Obtaining accurate and reliable images from low-dose computed tomography (CT) is challenging. Regression convolutional neural network (CNN) models that are learned from training data are increasingly gaining attention in low-dose CT reconstruction. This paper modifies the architecture of an iterative regression CNN, BCD-Net, for fast, stable, and accurate low-dose CT reconstruction, and presents the convergence property of the modified BCD-Net. Numerical results with phantom data show that applying faster numerical solvers to model-based image reconstruction (MBIR) modules of BCD-Net leads to faster and more accurate BCD-Net; BCD-Net significantly improves the reconstruction accuracy, compared to the state-of-the-art MBIR method using learned transforms; BCD-Net achieves better image quality, compared to a state-of-the-art iterative NN architecture, ADMM-Net. Numerical results with clinical data show that BCD-Net generalizes significantly better than a state-of-the-art deep (non-iterative) regression NN, FBPConvNet, that lacks MBIR modules.
University of Michigan(密歇根大学);University of Hawai’i–Mānoa(夏威夷大学马诺阿分校)
Image reconstruction in low-count PET is particularly challenging because gammas from natural radioactivity in Lu-based crystals cause high random fractions that lower the measurement signal-to-noise-ratio (SNR). In model-based image reconstruction (MBIR), using more iterations of an unregularized method may increase the noise, so incorporating regularization into the image reconstruction is desirable to control the noise. New regularization methods based on learned convolutional operators are emerging in MBIR. We modify the architecture of an iterative neural network, BCD-Net, for PET MBIR, a…
With substantial public concerns on potential cancer risks and health hazards caused by the accumulated radiation exposure in medical imaging, reducing radiation dose in X-ray based medical imaging such as Computed Tomography Perfusion (CTP) has raised significant research interests. In this paper, we embrace the deep Convolutional Neural Networks (CNN) based approaches and introduce Smoothed Dense-Convolution Neural Network (SDCNet) to recover high-dose quality CTP images from low-dose ones. SDCNet is composed of sub-network blocks cascaded by skip-connections to infer the noise (differential…
University of São Paulo(圣保罗大学);Rensselaer Polytechnic Institute(伦斯勒理工学院);Fudan University(复旦大学);Shanghai Center for Brain Science and Brain-inspired Technology(上海脑科学与类脑研究中心)
Digital breast tomosynthesis (DBT) exams should utilize the lowest possible radiation dose while maintaining sufficiently good image quality for accurate medical diagnosis. In this work, we propose a convolution neural network (CNN) to restore low-dose (LD) DBT projections to achieve an image quality equivalent to a standard full-dose (FD) acquisition. The proposed network architecture benefits from priors in terms of layers that were inspired by traditional model-based (MB) restoration methods, considering a model-based deep learning approach, where the network is trained to operate in the va…
Northeastern University(东北大学);Shenzhen Institutes of Advanced Technology, Chinese Academy of Sciences(中国科学院深圳先进技术研究院);Rensselaer Polytechnic Institute(伦斯勒理工学院)
Due to the widespread use of positron emission tomography (PET) in clinical practice, the potential risk of PET-associated radiation dose to patients needs to be minimized. However, with the reduction in the radiation dose, the resultant images may suffer from noise and artifacts that compromise diagnostic performance. In this paper, we propose a parameter-transferred Wasserstein generative adversarial network (PT-WGAN) for low-dose PET image denoising. The contributions of this paper are twofold: i) a PT-WGAN framework is designed to denoise low-dose PET images without compromising structural…
Korea Advanced Institute of Science and Technology(韩国科学技术院)
Recently, deep learning approaches have been extensively studied for low-dose CT denoising thanks to its superior performance despite the fast computational time. In particular, cycleGAN has been demonstrated as a powerful unsupervised learning scheme to improve the low-dose CT image quality without requiring matched high-dose reference data. Unfortunately, one of the main limitations of the cycleGAN approach is that it requires two deep neural network generators at the training phase, although only one of them is used at the inference phase. The secondary auxiliary generator is needed to enfo…
