Variable Stiffness & Dynamic Force Sensor for Tissue Palpation
AI总结:
针对微创手术中组织触诊受通道限制的问题,研究提出一种利用光反射测形变推算作用力的可变刚度动态力程传感器,实验验证其刚度和力程随压力升高而增大,校准精度较高。
摘要:
微创手术(Minimally Invasive Surgery)过程中人体组织的触诊因操作通道受限而受阻。在这篇扩展摘要中,我们提出了一种可变刚度、动态力程的传感器,有望解决这一难题。该传感器利用光反射估算自身形变,进而推算所受的作用力。在0、0.5、1 PSI三种不同压力下的实验测试表明,传感器的刚度和力程随压力升高而增大。与实测力对比,三种压力下力校准结果的平均RMSE分别为0.016 N、0.0715 N和0.1284 N。
英文摘要:
Palpation of human tissue during Minimally Invasive Surgery is hampered due to restricted access. In this extended abstract, we present a variable stiffness and dynamic force range sensor that has the potential to address this challenge. The sensor utilises light reflection to estimate sensor deformation, and from this, the force applied. Experimental testing at different pressures (0, 0.5 and 1 PSI) shows that stiffness and force range increases with pressure. The force calibration results when compared with measured forces produced an average RMSE of 0.016, 0.0715 and 0.1284 N respectively, for these pressures.