arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 2107.00883physics.ins-dethep-ex
收藏

单掩模三重GEM探测器的火花概率测量

Spark probability measurement of a single mask triple GEM detector

S. Chatterjee, U. Frankenfeld, C. Garabatos, J. Hehner, T. Morhardt, C. J. Schmidt, H. R. Schmidt, C. A. Lymanets, S. Biswas

摘要:

三重气体电子倍增器探测器将在CBM MUCH的前两个台站用作跟踪装置,在那里,对于8 AGeV的中心Au-Au碰撞,最大粒子速率预计将达到~1 MHz/cm 2。因此,探测器的稳定运行非常重要。在CERN SPS/H4束线装置上测量了单掩模三重GEM探测器在~150 GeV/c π介子束流和高电离簇射粒子环境中的放电几率。研究了不同粒子率下的火花几率与增益的关系。本文详细介绍了实验装置,火花识别方法和结果。

英文摘要:

Triple Gas Electron Multiplier (GEM) detectors will be used as a tracking device in the first two stations of CBM MUon CHamber (MUCH), where the maximum particle rate is expected to reach ~1 MHz/cm2 for central Au-Au collisions at 8 AGeV. Therefore, the stable operation of the detector is very important. Discharge probability has been measured of a single mask triple GEM detector at the CERN SPS/H4 beam-line facility with a pion beam of ~150 GeV/c and also in an environment of highly ionizing shower particles. The spark probability as a function of gain has been studied for different particle rates. The details of the experimental setup, method of spark identification and results are presented in this paper.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