Gating is commonly used in PET imaging to reduce respiratory motion blurring and facilitate more sophisticated motion correction methods. In the applications of low dose PET, however, reducing injection dose causes increased noise and reduces signal-to-noise ratio (SNR), subsequently corrupting the motion estimation/correction steps, causing inferior image quality. To tackle these issues, we first propose a Siamese adversarial network (SAN) that can efficiently recover high dose gated image volume from low dose gated image volume. To ensure the appearance consistency between the recovered gate…
Friedrich-Alexander Universität Erlangen-Nürnberg (FAU)(埃尔朗根-纽伦堡弗里德里希-亚历山大大学);Siemens Healthcare GmbH(西门子医疗有限公司)
Deep neural networks have shown great success in low dose CT denoising. However, most of these deep neural networks have several hundred thousand trainable parameters. This, combined with the inherent non-linearity of the neural network, makes the deep neural network diffcult to understand with low accountability. In this study we introduce JBFnet, a neural network for low dose CT denoising. The architecture of JBFnet implements iterative bilateral filtering. The filter functions of the Joint Bilateral Filter (JBF) are learned via shallow convolutional networks. The guidance image is estimated…
Ryerson University(瑞尔森大学);University of Saskatoon Health Region(萨斯卡通健康区域大学)
Low-dose CT denoising is a challenging task that has been studied by many researchers. Some studies have used deep neural networks to improve the quality of low-dose CT images and achieved fruitful results. In this paper, we propose a deep neural network that uses dilated convolutions with different dilation rates instead of standard convolution helping to capture more contextual information in fewer layers. Also, we have employed residual learning by creating shortcut connections to transmit image information from the early layers to later ones. To further improve the performance of the netwo…
Indian Institute of Technology Kharagpur(印度理工学院卡拉格布尔分校)
The resurgence of deep neural networks has created an alternative pathway for low-dose computed tomography denoising by learning a nonlinear transformation function between low-dose CT (LDCT) and normal-dose CT (NDCT) image pairs. However, those paired LDCT and NDCT images are rarely available in the clinical environment, making deep neural network deployment infeasible. This study proposes a novel method for self-supervised low-dose CT denoising to alleviate the requirement of paired LDCT and NDCT images. Specifically, we have trained an invertible neural network to minimize the pixel-based m…
Computer-Aided-Diagnosis (CADx) systems assist radiologists with identifying and classifying potentially malignant pulmonary nodules on chest CT scans using morphology and texture-based (radiomic) features. However, radiomic features are sensitive to differences in acquisitions due to variations in dose levels and slice thickness. This study investigates the feasibility of generating a normalized scan from heterogeneous CT scans as input. We obtained projection data from 40 low-dose chest CT scans, simulating acquisitions at 10%, 25% and 50% dose and reconstructing the scans at 1.0mm and 2.0mm…
Low dose computed tomography is a mainstream for clinical applications. How-ever, compared to normal dose CT, in the low dose CT (LDCT) images, there are stronger noise and more artifacts which are obstacles for practical applications. In the last few years, convolution-based end-to-end deep learning methods have been widely used for LDCT image denoising. Recently, transformer has shown superior performance over convolution with more feature interactions. Yet its ap-plications in LDCT denoising have not been fully cultivated. Here, we propose a convolution-free T2T vision transformer-based Enc…
Technical University of Munich(慕尼黑工业大学);Istanbul Technical University(伊斯坦布尔理工大学)
In this paper, we introduced a novel deep learning-based reconstruction technique for low-dose CT imaging using 3 dimensional convolutions to include the sagittal information unlike the existing 2 dimensional networks which exploits correlation only in transverse plane. In the proposed reconstruction technique, sparse and noisy sinograms are back-projected to the image domain with FBP operation, then the denoising process is applied with a U-Net like 3-dimensional network called 3D U-NetR. The proposed network is trained with synthetic and real chest CT images, and 2D U-Net is also trained wit…
