arXivDaily arXiv每日学术速递 周一至周五更新
感恩节,周五至周末无更新,大家顺遂。周一再来吧
arXiv 2106.05877physics.atom-phphysics.ins-det
收藏

模块化光泵浦磁力计系统

A modular optically pumped magnetometer system

University of Sussex(萨塞克斯大学);Brighton and Sussex Medical School, University of Sussex(布莱顿和萨塞克斯医学院);UKRI-STFC Rutherford Appleton Laboratory(英国研究与创新部STFC拉瑟福德阿普尔顿实验室)
Thomas Coussens, Aikaterini Gialopsou, Christopher Abel, Mark G. Bason, Tim M. James, William Evans, Michael T. M. Woodley, Denilson Nicolau, Leigh Page, Fedja … 展开作者

Thomas Coussens, Aikaterini Gialopsou, Christopher Abel, Mark G. Bason, Tim M. James, William Evans, Michael T. M. Woodley, Denilson Nicolau, Leigh Page, Fedja Orucevic, Peter Kruger

AI总结:

针对医疗和工业的空间分辨磁成像需求,提出含高灵敏度量子磁力计阵列的模块化OPM系统,支持快速重构与多种阵列布局,测试显示其梯度灵敏度达10 fT/cm/√Hz、带宽超200Hz,可测脊髓信号。

摘要:

为满足医疗和工业场景中空间分辨磁成像的需求,我们提出一种模块化光泵浦磁力计(OPM)系统,该系统包含由高灵敏度量子磁力计组成的多传感器阵列。此系统通过实现独立激光与传感器模块的快速重构、利用共享光源构建多种阵列布局,旨在促进新测量方案的快速原型设计与测试。系统的模块化特性有助于开发高密度阵列磁成像的管理方法。我们首先在单通道和差分梯度计配置下表征磁力计的灵敏度和带宽,随后在真实脑磁图环境中通过测量人类参与者大脑的阿尔法节律进行测试。我们展示了一阶轴向梯度计配置的OPM系统,其磁场梯度灵敏度达10 fT/cm/√Hz;两个独立模块实现了超过200 Hz的带宽。该系统更高的时间分辨率可用于测量脊髓信号,我们通过体模信号试验并与现有商用传感器对比验证了这一点。

英文摘要:

To address the demands in healthcare and industrial settings for spatially resolved magnetic imaging, we present a modular optically pumped magnetometer (OPM) system comprising a multi-sensor array of highly sensitive quantum magnetometers. This system is designed and built to facilitate fast prototyping and testing of new measurement schemes by enabling quick reconfiguration of the self-contained laser and sensor modules as well as allowing for the construction of various array layouts with a shared light source. The modularity of this system facilitates the development of methods for managing high-density arrays for magnetic imaging. The magnetometer sensitivity and bandwidth are first characterised in both individual channel and differential gradiometer configurations before testing in a real-world magnetoencephalography environment by measuring alpha rhythms from the brain of a human participant. We demonstrate the OPM system in a first-order axial gradiometer configuration with a magnetic field gradient sensitivity of 10 $\mathrm{fT/cm/\sqrt{Hz}}$. Bandwidths exceeding 200 Hz were achieved for two independent modules. The system's increased temporal resolution allows for the measurement of spinal cord signals, which we demonstrate by using phantom signal trials and comparing with an existing commercial sensor.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