We report on a compact, tunable, and scalable to large arrays imaging device, based on a radio-frequency optically pumped atomic magnetometer operating in magnetic induction tomography modality. Imaging of conductive objects is performed at room temperature, in an unshielded environment and without background subtraction. Conductivity maps of target objects exhibit not only excellent performance in terms of shape reconstruction but also demonstrate detection of sub-millimetric cracks and penetration of conductive barriers. The results presented here demonstrate the potential of a future generation of imaging instruments, which combine magnetic induction tomography and the unmatched performance of atomic magnetometers.
We demonstrate Magnetic Induction Tomography (MIT) with an all-optical atomic magnetometer. Our instrument creates a conductivity map of conductive objects. Both shape and size of the imaged samples compare very well with the actual shape and size. Given the potential of all-optical atomic magnetometers for miniaturization and extreme sensitivity, the proof-of-principle presented here opens up promising avenues in the development of instrumentation for magnetic induction tomography.
In this work we propose, and experimentally demonstrate, the use of a self-oscillating all-optical atomic magnetometer for magnetic induction measurements. Given the potential for miniaturization of atomic magnetometers, and their extreme sensitivity, the present work shows that atomic magnetometers may play a key role in the development of instrumentation for magnetic induction tomography.
Electromagnetic induction imaging with atomic magnetometers has disclosed unprecedented domains for imaging, from security screening to material characterization. However, applications to low-conductivity specimens -- most notably for biomedical imaging -- require sensitivity, stability, and tunability only speculated thus far. Here, we demonstrate contactless and non-invasive imaging down to 50 S/m using a 50 fT/Hz$^{-1/2}$ $^{87}$Rb radio-frequency atomic magnetometer operating in an unshielded environment and near room temperature. Two-dimensional images of test objects are obtained with a …
We report on a 2x2 array of radio-frequency atomic magnetometers in magnetic induction tomography configuration. Active detection, localization, and real-time tracking of conductive, non-magnetic targets are demonstrated in air and saline water. Penetration in different media and detection are achieved thanks to the sensitivity and tunability of the sensors, and to the active nature of magnetic induction probing. We obtained a 100% success rate for automatic detection and 93% success rate for automatic localization in air and water, up to 190 mm away from the sensors' plane (100 mm underwater)…
We present an apparatus to overcome the limitations of mechanical raster-scanning in electromagnetic induction imaging (EMI) techniques by instead performing a 2D optical raster-scan within the vapour cell of a radio-frequency atomic magnetometer (RF-AM). A large cuboidal 87Rb vapour cell is employed to act as the medium of an RF-AM with the pump and probe beams translated in the cell via acousto-optics. The technique is shown to give robust and repeatable magnetic measurements over the cell volume and successfully resolves conductive targets with EMI. Optical raster-scanning removes the limit…
We demonstrate electromagnetic induction imaging with an unshielded, portable radio-frequency atomic magnetometer scanning over the target object. This configuration satisfies standard requirements in typical applications, from security screening to medical imaging. The ability to scan the magnetometer over the object relies on the miniaturization of the sensor head and on the active compensation of the ambient magnetic field. Additionally, a procedure is implemented to extract high-quality images from the recorded spatial dependent magnetic resonance. The procedure is shown to be effective in…
The sensitive detection of either static or radio-frequency \textsc{(rf)} magnetic fields is essential to many fundamental studies and applications. Here, we demonstrate the operation of a cold-atom-based, \textsc{rf} magnetometer in performing {1-D and 2-D imaging} of small metallic objects. It is based on a cold $^{85}$Rb atomic sample, and operates in an unshielded environment with no active field stabilization. It shows a sensitivity up to $200\, \rm{pT}/\sqrt{\rm Hz}$ in the $5-35\, \rm{kHz}$ range bandwidth and can resolve a cut $0.4\, \rm{mm}$ wide in a $0.8\, \rm{mm}$ thick metallic fo…
Progress in electromagnetic induction imaging with atomic magnetometers has brought its domain to the edge of the regime useful for biomedical imaging. However, a demonstration of imaging below the required 1 Sm$^{-1}$ level is still missing. In this Letter, we use an $^{87}$Rb radio-frequency atomic magnetometer operating near room temperature in an unshielded environment to image calibrated solutions mimicking the electric conductivity of live tissues. By combining the recently introduced near-resonant imaging technique with a dual radio-frequency coil excitation scheme, we image 5 mL of sol…
We describe our research programme on the use of atomic magnetometers to detect conductive objects via electromagnetic induction. The extreme sensitivity of atomic magnetometers at low frequencies, up to seven orders of magnitude higher than a coil-based system, permits deep penetration through different media and barriers, and in various operative environments. This eliminates the limitations usually associated with electromagnetic detection.