The explosive rise of the use of Computer tomography (CT) imaging in medical practice has heightened public concern over the patient's associated radiation dose. However, reducing the radiation dose leads to increased noise and artifacts, which adversely degrades the scan's interpretability. Consequently, an advanced image reconstruction algorithm to improve the diagnostic performance of low dose ct arose as the primary concern among the researchers, which is challenging due to the ill-posedness of the problem. In recent times, the deep learning-based technique has emerged as a dominant method…
Harbin Institute of Technology(哈尔滨工业大学);The Second Affiliated Hospital of Harbin Medical University(哈尔滨医科大学附属第二医院)
Low-dose computed tomography (CT) allows the reduction of radiation risk in clinical applications at the expense of image quality, which deteriorates the diagnosis accuracy of radiologists. In this work, we present a High-Quality Imaging network (HQINet) for the CT image reconstruction from Low-dose computed tomography (CT) acquisitions. HQINet was a convolutional encoder-decoder architecture, where the encoder was used to extract spatial and temporal information from three contiguous slices while the decoder was used to recover the spacial information of the middle slice. We provide experimen…
Low-dose CT has been a key diagnostic imaging modality to reduce the potential risk of radiation overdose to patient health. Despite recent advances, CNN-based approaches typically apply filters in a spatially invariant way and adopt similar pixel-level losses, which treat all regions of the CT image equally and can be inefficient when fine-grained structures coexist with non-uniformly distributed noises. To address this issue, we propose a Structure-preserving Kernel Prediction Network (StructKPN) that combines the kernel prediction network with a structure-aware loss function that utilizes t…
University of Michigan - Shanghai Jiao Tong University Joint Institute(密歇根大学-上海交通大学联合学院);Shanghai Jiao Tong University(上海交通大学);University of Hawai’i at Manoa(夏威夷大学马诺阿分校)
This paper applies the recent fast iterative neural network framework, Momentum-Net, using appropriate models to low-dose X-ray computed tomography (LDCT) image reconstruction. At each layer of the proposed Momentum-Net, the model-based image reconstruction module solves the majorized penalized weighted least-square problem, and the image refining module uses a four-layer convolutional neural network (CNN). Experimental results with the NIH AAPM-Mayo Clinic Low Dose CT Grand Challenge dataset show that the proposed Momentum-Net architecture significantly improves image reconstruction accuracy,…
Two-Phase Multi-Dose-Level PET Image Reconstruction with Dose Level Awareness
School of Computer Science, Sichuan University(四川大学计算机学院);School of Biomedical Engineering & State Key Laboratory of Advanced Medical Materials and Devices, ShanghaiTech University(上海科技大学生物医学工程学院 & 先进医学材料与器件国家重点实验室);Shanghai United Imaging Intelligence Co., Ltd.(上海联影智能医疗科技有限公司)
Impact of loss functions on the performance of a deep neural network designed to restore low-dose digital mammography
Rensselaer Polytechnic Institute(伦斯勒理工学院);Fudan University(复旦大学);Shanghai Center for Brain Science and Brain-inspired Technology(上海脑科学与类脑研究中心);São Carlos School of Engineering, University of São Paulo(圣保罗大学圣卡洛斯工程学院);Ribeirão Preto School of Medicine, University of São Paulo(圣保罗大学里贝朗普雷图医学院)
ASCON: Anatomy-aware Supervised Contrastive Learning Framework for Low-dose CT Denoising
Institute of Science and Technology for Brain-Inspired Intelligence, Fudan University(复旦大学类脑智能科学与技术研究院);School of Cyber Science and Engineering, Sichuan University(四川大学网络空间安全学院)
University of California Davis Medical Center(加州大学戴维斯分校医学中心)
Deep neural networks have a great potential to improve image denoising in low-dose computed tomography (LDCT). Popular ways to increase the network capacity include adding more layers or repeating a modularized clone model in a sequence. In such sequential architectures, the noisy input image and end output image are commonly used only once in the training model, which however limits the overall learning performance. In this paper, we propose a parallel-clone neural network method that utilizes a modularized network model and exploits the benefit of parallel input, parallel-output loss, and cl…
