计算机断层成像中个体化剂量测定的综合方法
Comprehensive procedure for personalized dosimetry in computed tomography
摘要:
这项工作的目的是开发可行的程序,用于验证个性化剂量测定的适用性,在计算机断层成像(CT),使用蒙特卡洛模拟。开发了移动设备和定制软件,用于在临床条件下快速、非侵入性地确定CT扫描仪的等效源模型。扫描了在五个代表性位置配备实时CT剂量探头的标准和拟人CT剂量体模。在扫描期间在五个位置测量累积剂量。ImpactMC是一种基于Monte Carlo的CT剂量软件程序,用于模拟扫描。从扫描参数、等效源模型和体模的材料分割CT图像中获得必要的输入。模拟体模中的扫描后3D剂量分布,并将体模内五个位置处计算的剂量值与测量的剂量值进行比较。通过General Electric Optima CT 660和Toshiba(Canon)Aqueolone获得初步结果。一般而言,测量和计算的剂量值在相对不确定性范围内,估计不超过10%。开发的程序,它允许后CT扫描剂量进行测量和计算,在五个点内拟人化幻影,被认为是可行的和快速的。该程序适用于临床条件下的任何扫描仪类型。结果表明,该程序非常适合验证个性化CT剂量的适用性扫描后蒙特卡罗计算的基础上。
英文摘要:
The purpose of this work is to develop viable procedures for verifying the applicability of personalized dosimetry in computed tomography (CT) using Monte Carlo-based simulations. Mobile equipment together with customized software was developed and used for rapid, non-invasive determination of equivalent source models of CT scanners under clinical conditions. Standard and anthropomorphic CT dose phantoms equipped with real-time CT dose probes at five representative positions were scanned. The accumulated dose was measured during the scan at the five positions. ImpactMC, a Monte Carlo-based CT dose software program, was used to simulate the scan. The necessary inputs were obtained from the scan parameters, from the equivalent source models and from the material-segmented CT images of the phantoms. Post-scan 3D dose distributions in the phantoms were simulated and the dose values calculated at the five positions inside the phantom were compared to measured dose values. Initial results were obtained by means of a General Electric Optima CT 660 and a Toshiba (Canon) Aquilion ONE. In general, the measured and calculated dose values were within relative uncertainties that had been estimated to be less than 10%. The procedures developed, which allow the post-CT scan dose to be measured and calculated at five points inside anthropomorphic phantoms, were found to be viable and rapid. The procedures are applicable to any scanner type under clinical conditions. Results show that the procedures are well suited for verifying the applicability of personalized CT dosimetry based on post-scan Monte Carlo calculations.
机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。