A recently introduced tuning-dressed scheme makes a Bell and Bloom magnetometer suited to detect weak variations of a radio-frequency (RF) magnetic field. We envisage the application of such innovative detection scheme as an alternative (or rather as a complement) to RF atomic magnetometers in electromagnetic-induction-imaging apparatuses.
A free-induction-decay (FID) type optically-pumped rubidium atomic magnetometer driven by a radio-frequency (RF) magnetic field is presented in this paper. Influences of parameters, such as the temperature of rubidium vapor cell, the power of pump beam, and the strength of RF magnetic field and static magnetic field on the amplitude and the full width at half maximum (FWHM) of the FID signal, have been investigated in the time domain and frequency domain. At the same time, the sensitivities of the magnetometer for the single-pass and the triple-pass probe beam cases have been compared by chang…
We report on the use of radio-frequency optical atomic magnetometers for magnetic induction tomography measurements. We demonstrate the imaging of dummy targets of varying conductivities placed in the proximity of the sensor, in an unshielded environment at room-temperature and without background subtraction. The images produced by the system accurately reproduce the characteristics of the actual objects. Furthermore, we perform finite element simulations in order to assess the potential for measuring low-conductivity biological tissues with our system. Our results demonstrate the feasibility …
We describe a system for the compensation of time-dependent stray magnetic fields using a dual channel scalar magnetometer based on non-linear Faraday rotation in synchronously optically pumped Cs vapour. We detail the active control strategy, with an emphasis on the electronic circuitry, based on a simple phase-locked-loop integrated circuit. The performance and limits of the system developed are tested and discussed. The system was applied to significantly improve the detection of free induction decay signals from protons of remotely magnetized water precessing in an ultra-low magnetic field…
We report on a two-channel magnetometer based on nonlinear magneto-optical rotation in a Cs glass cell with buffer gas. The Cs atoms are optically pumped and probed by free running diode lasers tuned to the D$_2$ line. A wide frequency modulation of the pump laser is used to produce both synchronous Zeeman optical pumping and hyperfine repumping. The magnetometer works in an unshielded environment and spurious signal from distant magnetic sources is rejected by means of differential measurement. In this regime the magnetometer simultaneously gives the magnetic field modulus and the field diffe…
We demonstrate detection of NMR signals using a non-cryogenic atomic magnetometer and describe several novel applications of this technique. A water free induction decay (FID) signal in a 0.5 $μ$T field is detected using a spin-exchange-relaxation-free K magnetometer and the possibility of using a multi-channel magnetometer for 3-D MRI requiring only a single FID signal is described. We also demonstrate detection of less than $10^{13}$ $^{129}$Xe atoms whose NMR signal is enhanced by a factor of 540 due to Fermi-contact interaction with K atoms. This technique allows detection of less than $10…
We demonstrate identification of position, material, orientation and shape of objects imaged by an $^{85}$Rb atomic magnetometer performing electromagnetic induction imaging supported by machine learning. Machine learning maximizes the information extracted from the images created by the magnetometer, demonstrating the use of hidden data. Localization 2.6 times better than the spatial resolution of the imaging system and successful classification up to 97$\%$ are obtained. This circumvents the need of solving the inverse problem, and demonstrates the extension of machine learning to diffusive …