Noble gases and liquids are excellent scintillators and this opens a unique opportunity to directly detect the primary scintillation light produced in these media by photons or particles. This signal can be used for several purposes, for example as a start signal for TPCs or for particles identification. Usually photomultipliers (PMs) are used for the detection of the scintillation light. In our previous work we have demonstrated that costly PMs could be replaced by gaseous detectors with CsI photocathodes . Such detectors have the same quantum efficiency as the best PMs but at the same time are cheap, simple and have high position and time resolutions. The aim of this work is to evaluate various planar type gaseous detectors with CsI photocahodes in order to choose the best one for the detection of the primary scintillation light from noble gases and liquids.
Large Area Avalanche Photodiodes (LAAPDs) were used for a series of systematic measurements of the scintillation light in Ar, Kr, and Xe gas. Absolute quantum efficiencies are derived. Values for Xe and Kr are consistent with those given by the manufacturer. For the first time we show that argon scintillation (128 nm) can be detected at a quantum efficiency above 40%. Low-pressure argon gas is shown to emit significant amounts of non-UV radiation. The average energy expenditure for the creation of non-UV photons in argon gas at this pressure is measured to be below 378 eV.
There are several applications and fundamental research areas which require the detection of VUV light at cryogenic temperatures. For these applications we have developed and successfully tested special designs of gaseous detectors with solid photocathodes able to operate at low temperatures: sealed gaseous detectors with MgF2 windows and windowless detectors. We have experimentally demonstrated, that both primary and secondary (due to the avalanche multiplication inside liquids) scintillation lights could be recorded by photosensitive gaseous detectors. The results of this work may allow one …
Radiation detectors with noble gasses as the active medium are becoming increasingly common in experimental programs searching for physics beyond the standard model. Nearly all of these experiments rely to some degree on collecting scintillation light from noble gasses. The VUV wavelengths associated with noble gas scintillation mean that most of these experiments use a fluorescent material to shift the direct scintillation light into the visible or near UV band. We present an overview of the R&D program at LBL related to noble gas detectors for neutrino physics, double beta decay, and dark ma…
The performance of a two-phase Ar avalanche detector with CsI photocathode was studied, with regard to potential application in coherent neutrino-nucleus scattering and dark matter search experiments. The detector comprised a 1 cm thick liquid Ar layer and a triple-GEM multiplier operated in the saturated vapor above the liquid phase; the CsI photocathode was deposited on the first GEM. Successful detection of both primary scintillation and ionization signals, produced by beta-particles in liquid Ar, has for the first time been demonstrated in the two-phase avalanche mode.
Wavelength shifting fibers are widely used for light collection from scintillation counters, which allow connection of various scintillation planes to relatively small photocathodes of photodetectors and especially tiny photocathodes of silicon photo-multipliers. In October 2020 Kuraray announced production of a new branch of faster fibers. We performed a comparison of the new fiber YS-2 to a mature Y-11. The fiber YS-2 demonstrated decay time nearly two times shorter than that of Y-11: $\approx$4.0 ns versus $\approx$7.4 ns. At the same time its light yield and attenuation length are as good …
A novel concept is proposed for large-volume single-phase noble-liquid TPC detectors for rare events. Both radiation-induced scintillation-light and ionization-charge are detected by Liquid Hole-Multipliers, LHM, immersed in the noble liquid. The latter may consist of cascaded Gas Electron Multipliers (GEM), Thick Gas Electron Multiplier (THGEM) electrodes or others, coated with CsI UV-photocathodes. Electrons, photo-induced on CsI by primary scintillation in the noble liquid, and event-correlated drifting ionization electrons are amplified in the cascaded elements primarily through electrolum…
