arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2406.05657physics.ins-det
收藏

具有改进时间分辨率的单通道PICOSEC Micromegas探测器

Single channel PICOSEC Micromegas detector with improved time resolution

A. Utrobicic, R. Aleksan, Y. Angelis, J. Bortfeldt, F. Brunbauer, M. Brunoldi, E. Chatzianagnostou, J. Datta, K. Dehmelt, G. Fanourakis, D. Fiorina, K. J. Floet… 展开作者

A. Utrobicic, R. Aleksan, Y. Angelis, J. Bortfeldt, F. Brunbauer, M. Brunoldi, E. Chatzianagnostou, J. Datta, K. Dehmelt, G. Fanourakis, D. Fiorina, K. J. Floethner, M. Gallinaro, F. Garcia, I. Giomataris, K. Gnanvo, F. J. Iguaz, D. Janssens, A. Kallitsopoulou, M. Kovacic, B. Kross, P. Legou, M. Lisowska, J. Liu, M. Lupberger, I. Maniatis, J. McKisson, Y. Meng, H. Muller, E. Oliveri, G. Orlandini, A. Pandey, T. Papaevangelou, M. Pomorski, L. Ropelewski, D. Sampsonidis, L. Scharenberg, T. Schneider, L. Sohl, M. van Stenis, Y. Tsipolitis, S. E. Tzamarias, I. Vai, R. Veenhof, P. Vitulo, X. Wang, S. White, W. Xi, Z. Zhang, Y. Zhou

AI总结:

本文提出并验证了一种改进时间分辨率的单通道PICOSEC Micromegas探测器设计,通过优化结构与电气连接、简化重组装流程,并在束流测试中实现12.5±0.8 ps的创纪录时间分辨率。

摘要:

本文介绍了一种具有改进时间分辨率的单通道PICOSEC Micromegas(MM)探测器的设计指南与实验验证。该设计涵盖探测器板、腔体、辅助机械部件以及高压(HV)和信号的电气连接,重点在于提高稳定性、降低噪声并确保信号完整性,以优化定时性能。一个显著特点是简单快速的重新组装流程,便于快速更换探测器内部组件,从而支持涉及不同探测器组件的高效测量策略。本文还研究了寄生参数对输出信号完整性的影响。为验证该设计,制造了一个原型组件和三块可互换的探测器板,其读出焊盘直径各不相同。这些探测器首先在实验室环境中进行了测试。最后,利用最小电离粒子(MIP)束流测试验证了不同焊盘尺寸探测器的定时性能。值得注意的是,采用10 mm直径读出焊盘尺寸的探测器,在CsI光电阴极的PICOSEC Micromegas探测器技术上实现了12.5±0.8 ps的创纪录时间分辨率。

英文摘要:

This paper presents design guidelines and experimental verification of a single-channel PICOSEC Micromegas (MM) detector with an improved time resolution. The design encompasses the detector board, vessel, auxiliary mechanical parts, and electrical connectivity for high voltage (HV) and signals, focusing on improving stability, reducing noise, and ensuring signal integrity to optimize timing performance. A notable feature is the simple and fast reassembly procedure, facilitating quick replacement of detector internal components that allows for an efficient measurement strategy involving different detector components. The paper also examines the influence of parasitics on the output signal integrity. To validate the design, a prototype assembly and three interchangeable detector boards with varying readout pad diameters were manufactured. The detectors were initially tested in the laboratory environment. Finally, the timing performance of detectors with different pad sizes was verified using a Minimum Ionizing Particle (MIP) beam test. Notably, a record time resolution for a PICOSEC Micromegas detector technology with a CsI photocathode of 12.5$\pm$0.8 ps was achieved with a 10 mm diameter readout pad size detector.