无创通气中的双参数漏气估计
Two parameter Leak Estimation in Non invasive Ventilation
摘要:
在本文中,我们提出了一种方法来估计泄漏的无创通气。泄漏的准确估计是呼吸机的关键组成部分,因为它决定了呼吸机在患者-呼吸机同步性和风量输送方面的性能。特别地,在无创通气中,患者流量与在通气机出口处测量的流量显著不同。这主要是由于沿着用于呼气的管的通气孔,但也由于在回路中的其他地方发生的非故意泄漏(例如,面具)。传统上,这种泄漏通过具有两个参数的模型来量化,但其中只有一个参数不断更新-另一个是固定的。新算法允许两个参数的逐呼吸更新。这是通过利用描述患者呼吸力学的模型实现的。
英文摘要:
In this paper we present a method for the estimation of leaks in non-invasive ventilation. Accurate estimation of leaks is a key component of a ventilator, since it determines the ventilator performance in terms of patient-ventilator synchrony and air volume delivery. In particular, in non-invasive ventilation, the patient flow is significantly different from the flow measured at the ventilator outlet. This is mostly due to the vent orifice along the tube that is used for exhalation, but also to the non-intentional leaks that occur elsewhere in the circuit (e.g., at the mask). Such leaks are traditionally quantified via a model with two parameters, but only one of them is continually updated - the other is fixed. The new algorithm allows for breath-by-breath update of both parameters. This was made possible by leveraging a model describing the patient respiratory mechanics.