at the CERN Gamma Irradiation Facility. It exceeds the one of the ATLAS muon drift tube (MDT) chambers, which are operated at unpr…
Surface Commissioning and Performance of the sMDT Muon Chambers for the ATLAS HL-LHC Upgrade
University of Michigan(密歇根大学);Max Planck Institute for Physics(马克斯·普朗克物理研究所);Tufts University(塔夫茨大学);Chinese University of Hong Kong(香港中文大学);Michigan State University(密歇根州立大学);University of Science and Technology of China(中国科学技术大学)
Thin gap chambers (TGCs) are built for the muon trigger chambers in the endcap region of the LHC experiment ATLAS. More than 2500 ATLAS TGCs are being produced at the Weizmann institute in Israel, and in Shandong University in China. Detailed testing of these chambers is performed at the Technion and at the Tel-Aviv University. Two cosmic ray hodoscopes for testing the operation of these detectors were built in Israel. In these hodoscopes the response of the chambers to energetic cosmic ray muons is recorded and analyzed. The hodoscopes measure the exact time and space location of the cosmic r…
This paper reports on the design and construction of infrastructure and test stations for small-diameter monitored drift tube (sMDT) assembly and testing at the University of Michigan (UM) to prepare for the ATLAS Muon Spectrometer upgrade for the high-luminosity program of the Large Hadron Collider. Procedures of the tube assembly and quality assurance and control (QA/QC) tests are described in detail. More than 99% of the tubes meet the tube QA/QC specifications based on 2100 tubes built at UM. The UM test stations are also used for QA/QC testing on the tubes constructed at Michigan State Un…
The LHC upgrade scheduled for 2018 is expected to significantly increase the accelerator's luminosity, and as a result the radiation background rates in the ATLAS Muon Spectrometer will increase too. Some of its components will have to be replaced in order to cope with these high rates. Newly designed small-strip Thin Gap chambers (sTGC) will replace them at the small wheel region. One of the differences between the sTGC and the currently used TGC is the alignment of the wires along the azimuthal direction. As a result, the outermost wires approach the detector's edge with a small angle. Such …
Pressurized drift-tube chambers are efficient detectors for high-precision tracking over large areas. The Monitored Drift-Tube (MDT) chambers of the muon spectrometer of the ATLAS detector at the Large Hadron Collider (LHC) reach a spatial resolution of 35 micons and almost 100% tracking efficiency with 6 layers of 30 mm diameter drift tubes operated with Ar:CO2 (93:7) gas mixture at 3 bar and a gas gain of 20000. The ATLAS MDT chambers are designed to cope with background counting rates due to neutrons and gamma-rays of up to about 300 kHz per tube which will be exceeded for LHC luminosities …
The ATLAS muon spectrometer is designed to measure muon momenta with a resolution of 4% @ 100 GeV/c rising to 10% @ 1 TeV/c track momentum. The spectrometer consists of precision tracking and trigger chambers embedded in a 2T magnetic field generated by three large air-core superconducting toroids. The precision detectors provide 50 micron tracking resolution to a pseudo-rapidity of 2.7. The system also includes an optical monitoring system which measures detector positions with 40 micron precision. This paper reports on the calibration and performance of the ATLAS muon spectrometer.
The Neutrinos at the Main Injector (NuMI) facility is a conventional neutrino beam which produces muon neutrinos by focusing a beam of mesons into a long evacuated decay volume. We have built four arrays of ionization chambers to monitor the position and intensity of the hadron and muon beams associated with neutrino production at locations downstream of the decay volume. This article describes the chambers' construction, calibration, and commissioning in the beam.
The High-Luminosity LHC (HL-LHC) will provide the unique opportunity to explore the nature of physics beyond the Standard Model of strong and electroweak interactions. Highly selective first-level triggers are essential for the physics programme of the ATLAS experiment at HL-LHC, where the instantaneous luminosity will exceed the instantaneous LHC Run 1 luminosity by about an order of magnitude. The ATLAS first-level muon trigger rate is dominated by low momentum muons, which are accepted because of the moderate momentum resolution of the RPC and TGC trigger chambers. This limitation can be ov…
Gamma-Beams at the HIgS facility in the USA and anticipated at the ELI-NP facility, now constructed in Romania, present unique new opportunities to advance research in nuclear astrophysics; not the least of which is resolving open questions in oxygen formation during stellar helium burning via a precise measurement of the 12C(a,g) reaction. Time projection chamber (TPC) detectors operating with low pressure gas (as an active target) are ideally suited for such studies. We review the progress of the current research program and plans for the future at the HIγS facility with the optical readout …
The Compressed Baryonic Matter~(CBM) experiment in the upcoming Facility for Antiproton and Ion Research~(FAIR), designed to take data in nuclear collisions at very high interaction rates of up to 10 MHz, will employ a free-streaming data acquisition with self-triggered readout electronics, without any hardware trigger. A simulation framework with a realistic digitization of the detectors in the muon chamber (MuCh) subsystem in CBM has been developed to provide a realistic simulation of the time-stamped data stream. In this article, we describe the implementation of the free-streaming detector…
Small-diameter muon drift tube (sMDT) chambers are a cost-effective technology for high-precision muon tracking. The rate capability of the sMDT chambers has been extensively tested at the Gamma Irradiation Facility at CERN in view of expected rates at future high-energy hadron colliders. Results show that it fulfills the requirements over most of the acceptance of muon detectors. The optimization of the read-out electronics to further increase the rate capability of the detectors is discussed. Chambers of this type are under construction for upgrades of the muon spectrometer of the ATLAS dete…