Purity monitor devices are increasingly used in noble gas time projection chambers to measure the lifetime of drifting electrons. Purity monitors work by emitting electrons from a photocathode material via the photoelectric effect. The electrons are then drifted towards an anode by means of an applied electric drift field. By measuring the difference in charge between the cathode and the anode, one can extract the lifetime of the drifting electrons in the medium. For the first time, we test the performance of different photocathode materials-- silver, titanium, and aluminium--and compare them …
We present a model of the ionization efficiency, or quenching factor, for low-energy nuclear recoils, based on a solution to Lindhard integral equation with binding energy and apply it to the calculation of the relative scintillation efficiency and charge yield for nuclear recoils in noble liquid detectors. The quenching model incorporates a constant average binding energy together with an electronic stopping power proportional to the ion velocity, and is an essential input in an analysis of charge recombination processes to predict the ionization and scintillation yields. Our results are comp…
Noble liquid detectors have become an attractive option for exploring physics beyond the standard model. Current experiments are using these detectors to search for dark matter interactions, neutrinoless double beta decay, and other phenomena. Improved energy resolution can be leveraged from an optimized combination of two detection channels: ionization and scintillation. Experimentally, a microscopic anti-correlation behavior between these signals has been observed, but it has not been described from first principles. Making measurements in a third channel would provide useful information abo…
Far-ultraviolet (FUV) scintillation signals have been measured in heavy noble gases (argon, krypton, xenon) following boron-neutron capture ($^{10}$B($n,α$)$^7$Li) in $^{10}$B thin films. The observed scintillation yields are comparable to the yields from some liquid and solid neutron scintillators. At noble gas pressures of 107 kPa, the number of photons produced per neutron absorbed following irradiation of a 1200 nm thick $^{10}$B film was 14,000 for xenon, 11,000 for krypton, and 6000 for argon. The absolute scintillation yields from the experimental configuration were calculated using dat…
The detection of primary scintillation light in combination with the charge or secondary scintillation signals is an efficient technique to determine the events t=0 as well as particle / photon separation in large mass TPC detectors filled with noble gases and/or condensed noble gases. The aim of this work is to demonstrate that costly photo-multipliers could be replaced by cheap novel photosensitive gaseous detectors: wire counters, GEMs or glass capillary tubes coupled with CsI photocathodes. We have performed systematic measurements with Ar, Kr and Xe gas at pressures in the range of 1-50 a…
A comprehensive model for explaining scintillation yield in liquid xenon is introduced. We unify various definitions of work function which abound in the literature and incorporate all available data on electron recoil scintillation yield. This results in a better understanding of electron recoil, and facilitates an improved description of nuclear recoil. An incident gamma energy range of O(1 keV) to O(1 MeV) and electric fields between 0 and O(10 kV/cm) are incorporated into this heuristic model. We show results from a Geant4 implementation, but because the model has a few free parameters, im…
Liquid noble based particle detectors often use the organic wavelength shifter 1,1,4,4- tetraphenyl-1,3-butadiene (TPB) which shifts UV scintillation light to the visible regime, facilitating its detection, but which also can scintillate on its own. Dark matter searches based on this type of detector commonly rely on pulse-shape discrimination (PSD) for background mitigation. Alpha-induced scintillation therefore represents a possible background source in dark matter searches. The timing characteristics of this scintillation determine whether this background can be mitigated through PSD. We ha…
Since the 1970s it has been known that noble gases scintillate in the near infrared (NIR) region of the spectrum (0.7 $μ$m < $λ$; < 1.5$μ$m). More controversial has been the question of the NIR light yield for condensed noble gases. We first present the motivation for using the NIR scintillation in liquid argon detectors, then briefly review early as well as more recent efforts and finally show encouraging preliminary results of a test performed at Fermilab.
The properties of UV-photon imaging detectors consisting of CsI-coated THGEM electron multipliers are summarized. New results related to detection of Cherenkov light (RICH) and scintillation photons in noble liquid are presented.
