用于1.5T医用肺扫描仪的高场~ 3 He亚稳态交换光泵偏振器MRI

A high-field 3He Metastability Exchange Optical Pumping polarizer operating in a 1.5 T medical scanner for lung MRI

G. Collier, T. Pałasz, A. Wojna, B. Głowacz, M. Suchanek, Z. Olejniczak, T. Dohnalik

摘要:

利用亚稳态交换光泵(MEOP)技术对3 He进行超极化处理后,可以作为肺磁共振成像(MRI)的对比剂。MEOP通常在低磁场(~ 1 mT)和低压(~ 1 mbar)下进行,这导致低磁化产生速率。还需要精细的偏振保持压缩步骤。在密封室中证明,如果在高磁场(非标准条件)下进行,则可以在较高压力下使用MEOP获得高的核极化值。在这项工作中,建立一个高场偏振器,在商业1.5 T扫描仪的可行性进行了评估。在不同的3 He气体压力(1.33至267毫巴)的密封细胞进行核极化的初步测量。与高斯光束形状相比,激光束的环形形状的使用使在32和67毫巴下可实现的核极化平衡值(Meq)增加了25%。在32和267毫巴下分别获得66.4和31%的Meq值,并且与在标准条件下获得的最佳结果相比,磁化产生速率增加了10倍。为了研究该方法在极化系统中的再现性,用连接到气体处理系统的小电池进行相同的实验。在20-30毫巴压力范围内的极化建立时间比32毫巴密封电池获得的极化建立时间长约3倍。然而,在90 mL开放池中实现了40- 60%的合理Meq值。在此基础上,设计并搭建了一种新型的小型偏光系统。它的典型输出是3 He气体流速为15 sccm,极化率为33%。在体内肺MRI通气图像采集,以证明偏振器的应用。

英文摘要:

After being hyperpolarized using the technique of Metastability Exchange Optical Pumping (MEOP), 3He can be used as a contrast agent for lung magnetic resonance imaging (MRI). MEOP is usually performed at low magnetic field (~ 1 mT) and low pressure (~ 1 mbar), which results in a low magnetization production rate. A delicate polarization-preserving step of compression is also required. It was demonstrated in sealed cells that high nuclear polarization values can be obtained at higher pressures with MEOP, if performed at high magnetic field (non-standard conditions). In this work the feasibility of building a high-field polarizer that operates within a commercial 1.5 T scanner was evaluated. Preliminary measurements of nuclear polarization with sealed cells filled at different 3He gas pressures (1.33 to 267 mbar) were performed. The use of an annular shape for the laser beam increased by 25 % the achievable nuclear polarization equilibrium value (Meq) at 32 and 67 mbar as compared to a Gaussian beam shape. Meq values of 66.4 and 31 % were obtained at 32 and 267 mbar respectively and the magnetization production rate was increased by a factor of 10 compared to the best results obtained under standard conditions. To study the reproducibility of the method in a polarizing system, the same experiments were performed with small cells connected to a gas handling system. The polarization build-up times were approximately 3 times longer in the 20-30 mbar range of pressure than those obtained for the 32 mbar sealed cell. However, reasonable Meq values of 40-60 % were achieved in a 90 mL open cell. Based on these findings, a novel compact polarizing system was designed and built. Its typical output is a 3He gas flow rate of 15 sccm with a polarization of 33 %. In-vivo lung MRI ventilation images were acquired to demonstrate the polarizer s application.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