包含心脏运动效应的患者特异性胸主动脉模型机械边界条件的校准

Calibration of the mechanical boundary conditions for a patient-specific thoracic aorta model including the heart motion effect

Leonardo Geronzi, Aline Bel-Brunon, Antonio Martinez, Michel Rochette, Marco Sensale, Olivier Bouchot, Alain Lalande, Siyu Lin, Pier Paolo Valentini, Marco Evangelos Biancolini

AI总结:

提出一种校准胸主动脉模型机械边界条件的程序,通过纯结构模拟迭代优化参数以包含心脏运动效应,显著降低了边界距离误差,提高了流固耦合模型重现真实主动脉根部运动学的保真度。

摘要:

目的:我们提出一种校准4个参数的程序,这些参数控制着源自一名升主动脉瘤患者的胸主动脉(TA)模型的机械边界条件(BCs)。该BCs重现了软组织和脊柱提供的粘弹性结构支撑,并允许包含心脏运动效应。方法:我们首先从磁共振成像(MRI)血管造影中分割出TA,并通过追踪cine-MRI中的主动脉瓣环来推导心脏运动。执行刚体壁流体动力学模拟以推导随时间变化的壁面压力场。我们构建有限元模型,考虑患者特异性材料属性,并施加推导出的压力场和瓣环边界处的运动。涉及零压力状态计算的校准基于纯结构模拟。从cine-MRI序列中获取血管边界后,执行迭代程序以最小化这些边界与从变形结构模型推导出的相应边界之间的距离。最后使用调谐参数执行强耦合流固耦合(FSI)分析,并与纯结构模拟进行比较。结果与结论:使用结构模拟的校准将图像推导与模拟推导边界之间的最大和平均距离分别从8.64 mm减少至6.37 mm和从2.24 mm减少至1.83 mm。变形结构与FSI表面网格之间的最大均方根误差为0.19 mm。该程序可能对提高模型在重现真实主动脉根部运动学方面的保真度至关重要。

英文摘要:

Objective: we propose a procedure for calibrating 4 parameters governing the mechanical boundary conditions (BCs) of a thoracic aorta (TA) model derived from one patient with ascending aortic aneurysm. The BCs reproduce the visco-elastic structural support provided by the soft tissue and the spine and allow for the inclusion of the heart motion effect. Methods: we first segment the TA from magnetic resonance imaging (MRI) angiography and derive the heart motion by tracking the aortic annulus from cine-MRI. A rigid-wall fluid-dynamic simulation is performed to derive the time-varying wall pressure field. We build the finite element model considering patient-specific material properties and imposing the derived pressure field and the motion at the annulus boundary. The calibration, which involves the zero-pressure state computation, is based on purely structural simulations. After obtaining the vessel boundaries from the cine-MRI sequences, an iterative procedure is performed to minimize the distance between them and the corresponding boundaries derived from the deformed structural model. A strongly-coupled fluid-structure interaction (FSI) analysis is finally performed with the tuned parameters and compared to the purely structural simulation. Results and Conclusion: the calibration with structural simulations allows to reduce maximum and mean distances between image-derived and simulation-derived boundaries from 8.64 mm to 6.37 mm and from 2.24 mm to 1.83 mm, respectively. The maximum root mean square error between the deformed structural and FSI surface meshes is 0.19 mm. This procedure may prove crucial for increasing the model fidelity in replicating the real aortic root kinematics.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