We present the precision calibration of 35 Hamamatsu R11410-22 photomultiplier tubes (PMTs) with xenon scintillation light centred near 175 nm. This particular PMT variant was developed specifically for the LUX-ZEPLIN (LZ) dark matter experiment. A room-temperature xenon scintillation cell coupled to a vacuum cryostat was used to study the full-face PMT response at both room and low temperature $\textrm{($\sim$ -100$^\circ$C)}$, in particular to determine the quantum efficiency (QE) and double photoelectron emission (DPE) probability in LZ operating conditions. For our sample with an average Q…
We further study the performance of single, double and triple Gas Electron Multiplier (GEM) detectors in pure noble gases at high pressures, in the range of 1-10 atm. We confirm that light noble gases, in particular He and its mixtures with Kr, have the highest gain, reaching 10^6, and an unusual gain dependence on pressure. Effects of the number of GEMs, GEM hole diameter and pitch are investigated in detail. In He, avalanche-induced secondary scintillations are observed at high gains, using the metal photocathode. These results are relevant in the field of avalanche mechanism in noble gases …
Gaseous PhotoMultipliers (GPM) are a very promising alternative of vacuum PMTs especially for large-size noble-liquid detectors in the field of Functional Nuclear Medical Imaging and Direct Dark Matter Detection. We present recent characterization results of a Hybrid-GPM made of three Micropattern Gaseous Structures; a Thick Gaseous Electron Multiplier (THGEM), a Parallel Ionization Multiplier (PIM) and a MICROMesh GAseous Structure (MICROMEGAS),operating in Ne/CF4 (90:10). Gain values close to 10^7 were recorded in this mixture, with 5.9keV x-rays at 1100 mbar, both at room temperature and at…
Detection of the vacuum ultraviolet (VUV) scintillation light produced by liquid noble elements is a central challenge in order to fully exploit the available timing, topological, and calorimetric information in detectors leveraging these media. In this paper, we characterize a novel, windowless amorphous selenium based photodetector with direct sensitivity to VUV light. We present here the manufacturing and experimental setup used to operate this detector at low transport electric fields (2.7-5.2 V/$μ$m) and across a wide range of temperatures (77K-290K). This work shows that the first proof-…
Presented here are first tests of a Gaseous Photomultiplier based on a cascade of Thick GEM structures intended for gamma-ray position reconstruction in liquid Argon. The detector has a MgF$_2$ window, transparent to VUV light, and a CsI photocathode deposited on the first THGEM. A gain of $8\cdot10^{5}$ per photoelectron and $\sim100\%$ photoelectron collection efficiency are measured at stable operation settings. The excellent position resolution capabilities of the detector (better than 100 $μ$m) at 100 kHz readout rate, is demonstrated at room temperature. Structural integrity tests of the…
We review latest progress in gaseous photomultipliers (GPM) combining solid photocathodes and various types of novel electron multipliers. Cascaded gaseous electron multipliers (GEM) coated with CsI photocathodes can efficiently replace UVsensitive wire chambers for single-photon recording in Cherenkov and other detectors. Other hole-multipliers with patterned electrodes (Micro-Hole and Strip Plates) and improved ion-blocking properties are discussed; these permit reducing considerably photon- and ion-induced secondary effects. Photon detectors with other electron-multiplier techniques are bri…
The study of primary and secondary scintillations in noble gases and liquids is of paramount importance to rare-event experiments using noble gas media. In the present work, the scintillation yield in gaseous and liquid Ar has for the first time been measured in the near infrared (NIR) and visible region, both for primary and secondary (proportional) scintillations, using Geiger-mode avalanche photodiodes (G-APDs) and pulsed X-ray irradiation. The primary scintillation yield of the fast component was measured to be 17000 photon/MeV in gaseous Ar in the NIR, in the range of 690-1000 nm, and 510…
We present several novel ionization-electron and scintillation-photon recording concepts in noble-liquid detectors, for future applications in particle and astroparticle physics and in other fields. These involve both single- and dual-phase detector configurations with combined electroluminescence and small charge multiplication in gas and liquid media.
Many rare event searches including dark matter direct detection and neutrinoless double beta decay experiments take advantage of the high VUV reflective surfaces made from polytetrafluoroethylene (PTFE) reflector materials to achieve high light collection efficiency in their detectors. As the detectors have grown in size over the past decade, there has also been an increased need for ever thinner detector walls without significant loss in reflectance to reduce dead volumes around active noble liquids, outgassing, and potential backgrounds. We report on the experimental results to measure the d…
Scintillators are employed for particle detection and identification using light-pulse shapes and light quenching factors. We developed a comprehensive model describing the light generation and quenching in CaWO$_4$ single crystals used for direct dark matter search. All observed particle-dependent light-emission characteristics can be explained quantitatively, light-quenching factors and light-pulse shapes are calculated on a microscopic basis. This model can be extended to other scintillators such as inorganic crystal scintillators, liquid noble gases or organic liquid scintillators.
Detection of scintillation light in noble gases with wavelength-shifting optical fibers
Donostia International Physics Center, BERC Basque Excellence Research Centre(多诺西亚国际物理中心,BERC巴斯克卓越研究中心);Ikerbasque (Basque Foundation for Science)(伊克拉巴斯克(巴斯克科学基金会));Instituto de Física Corpuscular (IFIC), CSIC & Universitat de València(粒子物理研究所(IFIC),西班牙国家研究委员会与瓦伦西亚大学);Department of Physics, Universidad del Pais Vasco (UPV/EHU)(巴斯克国立大学(UPV/EHU)物理系);Nuclear Research Center Negev(内盖夫核研究中心)
ArDM is a new-generation WIMP detector which will measure simultaneously light and charge from scintillation and ionization of liquid argon. Our goal is to construct, characterize and operate a 1 ton liquid argon underground detector. The project relies on the possibility to extract the electrons produced by ionization from the liquid into the gas phase of the detector, to amplify and read out with Large Electron Multipliers detectors. Argon VUV scintillation light has to be converted with wavelength shifters such as TetraPhenyl Butadiene in order to be detected by photomultipliers with bialka…
Experiments searching for weak interacting massive particles with noble gases such as liquid argon require very low detection thresholds for nuclear recoils. A determination of the scintillation efficiency is crucial to quantify the response of the detector at low energy. We report the results obtained with a small liquid argon cell using a monoenergetic neutron beam produced by a deuterium-deuterium fusion source. The light yield relative to electrons was measured for six argon recoil energies between 11 and 120 keV at zero electric drift field.