In the past few decades, to reduce the risk of X-ray in computed tomography (CT), low-dose CT image denoising has attracted extensive attention from researchers, which has become an important research issue in the field of medical images. In recent years, with the rapid development of deep learning technology, many algorithms have emerged to apply convolutional neural networks to this task, achieving promising results. However, there are still some problems such as low denoising efficiency, over-smoothed result, etc. In this paper, we propose the Edge enhancement based Densely connected Convol…
We propose a dual-domain cascade of U-nets (i.e. a "W-net") operating in both the spatial frequency and image domains to enhance low-dose CT (LDCT) images without the need for proprietary x-ray projection data. The central slice theorem motivated the use of the spatial frequency domain in place of the raw sinogram. Data were obtained from the AAPM Low-dose Grand Challenge. A combination of Fourier space (F) and/or image domain (I) U-nets and W-nets were trained with a multi-scale structural similarity and mean absolute error loss function to denoise filtered back projected (FBP) LDCT images wh…
A deep dictionary network-based foundation model for ultra-low-dose CT denoising
Yanshan University(燕山大学);Capital Medical University(首都医科大学);Beijing Friendship Hospital(北京友谊医院);Shenzhen Institute of Advanced Technology, Chinese Academy of Sciences(中国科学院深圳先进技术研究院);Agency for Science, Technology and Research (A*STAR)(新加坡科技研究局)
The Plug-and-Play (PnP) framework was recently introduced for low-dose CT reconstruction to leverage the interpretability and the flexibility of model-based methods to incorporate various plugins, such as trained deep learning (DL) neural networks. However, the benefits of PnP vs. state-of-the-art DL methods have not been clearly demonstrated. In this work, we proposed an improved PnP framework to address the previous limitations and develop clinical-relevant segmentation metrics for quantitative result assessment. Compared with the DL alone methods, our proposed PnP framework was slightly inf…
Enhancing Low-dose CT Image Reconstruction by Integrating Supervised and Unsupervised Learning
University of Michigan Shanghai Jiao Tong University Joint Institute(密歇根大学上海交通大学联合学院);Shanghai Jiao Tong University(上海交通大学);Amazon Inc.(亚马逊公司);Michigan State University(密歇根州立大学)
Rensselaer Polytechnic Institute(伦斯勒理工学院);Massachusetts General Hospital(麻省总医院);Harvard Medical School(哈佛医学院);University of Massachusetts Lowell(马萨诸塞大学洛厄尔分校)
By the ALARA (As Low As Reasonably Achievable) principle, ultra-low-dose CT reconstruction is a holy grail to minimize cancer risks and genetic damages, especially for children. With the development of medical CT technologies, the iterative algorithms are widely used to reconstruct decent CT images from a low-dose scan. Recently, artificial intelligence (AI) techniques have shown a great promise in further reducing CT radiation dose to the next level. In this paper, we demonstrate that AI-powered CT reconstruction offers diagnostic image quality at an ultra-low-dose level comparable to that of…
Low Dose Computed Tomography suffers from a high amount of noise and/or undersampling artefacts in the reconstructed image. In the current article, a Deep Learning technique is exploited as a regularization term for the iterative reconstruction method SIRT. While SIRT minimizes the error in the sinogram space, the proposed regularization model additionally steers intermediate SIRT reconstructions towards the desired output. Extensive evaluations demonstrate the superior outcomes of the proposed method compared to the state of the art techniques. Comparing the forward projection of the reconstr…
University of Texas Southwestern Medical Centre(德克萨斯大学西南医学中心)
Despite the indispensable role of X-ray computed tomography (CT) in diagnostic medicine field, the associated ionizing radiation is still a major concern considering that it may cause genetic and cancerous diseases. Decreasing the exposure can reduce the dose and hence the radiation-related risk, but will also induce higher quantum noise. Supervised deep learning can be used to train a neural network to denoise the low-dose CT (LDCT). However, its success requires massive pixel-wise paired LDCT and normal-dose CT (NDCT) images, which are rarely available in real practice. To alleviate this pro…
Cycle-Constrained Adversarial Denoising Convolutional Network for PET Image Denoising: Multi-Dimensional Validation on Large Datasets with Reader Study and Real Low-Dose Data
Fudan University(复旦大学);Sichuan University(四川大学);Shanghai Center for Brain Science and Brain-inspired Technology(上海脑科学与类脑研究中心)
