磁锚定手术器械的充气腹壁组织动力学

Dynamics of Insufflated Abdominal Wall Tissue for Magnetically Anchored Surgical Instruments

Florence Leong, Alireza Mohammadi, Vijay Rajagopal, Ying Tan, Dhan Thiruchelvam, Pietro Valdastri, Denny Oetomo

摘要:

磁锚定手术装置最近在腹部手术中获得关注,其中使用磁体将手术装置锚定到吹入的腹壁上。这些锚钉已用于固定无源和有源设备,其中机器人操纵器等有源设备会产生激发非刚性腹壁动态的运动。因此,需要研究与磁锚定相结合的吹入状态下的腹壁组织的机械动力学,特别是其对机械激励的响应和对锚定装置的操作的预期干扰。本文对相应的猪试件进行了加载和卸载试验,以进行动力学识别。实验装置被构造成模拟具有磁锚定机构的腹部的充气状态。卸载过程中的组织响应被捕获,并近似与一般的数值模型,这反过来又用于组织的动态分析,使用波德图。结果表明,在这种拉伸和压缩状态下,组织的稳态位移近似为零。使用高磁锚定力时,最大瞬态误差为1mm。由于腹壁的高刚度和阻尼,在频率响应中从100rad/s观察到干扰的显著衰减。如果机器人操纵器连接到锚定装置,则典型的运动操作频率仍将产生未衰减的干扰。预计需要通过适当的控制策略进行一些误差补偿。这些结果为有源磁锚定手术器械的控制器设计和设计规范奠定了基础,以使对腹壁组织的干扰影响最小。

英文摘要:

Magnetically-anchored surgical devices have recently gained attention in abdominal surgery, with the use of magnets to anchor surgical devices onto the insufflated abdominal wall. These anchors have been used to secure passive and active devices, where active device such as robotic manipulators produce motions that would excite the dynamics of the non-rigid abdominal wall. Hence, there is a need to investigate the mechanical dynamics of the abdominal wall tissue in insufflated state, combined with magnetic anchoring, specifically its response to mechanical excitations and the expected disturbances to the operation of the anchored devices. In this paper, loading and unloading tests are performed on a corresponding porcine specimen for dynamics identification. The experiment setup was constructed to emulate the insufflated state of the abdomen with the magnetically anchored mechanism. The tissue responses during unloading are captured and approximated with a general numerical model, which is in turn used for the dynamic analysis of the tissue using Bode plot. The results showed that in such stretched and compressed state, the steady state displacement of the tissue is approximately zero. The maximum transient error was found to be 1mm in displacement using a high magnetic anchoring force. Significant attenuation of the disturbances due to the high stiffness and damping of the abdominal wall, was observed from 100rad/s in the frequency response. If a robotic manipulator was attached to the anchoring device, the typical operating frequency of movements would still produce unattenuated disturbances. It is expected that some error compensation through suitable control strategies is required. These outcomes establish the basis for the controller design and the design specification of the active magnetically-anchored surgical devices for minimal disturbance impact onto the abdominal wall tissue.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