We discuss recent advances in the development of cryogenic gaseous photomultipliers (GPM), for possible use in dark matter and other rare-event searches using noble-liquid targets. We present results from a 10 cm diameter GPM coupled to a dual-phase liquid xenon (LXe) TPC, demonstrating - for the first time - the feasibility of recording both primary ("S1") and secondary ("S2") scintillation signals. The detector comprised a triple Thick Gas Electron Multiplier (THGEM) structure with cesium iodide photocathode on the first element; it was shown to operate stably at 180 K with gains above 10^5,…
Dark Matter and Double Beta Decay experiments require extremely low radioactivity within the detector materials. For this purpose, the University of California, Los Angeles and Hamamatsu Photonics have developed the QUartz Photon Intensifying Detector (QUPID), an ultra-low background photodetector based on the Hybrid Avalanche Photo Diode (HAPD) and entirely made of ultraclean synthetic fused silica. In this work we present the basic concept of the QUPID and the testing measurements on QUPIDs from the first production line. Screening of radioactivity at the Gator facility in the Laboratori Naz…
We have developed commercial prototypes of sealed gaseous detectors combined with CsI photocathodes and/or filled with photosensitive vapors. The rirst results of application of these devices for the detection of flames in daylight conditions and for the detection of scintillation lights from noble liquids will be presented. The main conclusion from our studies is that for some applications the sealed UV sensitive gaseous detectors have superior performance (higher practical quantum efficiency and better signal to noise ratio) than existing commercial UV sensitive detectors. Additionally, they…
We have demonstrated that hole-type gaseous detectors, GEMs and capillary plates, can operate up to 77 K. For example, a single capillary plate can operate at gains of above 10E3 in the entire temperature interval between 300 until 77 K. The same capillary plate combined with CsI photocathodes could operate perfectly well at gains (depending on gas mixtures) of 100-1000. Obtained results may open new fields of applications for capillary plates as detectors of UV light and charge particles at cryogenic temperatures: noble liquid TPCs, WIMP detectors or LXe scintillating calorimeters and cryogen…
The field of dark matter direct detection has seen important contributions in recent years from experiments involving liquid noble gases, specifically liquid argon and liquid xenon. These detection media offer many properties deemed useful in this search, including fast scintillation response, charge readout, 3-D position reconstruction, and nuclear recoil discrimination. Part of the very rapid emergence and dominance of noble liquids is due to the fact that these technologies are easily scalable to nearly arbitrary size and mass. However, the physics impact of recent results has called into q…
Particle detectors making use of noble elements in gaseous, liquid, or solid phases are prevalent in neutrino and dark matter experiments and are also used to a lesser extent in collider-based particle physics experiments. These experiments take advantage of both the very large, ultra-pure target volumes achievable and the multiple observable signal pathways possible in noble-element based particle detectors. As these experiments seek to increase their sensitivity, novel and improved technologies will be needed to enhance the precision of their measurements and to broaden the reach of their ph…
We report the preparation of a Kr-83m source and its subsequent use in calibrating a liquid xenon detector. Kr-83m atoms were produced through the decay of Rb-83 atoms trapped in zeolite molecular sieve and were then introduced into liquid xenon. Decaying Kr-83m nuclei were detected through liquid xenon scintillation. Conversion electrons with energies of 9.4 keV and 32.1 keV from the decay of Kr-83m were both observed. This calibration source will allow the characterization of the scintillation and ionization response of noble liquid detectors at low energies, highly valuable for the search f…
The ARAPUCA concept has been proposed as a simple and neat solution for increasing the effective collection area of SiPMs through the shifting and trapping of scintillation light in noble liquids, thus with great potential for improving timing and calorimetry resolution in neutrino and dark matter search experiments using time projection chambers. It is expected to achieve a single photon detection efficiency larger than 1\%. The initial design consists of a box made of highly reflective internal surface material and with an acceptance window for photons composed of two shifters and a dichroic…