LDCT has drawn major attention in the medical imaging field due to the potential health risks of CT-associated X-ray radiation to patients. Reducing the radiation dose, however, decreases the quality of the reconstructed images, which consequently compromises the diagnostic performance. Various deep learning techniques have been introduced to improve the image quality of LDCT images through denoising. GANs-based denoising methods usually leverage an additional classification network, i.e. discriminator, to learn the most discriminate difference between the denoised and normal-dose images and, …
Concordia University(康考迪亚大学);University of Toronto(多伦多大学);McGill University Health Center-Research Institute(麦吉尔大学健康中心-研究所);Sunnybrook Health Sciences Centre(新宁健康科学中心);University of Montreal(蒙特利尔大学);National Cancer Institute(国家癌症研究所)
Reverse transcription-polymerase chain reaction (RT-PCR) is currently the gold standard in COVID-19 diagnosis. It can, however, take days to provide the diagnosis, and false negative rate is relatively high. Imaging, in particular chest computed tomography (CT), can assist with diagnosis and assessment of this disease. Nevertheless, it is shown that standard dose CT scan gives significant radiation burden to patients, especially those in need of multiple scans. In this study, we consider low-dose and ultra-low-dose (LDCT and ULDCT) scan protocols that reduce the radiation exposure close to tha…
The extensive use of medical CT has raised a public concern over the radiation dose to the patient. Reducing the radiation dose leads to increased CT image noise and artifacts, which can adversely affect not only the radiologists judgement but also the performance of downstream medical image analysis tasks. Various low-dose CT denoising methods, especially the recent deep learning based approaches, have produced impressive results. However, the existing denoising methods are all downstream-task-agnostic and neglect the diverse needs of the downstream applications. In this paper, we introduce a…
Low-dose CT (LDCT) imaging is desirable in many clinical applications to reduce X-ray radiation dose to patients. Inspired by deep learning (DL), a recent promising direction of model-based iterative reconstruction (MBIR) methods for LDCT is via optimization-unrolling DL-regularized image reconstruction, where pre-defined image prior is replaced by learnable data-adaptive prior. However, LDCT is clinically multilevel, since clinical scans have different noise levels that depend of scanning site, patient size, and clinical task. Therefore, this work aims to develop an adaptive-hyper-parameter D…
End-to-end Triple-domain PET Enhancement: A Hybrid Denoising-and-reconstruction Framework for Reconstructing Standard-dose PET Images from Low-dose PET Sinograms
ShanghaiTech University(上海科技大学);Shanghai United Imaging Intelligence Co., Ltd.(上海联影智能医疗科技有限公司);Shanghai Clinical Research and Trial Center(上海临床研究与试验中心);Shanghai Artificial Intelligence Laboratory(上海人工智能实验室)
提出端到端三域 TriPLET 框架,融合投影域去噪、小波域重建和图像域对抗评估,从低剂量 PET 正弦图恢复标准剂量图像,并取得最优相似度与信噪比。
University of Texas Southwestern Medical Centre(德克萨斯大学西南医学中心)
Low Dose Computed Tomography (LDCT) is clinically desirable due to the reduced radiation to patients. However, the quality of LDCT images is often sub-optimal because of the inevitable strong quantum noise. Inspired by their unprecedent success in computer vision, deep learning (DL)-based techniques have been used for LDCT denoising. Despite the promising noise removal ability of DL models, people have observed that the resolution of the DL-denoised images is compromised, decreasing their clinical value. Aiming at relieving this problem, in this work, we developed a more effective denoiser by …
Restoring high-quality CT images from low dose CT counterparts is an ill-posed, nonlinear problem to which Deep Learning approaches have been giving superior solutions compared to classical model-based approaches. In this article, a framework is presented wherein a Recurrent Neural Network (RNN) is utilized to remove the streaking artefacts from low projection number CT imaging. The results indicate similar image restoration performance for the RNN compared to the feedforward network in low noise cases while in high noise levels the RNN returns better results. The computational costs are also …
