非线性偶极反演(NDI)可实现定量磁化率映射(QSM),无需参数调整

Nonlinear Dipole Inversion (NDI) enables Quantitative Susceptibility Mapping (QSM) without parameter tuning

Heidelberg University(海德堡大学);Massachusetts General Hospital(麻省总医院);Siemens Healthcare GmbH(西门子医疗有限公司);National Nanotechnology Center(国家纳米技术中心);Seoul National University(首尔大学);Massachusetts Institute of Technology(麻省理工学院);German Cancer Research Center (DKFZ)(德国癌症研究中心);Harvard Medical School(哈佛医学院);Harvard-MIT Health Sciences and Technology(哈佛-麻省理工健康科学与技术学院)

Daniel Polak, Itthi Chatnuntawech, Jaeyeon Yoon, Siddharth Srinivasan Iyer, Jongho Lee, Peter Bachert, Elfar Adalsteinsson, Kawin Setsompop, Berkin Bilgic

摘要:

我们提出了非线性偶极子反演(NDI),用于高质量的定量磁化率映射(QSM),无需正则化调整,同时匹配最先进的重建技术的图像质量。除了避免这些技术经常遭受的过度平滑之外,我们还考虑了参数选择的需要。NDI足够灵活,可以从任意数量的头部方向进行重建,即使使用少至1个方向的数据,也优于COSMOS。这是通过一个非线性的前向模型,使用的幅度作为一个有效的先验,我们推导出一个简单的梯度下降更新规则。我们将此物理模型与变分网络(VN)协同结合,以利用VaNDI算法中的深度学习功能。该技术采用NDI中简单的梯度下降规则,并在训练过程中学习网络参数,因此不需要额外的参数调整。此外,我们使用高度加速的Wave-CAIPI采集在0.5 mm各向同性分辨率下评估了7 T下的NDI,并从少至2个方向的数据中展示了高质量的QSM。

英文摘要:

We propose Nonlinear Dipole Inversion (NDI) for high-quality Quantitative Susceptibility Mapping (QSM) without regularization tuning, while matching the image quality of state-of-the-art reconstruction techniques. In addition to avoiding over-smoothing that these techniques often suffer from, we also obviate the need for parameter selection. NDI is flexible enough to allow for reconstruction from an arbitrary number of head orientations, and outperforms COSMOS even when using as few as 1-direction data. This is made possible by a nonlinear forward-model that uses the magnitude as an effective prior, for which we derived a simple gradient descent update rule. We synergistically combine this physics-model with a Variational Network (VN) to leverage the power of deep learning in the VaNDI algorithm. This technique adopts the simple gradient descent rule from NDI and learns the network parameters during training, hence requires no additional parameter tuning. Further, we evaluate NDI at 7T using highly accelerated Wave-CAIPI acquisitions at 0.5 mm isotropic resolution and demonstrate high-quality QSM from as few as 2-direction data.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