arXivDaily arXiv每日学术速递 周一至周五更新
arXiv physics/0207129physics.med-phphysics.ins-det

In-Vivo NMR of Hyperpolarized Helium-3 in the Human Lung at Very Low Magnetic Fields

Christopher P. Bidinosti, Jamal Choukeife, Pierre-Jean Nacher, Geneviève Tastevin

英文摘要:

We present NMR measurements of the diffusion of hyperpolarized helium-3 in the human lung performed at fields much lower than those of conventional MRI scanners. The measurements were made on standing subjects using homebuilt apparatus operating at 3 mT. Oxygen-limited transverse relaxation (T2 up to 15-35 s) could be measured in-vivo. Accurate global diffusion measurements have been performed in-vivo and in-vitro. 1D ADC mapping with high SNR demonstrates the real possibility of performing quality lung imaging at extremely low fields.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