Low-dose computed tomography (LDCT) scans, which can effectively alleviate the radiation problem, will degrade the imaging quality. In this paper, we propose a novel LDCT reconstruction network that unrolls the iterative scheme and performs in both image and manifold spaces. Because patch manifolds of medical images have low-dimensional structures, we can build graphs from the manifolds. Then, we simultaneously leverage the spatial convolution to extract the local pixel-level features from the images and incorporate the graph convolution to analyze the nonlocal topological features in manifold…
Shanghai Jiao Tong University(上海交通大学);National University of Singapore(新加坡国立大学);South China University of Technology(华南理工大学)
Low-dose CT (LDCT) imaging attracted a considerable interest for the reduction of the object's exposure to X-ray radiation. In recent years, supervised deep learning (DL) has been extensively studied for LDCT image reconstruction, which trains a network over a dataset containing many pairs of normal-dose and low-dose images. However, the challenge on collecting many such pairs in the clinical setup limits the application of such supervised-learning-based methods for LDCT image reconstruction in practice. Aiming at addressing the challenges raised by the collection of training dataset, this pap…
Low-electron-dose observation is indispensable for observing various samples using a transmission electron microscope; consequently, image processing has been used to improve transmission electron microscopy (TEM) images. To apply such image processing to in situ observations, we here apply a convolutional neural network to TEM imaging. Using a dataset that includes short-exposure images and long-exposure images, we develop a pipeline for processed short-exposure images, based on end-to-end training. The quality of images acquired with a total dose of approximately 5 e- per pixel becomes compa…
Maastricht University Medical Centre+(马斯特里赫特大学医学中心+)
Radiomics is an active area of research focusing on high throughput feature extraction from medical images with a wide array of applications in clinical practice, such as clinical decision support in oncology. However, noise in low dose computed tomography (CT) scans can impair the accurate extraction of radiomic features. In this article, we investigate the possibility of using deep learning generative models to improve the performance of radiomics from low dose CTs. We used two datasets of low dose CT scans -NSCLC Radiogenomics and LIDC-IDRI - as test datasets for two tasks - pre-treatment s…
A Low-dose CT Reconstruction Network Based on TV-regularized OSEM Algorithm
School of Mathematical Sciences, Capital Normal University(首都师范大学数学科学学院);Centre for Mathematical Imaging Techniques, University of Liverpool(利物浦大学数学成像技术中心);Academy of Mathematics and Systems Science, Chinese Academy of Sciences(中国科学院数学与系统科学研究院);Department of Mathematics and Statistics, University of Strathclyde(思克莱德大学数学与统计系)
Sichuan University(四川大学);Fudan University(复旦大学);Shanghai Center for Brain Science and Brain-Inspired Technology(上海脑科学与类脑研究中心);Rensselaer Polytechnic Institute(伦斯勒理工学院)
Since 2016, deep learning (DL) has advanced tomographic imaging with remarkable successes, especially in low-dose computed tomography (LDCT) imaging. Despite being driven by big data, the LDCT denoising and pure end-to-end reconstruction networks often suffer from the black box nature and major issues such as instabilities, which is a major barrier to apply deep learning methods in low-dose CT applications. An emerging trend is to integrate imaging physics and model into deep networks, enabling a hybridization of physics/model-based and data-driven elements. %This type of hybrid methods has be…
Purpose: CBCT-based adaptive radiotherapy requires daily images for accurate dose calculations. This study investigates the feasibility of applying a single convolutional network to facilitate CBCT-to-CT synthesis for head-and-neck, lung, and breast cancer patients. Methods: Ninety-nine patients diagnosed with head-and-neck, lung or breast cancer undergoing radiotherapy with CBCT-based position verification were included in this study. CBCTs were registered to planning CTs according to clinical procedures. Three cycle-consistent generative adversarial networks (cycle-GANs) were trained in an u…
University of Electronic Science and Technology of China(电子科技大学);West China Hospital, Sichuan University(四川大学华西医院);West China No.4 Hospital, Sichuan University(四川大学华西第四医院)
Being low-level radiation exposure and less harmful to health, low-dose computed tomography (LDCT) has been widely adopted in the early screening of lung cancer and COVID-19. LDCT images inevitably suffer from the degradation problem caused by complex noises. It was reported that deep learning (DL)-based LDCT denoising methods using convolutional neural network (CNN) achieved impressive denoising performance. Although most existing DL-based methods (e.g., encoder-decoder framework) can implicitly utilize non-local and contextual information via downsampling operator and 3D CNN, the explicit mu…