Intense infrared (IR) light emission from liquid Ar-Xe mixtures has been observed using 12 keV electron-beam excitation. The emission peaks at a wavelength of 1.18 $μ$m and the half-width of the emission band is 0.1 $μ$m. Maximum intensity has been found for a 10 ppm xenon admixture in liquid argon. The conversion efficiency of electron beam-power to IR-light is about 1% (10000 photons per MeV electron energy deposited). A possible application of this intense IR emission for a new particle discrimination concept in liquid noble gas detectors is discussed. No light emission was found for perfec…
We have exposed a dual-phase Liquid Argon Time Projection Chamber (LAr-TPC) to a low energy pulsed narrowband neutron beam, produced at the Notre Dame Institute for Structure and Nuclear Astrophysics to study the scintillation light yield of recoiling nuclei in a LAr-TPC. A liquid scintillation counter was arranged to detect and identify neutrons scattered in the LAr-TPC target and to select the energy of the recoiling nuclei. We report the observation of a significant dependence on drift field of liquid argon scintillation from nuclear recoils of 11 keV. This observation is important because,…
The Deep Underground Neutrino Experiment (DUNE) will be a premier facility for exploring long-standing questions about the boundaries of the standard model. Acting in concert with the liquid argon time projection chambers underpinning the far detector design, the DUNE photon detection system will capture ultraviolet scintillation light in order to provide valuable timing information for event reconstruction. To maximize the active area while maintaining a small photocathode coverage, the experiment will utilize a design based on plastic light guides coated with a wavelength-shifting compound, …
A light-yield calibration of an NE 213A organic liquid scintillator detector has been performed using both monoenergetic and polyenergetic gamma-ray sources. Scintillation light was detected in a photomultiplier tube, and the corresponding pulses were subjected to waveform digitization on an event-by-event basis. The resulting Compton edges have been analyzed using a Geant4 simulation of the detector which models both the interactions of the ionizing radiation as well as the transport of scintillation photons. The simulation is calibrated and also compared to well-established prescriptions use…
We performed a time-resolved spectroscopic study of the VUV/UV argon scintillation as a function of pressure and electric field, by means of a wavelength sensitive detector operated with different radioactive sources. Our work conveys new evidence of distinctive features of the argon light which are in contrast with the general assumption that, for particle detection purposes, the scintillation can be considered to be largely monochromatic at 128 nm (second continuum). The wavelength and the time-resolved analysis of the photon emission reveal that the dominant component of the argon scintilla…
Gamma-ray and fast-neutron imaging was performed with a novel liquid xenon (LXe) scintillation detector read out by a Gaseous Photomultiplier (GPM). The 100 mm diameter detector prototype comprised a capillary-filled LXe converter/scintillator, coupled to a triple-THGEM imaging-GPM, with its first electrode coated by a CsI UV-photocathode, operated in Ne/5%CH4 cryogenic temperatures. Radiation localization in 2D was derived from scintillation-induced photoelectron avalanches, measured on the GPM's segmented anode. The localization properties of Co-60 gamma-rays and a mixed fast-neutron/gamma-r…
Silicon photomultipliers (SiPMs) are photosensors widely used for imaging in a variety of high energy and nuclear physics experiments. In noble-gas detectors for double-beta decay and dark matter experiments, SiPMs are attractive photosensors for imaging. However they are insensitive to the VUV scintillation emitted by the noble gases (xenon and argon). This difficulty is overcome in the NEXT experiment by coating the SiPMs with tetraphenyl butadiene (TPB) to convert the VUV light into visible light. TPB requires stringent storage and operational conditions to prevent its degradation by enviro…
Measuring the scintillation light in noble gases is an important detection technique in particle physics. Numerous rare event searches like neutrino beam experiments, neutrino-less double beta-decay, and dark matter searches use argon-based detectors. In liquid argon, the light yield can be enhanced by the addition of a small quantity of xenon, where $\sim 10 - 1000$ ppm are added. The general enhancement mechanism and its pathway via an energy transfer between argon and xenon excimers is well known, however the importance of absorption of argon excimer emission by atomic xenon has not been fu…
