HUBS探测器开发 I:小面积TES微量热计的初步测试

Detector Development for HUBS I: Initial Testing of Small-Area TES Microcalorimeters

Low Temperature Detector Lab, Department of Astronomy, Tsinghua University(清华大学天文系低温探测器实验室)

Naihui Chen, Jian Ma, Sifan Wang, Qian Wang, Guanhua Gao, Qing Yu, Yajie Liang, Yaowu Song, Jingyi Zhang, Rui Huang, Wei Cui

AI总结:

报告HUBS任务中微量热计探测器的开发进展,测试了Mo/Cu TES阵列像素,在5.9 keV下实现3.8±0.2 eV能量分辨率,并分析了噪声和热串扰效应。

摘要:

我们报告了为热宇宙重子巡天器(HUBS)任务正在进行的微量热计探测器技术的开发进展。我们展示了对10×10微量热计阵列中选定像素进行测试和表征的结果。该微量热计基于与Au吸收体耦合的Mo/Cu过渡边缘传感器(TES)。为了更好地理解器件的特性,我们首先使用产生3 eV光子的脉冲激光系统测量了相同设计TES阵列中选定像素的能量分辨率,发现单个光子峰很容易被TES分辨,表明性能良好。然后,我们将微量热计阵列暴露于$^{55}$Fe源的辐射中,发现测试的像素在5.9 keV下的能量分辨率高达3.8±0.2 eV。所有器件的能量分辨率随偏置点单调变化,几乎没有证据表明存在所谓的过量噪声。这与对测量噪声谱进行建模的结果一致。热串扰效应明显,导致能量分辨率下降。

英文摘要:

We report progress on the ongoing development of microcalorimeter detector technology for the Hot Universe Baryon Surveyor (HUBS) mission. We show the results from testing and characterizing selected pixels in a 10$\times$10 microcalorimeter array. The microcalorimeter is based on a Mo/Cu transition-edge sensor (TES) coupled to an Au absorber. To better understand the properties of the devices, we have first measured the energy resolution of a selected pixel in a TES array of the same design with a pulsed laser system that produces 3 eV photons, and found that individual photon peaks are easily resolved with the TES, indicating good performance. We have then exposed the microcalorimeter array to radiation from a $^{55}$Fe source, and found that the pixels tested show energy resolutions as good as 3.7$\pm$0.1 eV at 5.9 keV. The energy resolution is found to vary monotonically with the bias point for all the devices, showing little evidence for the presence of the so-called excess noise. This is consistent with the results from modeling the measured noise spectrum. The effects of thermal crosstalk are evident, leading to the degradation of energy resolution.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