University of Florida(佛罗里达大学);Sichuan University(四川大学);Georgia State University(佐治亚州立大学)
We propose a provably convergent method, called Efficient Learned Descent Algorithm (ELDA), for low-dose CT (LDCT) reconstruction. ELDA is a highly interpretable neural network architecture with learned parameters and meanwhile retains convergence guarantee as classical optimization algorithms. To improve reconstruction quality, the proposed ELDA also employs a new non-local feature mapping and an associated regularizer. We compare ELDA with several state-of-the-art deep image methods, such as RED-CNN and Learned Primal-Dual, on a set of LDCT reconstruction problems. Numerical experiments demo…
Low-dose computed tomography (CT) plays a significant role in reducing the radiation risk in clinical applications. However, lowering the radiation dose will significantly degrade the image quality. With the rapid development and wide application of deep learning, it has brought new directions for the development of low-dose CT imaging algorithms. Therefore, we propose a fully unsupervised one sample diffusion model (OSDM)in projection domain for low-dose CT reconstruction. To extract sufficient prior information from single sample, the Hankel matrix formulation is employed. Besides, the penal…
Low-dose computed tomography (CT) denoising algorithms aim to enable reduced patient dose in routine CT acquisitions while maintaining high image quality. Recently, deep learning~(DL)-based methods were introduced, outperforming conventional denoising algorithms on this task due to their high model capacity. However, for the transition of DL-based denoising to clinical practice, these data-driven approaches must generalize robustly beyond the seen training data. We, therefore, propose a hybrid denoising approach consisting of a set of trainable joint bilateral filters (JBFs) combined with a co…
Traditional model-based image reconstruction (MBIR) methods combine forward and noise models with simple object priors. Recent application of deep learning methods for image reconstruction provides a successful data-driven approach to addressing the challenges when reconstructing images with measurement undersampling or various types of noise. In this work, we propose a hybrid supervised-unsupervised learning framework for X-ray computed tomography (CT) image reconstruction. The proposed learning formulation leverages both sparsity or unsupervised learning-based priors and neural network recon…
Cone-beam breast computed tomography (CT) provides true 3D breast images with isotropic resolution and high-contrast information, detecting calcifications as small as a few hundred microns and revealing subtle tissue differences. However, breast is highly sensitive to x-ray radiation. It is critically important for healthcare to reduce radiation dose. Few-view cone-beam CT only uses a fraction of x-ray projection data acquired by standard cone-beam breast CT, enabling significant reduction of the radiation dose. However, insufficient sampling data would cause severe streak artifacts in CT imag…
Synchrotron-based X-ray computed tomography is widely used for investigating inner structures of specimens at high spatial resolutions. However, potential beam damage to samples often limits the X-ray exposure during tomography experiments. Proposed strategies for eliminating beam damage also decrease reconstruction quality. Here we present a deep learning-based method to enhance low-dose tomography reconstruction via a hybrid-dose acquisition strategy composed of extremely sparse-view normal-dose projections and full-view low-dose projections. Corresponding image pairs are extracted from low-…
Computed tomography (CT) has played a vital role in medical diagnosis, assessment, and therapy planning, etc. In clinical practice, concerns about the increase of X-ray radiation exposure attract more and more attention. To lower the X-ray radiation, low-dose CT is often used in certain scenarios, while it will induce the degradation of CT image quality. In this paper, we proposed a training method that trained denoising neural networks without any paired clean data. we trained the denoising neural network to map one noise LDCT image to its two adjacent LDCT images in a singe 3D thin-layer low…