We report the testing of a charcoal-based Kr-83m source for use in calibrating a low background two-phase liquid xenon detector. Kr-83m atoms produced through the decay of Rb-83 are introduced into a xenon detector by flowing xenon gas past the Rb-83 source. 9.4 keV and 32.1 keV transitions from decaying 83Krm nuclei are detected through liquid xenon scintillation and ionization. The characteristics of the Kr-83m source are analyzed and shown to be appropriate for a low background liquid xenon detector. Introduction of Kr-83m allows for quick, periodic calibration of low background noble liqui…
We present a model for the Global Quantum Efficiency (GQE) of the MicroBooNE optical units. An optical unit consists of a flat, circular acrylic plate, coated with tetraphenyl butadiene (TPB), positioned near the photocathode of a 20.2-cm diameter photomultiplier tube. The plate converts the ultra-violet scintillation photons from liquid argon into visible-spectrum photons to which the cryogenic phototubes are sensitive. The GQE is the convolution of the efficiency of the plates that convert the 128 nm scintillation light from liquid argon to visible light, the efficiency of the shifted light …
Measurements of electron drift properties in liquid and gaseous xenon are reported. The electrons are generated by the photoelectric effect in a semi-transparent gold photocathode driven in transmission mode with a pulsed ultraviolet laser. The charges drift and diffuse in a small chamber at various electric fields and a fixed drift distance of 2.0 cm. At an electric field of 0.5 kV/cm, the measured drift velocities and corresponding temperature coefficients respectively are $1.97 \pm 0.04$ mm/$μ$s and $(-0.69\pm0.05)$\%/K for liquid xenon, and $1.42 \pm 0.03$ mm/$μ$s and $(+0.11\pm0.01)$\%/K …
A new data analysis method based on physical observables for WIMP dark matter searches with noble liquid Xe dual-phase TPCs is presented. Traditionally, the nuclear recoil energy from a scatter in the liquid target has been estimated by means of the initial prompt scintillation light (S1) produced at the interaction vertex. The ionization charge (C2), or its secondary scintillation (S2), is combined with the primary scintillation in Log(S2/S1) vs. S1 only as a discrimination parameter against electron recoil background. Arguments in favor of C2 as the more reliable nuclear recoil energy estima…
Operation results are presented of a UV-sensitive gaseous photomultiplier (GPM) coupled through a MgF2 window to a liquid-xenon scintillator. It consisted of a reflective CsI photocathode deposited on top of a THick Gaseous Electron Multiplier (THGEM); further multiplication stages were either a second THGEM or a Parallel Ionization Multiplier (PIM) followed by a MICROMEsh GAseous Structure (MICROMEGAS). The GPM operated in gas-flow mode with non-condensable gas mixtures. Gains of 10^4 were measured with a CsI-coated double-THGEM detector in Ne/CH4 (95:5), Ne/CF4 (95:5) and Ne/CH4/CF4 (90:5:5)…
Vacuum ultraviolet light emission from xenon-doped liquid argon is described in the context of liquid noble gas particle detectors. Xenon concentrations in liquid argon from 0.1 ppm to 1000 ppm were studied. The energy transfer from the second excimer continuum of argon ($\sim$127 nm) to the second excimer continuum of xenon ($\sim$174 nm) is observed by recording optical emission spectra. The transfer almost saturates at a xenon concentration of $\sim$10 ppm for which, in addition, an intense emission in the infrared at a peak wavelength of 1.17 $μ$m with (13000$\pm$4000) photons per MeV depo…
Future tonne-scale liquefied noble gas detectors depend on efficient light detection in the VUV range. In the past years Silicon Photomultipliers (SiPMs) have emerged as a valid alternative to standard photomultiplier tubes or large area avalanche photodiodes. The next generation double beta decay experiment, nEXO, with a 5 tonne liquid xenon time projection chamber, will use SiPMs for detecting the $178\,\text{nm}$ xenon scintillation light, in order to achieve an energy resolution of $σ/ Q_{ββ} = 1\, \%$. This paper presents recent measurements of the VUV-HD generation SiPMs from Fondazione …
For the development of liquid argon dark matter detectors we assembled a setup in the laboratory to scatter neutrons on a small liquid argon target. The neutrons are produced mono-energetically (E_kin=2.45 MeV) by nuclear fusion in a deuterium plasma and are collimated onto a 3" liquid argon cell operating in single-phase mode (zero electric field). Organic liquid scintillators are used to tag scattered neutrons and to provide a time-of-flight measurement. The setup is designed to study light pulse shapes and scintillation yields from nuclear and electronic recoils as well as from α-particles …