Improvements of the systematic uncertainty, frequency instability, and long-term reliability of the two caesium fountain primary frequency standards CSF1 and CSF2 at PTB (Physikalisch-Technische Bundesanstalt) are described. We have further investigated many of the systematic effects and made a number of modifications of the fountains. With an optically stabilized microwave oscillator, the quantum projection noise limited frequency instabilities are improved to $7.2 \times 10^{-14} (τ/1\,\mathrm{s})^{-1/2}$ for CSF1 and $2.5 \times 10^{-14} (τ/1\,\mathrm{s})^{-1/2}$ for CSF2 at high atom density. The systematic uncertainties of CSF1 and CSF2 are reduced to $2.74 \times 10^{-16}$ and $1.71 \times 10^{-16}$, respectively. Both fountain clocks regularly calibrate the scale unit of International Atomic Time (TAI) and the local realization of Coordinated Universal Time, UTC(PTB), and serve as references to measure the frequencies of local and remote optical frequency standards.
We evaluate the frequency error from distributed cavity phase in the caesium fountain clock PTB-CSF2 at the Physikalisch-Technische Bundesanstalt with a combination of frequency measurements and ab initio calculations. The associated uncertainty is 1.3E-16, with a frequency bias of 0.4E-16. The agreement between the measurements and calculations explains the previously observed frequency shifts at elevated microwave amplitude. We also evaluate the frequency bias and uncertainty due to the microwave lensing of the atomic wavepackets. We report a total PTB-CSF2 systematic uncertainty of 4.1E-16.
A caesium fountain clock is operated utilizing a microwave oscillator that derives its frequency stability from a stable laser by means of a fiber-laser femtosecond frequency comb. This oscillator is based on the technology developed for optical clocks and replaces the quartz based microwave oscillator commonly used in fountain clocks. As a result, a significant decrease of the frequency instability of the fountain clock is obtained, reaching 0.74E-14 at 100 s averaging time. We could demonstrate that for a significant range of detected atom numbers the instability is limited by quantum projec…
We have observed that the collisional frequency shift in primary caesium fountain clocks varies with the clock state population composition and, in particular, is zero for a given fraction of the |F = 4, mF = 0> atoms, depending on the initial cloud parameters. We present a theoretical model explaining our observations. The possibility of the collisional shift cancellation implies an improvement in the performance of caesium fountain standards and a simplification in their operation. Our results also have implications for test operation of fountains at multiple pi/2 pulse areas.
A cesium atomic fountain clock is under development at Huazhong University of Science and Technology (HUST) in China. In this paper, we describe the construction of the entire fountain clock system and report the preliminary results. A frequency stability of $2.5\times 10^{-13} τ^{-1/2}$ has been achieved by inter-comparison with a hydrogen maser, and the factors limiting the frequency stability are also discussed.
We report the evaluation of the second order Zeeman shift in the continuous atomic fountain clock FoCS-2. Because of the continuous operation and of its geometrical constraints, the methods used in pulsed fountains are not applicable. We use here time-resolved Zeeman spectroscopy to probe the magnetic field profile in the clock. The pulses of ac magnetic excitation allow us to spatially resolve the Zeeman frequency and to evaluate the Zeeman shift with a relative uncertainty smaller than 10E-16.
An ensemble of rubidium atomic fountain clocks has been put into operation at the U.S. Naval Observatory (USNO). These fountains are used as continuous clocks in the manner of commercial cesium beams and hydrogen masers for the purpose of improved timing applications. Four fountains have been in operation for more than two years and are included in the ensemble used to generate the USNO master clock. Individual fountain performance is characterized by a white-frequency noise level below $2\times 10^{-13}$ and fractional-frequency stability routinely reaching the low $10^{-16}$s. The highest pe…
We performed an absolute frequency measurement of the $^1S_{0}$-$^3P_{0}$ transition in $^{87}$Sr with a fractional uncertainty of $1.2 \times 10^{-15}$, which is less than one third that of our previous measurement. A caesium fountain atomic clock was used as a transfer oscillator to reduce the uncertainty of the link between a strontium optical lattice clock and the SI second. The absolute value of the transition frequency is 429 228 004 229 873.56(49) Hz.
We describe an optical frequency stabilization scheme of a microwave oscillator that is used for the interrogation of primary caesium fountain clocks. Because of its superior phase noise properties, this scheme, which is based on an ultrastable laser and a femtosecond laser frequency comb, overcomes the frequency instability limitations of fountain clocks given by the previously utilized quartz-oscillator-based frequency synthesis. The presented scheme combines the transfer of the short-term frequency instability of an optical cavity and the long-term frequency instability of a hydrogen maser …
We present experimental work for improved atom loading in the optical molasses of a caesium fountain clock, employing a low-velocity intense source of atoms (LVIS) [Lu et al., Phys. Rev. Lett. 77, 3331 (1996)], which we modified by adding a "dark" state pump laser. With this modification the atom source has a mean flux of $4 \times 10^{8}$ atoms/s at a mean atom velocity of $8.6$ m/s. Compared to fountain operation using background gas loading, we achieved a significant increase of the loaded and detected atom number by a factor of 40. Operating the fountain clock with a total number of detect…
The fundamental limits of atomic fountains as operational clocks are considered. Four rubidium fountains in operation at the U.S. Naval Observatory for over 5.5 years have demonstrated unprecedented long-term stability for continuously running clocks [1,2]. With only these rubidium fountains, a post-processed timescale can be created that demonstrates superior long-term performance to any individual clock by compensating for occasional frequency steps. By comparing to the world's primary standards we demonstrate instability of this rubidium fountain timescale reaching the mid $10^{-17}$'s and …
We describe a new method to determine the position of the atomic cloud during its interaction with the microwave field in the cavity of a fountain clock. The positional information is extracted from the spectrum of the F=3,mF=0 to F=4,mF=-1 hyperfine transition, which shows a position dependent asymmetry when the magnetic C-field is tilted by a few degrees with respect to the cavity axis. Analysis of this spectral asymmetry provides the horizontal center-of-mass position for the ensemble of atoms contributing to frequency measurements. With an uncertainty on the order of 0.1 mm, the obtained i…
With 87-Sr atoms confined in a one dimensional optical lattice, the frequency of the optical clock transition 5s^2 ^1S_0 - 5s5p ^3P_0 has been determined to be 429 228 004 229 872.9(5) Hz. The transition frequency was measured with the help of a fs-frequency comb against one of PTB's H-masers whose frequency was measured simultaneously by the PTB Cs fountain clock CSF1. The Sr optical frequency standard contributes with a fractional uncertainty of 1.5 10^-16 to the total uncertainty. The agreement of the measured transition frequency with previous measurements at other institutes supports the …
We present a trajectory dynamically tracing compensation method to smooth the spatial fluctuation of the static magnetic field (C-field) that provides a quantization axis in the fountain clock. The C-field coil current is point-to-point adjusted in accordance to the atoms experienced magnetic field along the flight trajectory. A homogeneous field with a 0.2 nT uncertainty is realized compared to 5 nT under the static magnetic field with constant current during the Ramsey interrogation. The corresponding uncertainty associated with the second-order Zeeman shift that we calculate is improved by …
Four rubidium fountains at the U.S. Naval Observatory (USNO) have been in operation for 4.5 years. Each fountain has demonstrated periods of stability marked by Total or Theo deviation below 10^-16. Occasional frequency changes, on order of 1.5 times per year per fountain, introduce deviations from white-frequency noise behavior. Averaged together, the four fountains form an ensemble with a white-frequency noise level of 10^-13 and excellent long-term stability as compared to the primary frequency standards contributing to TAI. Progress on using the clocks at USNO for improving limits on coupl…
We describe the current status of the Ca optical frequency standards with laser-cooled neutral atoms realized in two different laboratories for the purpose of developing a possible future optical atomic clock. Frequency measurements performed at the Physikalisch-Technische Bundesanstalt (PTB) and the National Institute of Standards and Technology (NIST) make the frequency of the clock transition of 40Ca one of the best known optical frequencies (relative uncertainty 1.2e-14) and the measurements of this frequency in both laboratories agree to well within their respective uncertainties. Prospec…
We report the operation of a dual Rb/Cs atomic fountain clock. 133Cs and 87Rb atoms are cooled, launched, and detected simultaneously in LNE-SYRTE's FO2 double fountain. The dual clock operation occurs with no degradation of either the stability or the accuracy. We describe the key features for achieving such a simultaneous operation. We also report on the results of the first Rb/Cs frequency measurement campaign performed with FO2 in this dual atom clock configuration, including a new determination of the absolute 87Rb hyperfine frequency.
We report on the development and phase noise performance of a 9.1926 GHz microwave frequency synthesizer to be used as the local oscillator for a Cs fountain clock. It is based on frequency multiplication and synthesis from an ultralow phase noise 5 MHz Oven Controlled Crystal Oscillator (OCXO) and 100 MHz Voltage Controlled Crystal Oscillator (VCXO).The key component of the frequency multiplication is a non-linear transmission-line (NLTL) used as a frequency comb generator. The phase noise of the synthesizer is improved by carefully optimizing the input power, the input and output impedances …
We demonstrate agreement between measurements and ab initio calculations of the frequency shifts caused by distributed cavity phase variations in the microwave cavity of a primary atomic fountain clock. Experimental verification of the finite element models of the cavities gives the first quantitative evaluation of this leading uncertainty and allows it to be reduced to delta nu / nu = 8.4\times10^-17. Applying these experimental techniques to clocks with improved microwave cavities will yield negligible distributed cavity phase uncertainties, less than \pm1\times10^-17.
In this article, we report on the work done with the LNE-SYRTE atomic clock ensemble during the last 10 years. We cover progress made in atomic fountains and in their application to timekeeping. We also cover the development of optical lattice clocks based on strontium and on mercury. We report on tests of fundamental physical laws made with these highly accurate atomic clocks. We also report on work relevant to a future possible redefinition of the SI second.
We give an overview of the work done with the Laboratoire National de Métrologie et d'Essais-Systèmes de Référence Temps-Espace (LNE-SYRTE) fountain ensemble during the last five years. After a description of the clock ensemble, comprising three fountains, FO1, FO2, and FOM, and the newest developments, we review recent studies of several systematic frequency shifts. This includes the distributed cavity phase shift, which we evaluate for the FO1 and FOM fountains, applying the techniques of our recent work on FO2. We also report calculations of the microwave lensing frequency shift for the thr…
We used Precise Point Positioning, a well-established GPS carrier-phase frequency transfer method to perform a direct remote comparison of two optical frequency standards based on single laser-cooled $^{171}$Yb$^+$ ions operated at NPL, UK and PTB, Germany. At both institutes an active hydrogen maser serves as a flywheel oscillator; it is connected to a GPS receiver as an external frequency reference and compared simultaneously to a realization of the unperturbed frequency of the ${{}^2S_{1/2}(F=0)-{}^2D_{3/2}(F=2)}$ electric quadrupole transition in ${}^{171}$Yb${}^+$ via an optical femtoseco…
We demonstrate the use of a fiber-based femtosecond laser locked onto an ultra-stable optical cavity to generate a low-noise microwave reference signal. Comparison with both a liquid Helium cryogenic sapphire oscillator (CSO) and a Ti:Sapphire-based optical frequency comb system exhibit a stability about $3\times10^{-15}$ between 1 s and 10 s. The microwave signal from the fiber system is used to perform Ramsey spectroscopy in a state-of-the-art Cesium fountain clock. The resulting clock system is compared to the CSO and exhibits a stability of $3.5\times10^{-14}τ^{-1/2}$. Our continuously ope…
This paper describes advances in microwave frequency standards using laser-cooled atoms at BNM-SYRTE. First, recent improvements of the $^{133}$Cs and $^{87}$Rb atomic fountains are described. Thanks to the routine use of a cryogenic sapphire oscillator as an ultra-stable local frequency reference, a fountain frequency instability of $1.6\times 10^{-14}τ^{-1/2}$ where $τ$ is the measurement time in seconds is measured. The second advance is a powerful method to control the frequency shift due to cold collisions. These two advances lead to a frequency stability of $2\times 10^{-16}$ at $50,000s…
We stabilise a microwave oscillator at 9.6 GHz to an optical clock laser at 344 THz by using a fibre-based femtosecond laser frequency comb as a transfer oscillator. With a second frequency comb we measure independently the instability of the microwave source with respect to another optical clock laser frequency at 456 THz. The total fractional frequency instability of this optic-to-microwave and microwave-to-optic conversion resulted in an Allan deviation sigma_y, of sigma_y=1.2E-14 at 1 s averaging time (band width 50 kHz). The residual phase noise density is -97 dBc/Hz at 10 Hz offset from …
We report on the first absolute transition frequency measurement at the 10^{-15} level with a single, laser-cooled 40Ca+ ion in a linear Paul trap. For this measurement, a frequency comb is referenced to the transportable Cs atomic fountain clock of LNE-SYRTE and is used to measure the S1/2-D5/2 electric-quadrupole transition frequency. After the correction of systematic shifts, the clock transition frequency f_Ca+ = 411 042 129 776 393.2 (1.0) Hz is obtained, which corresponds to a fractional uncertainty within a factor of three of the Cs standard. Future improvements are expected to lead to …
We used a Cs atomic fountain frequency standard to measure the Stark shift on the ground state hyperfine transiton frequency in cesium (9.2 GHz) due to the electric field generated by the blackbody radiation. The measures relative shift at 300 K is -1.43(11)e-14 and agrees with our theoretical evaluation -1.49(07)e-14. This value differs from the currently accepted one -1.69(04)e-14. The difference has a significant implication on the accuracy of frequency standards, in clocks comparison, and in a variety of high precision physics tests such as the time stability of fundamental constants.
We report tests of local position invariance based on measurements of the ratio of the ground state hyperfine frequencies of 133Cs and 87Rb in laser-cooled atomic fountain clocks. Measurements extending over 14 years set a stringent limit to a possible variation with time of this ratio: d ln(nu_Rb/nu_Cs)/dt=(-1.39 +/- 0.91)x 10-16 yr-1. This improves by a factor of 7.7 over our previous report (H. Marion et al., Phys. Rev. Lett. 90, 150801 (2003)). Our measurements also set the first limit to a fractional variation of the Rb/Cs ratio with gravitational potential at the level of c^2 d ln(nu_Rb/…
Over five years we have compared the hyperfine frequencies of 133Cs and 87Rb atoms in their electronic ground state using several laser cooled 133Cs and 87Rb atomic fountains with an accuracy of ~10^{-15}. These measurements set a stringent upper bound to a possible fractional time variation of the ratio between the two frequencies : (d/dt)ln(nu_Rb/nu_Cs)=(0.2 +/- 7.0)*10^{-16} yr^{-1} (1 sigma uncertainty). The same limit applies to a possible variation of the quantity (mu_Rb/mu_Cs)*alpha^{-0.44}, which involves the ratio of nuclear magnetic moments and the fine structure constant.
Coherent manipulation of atomic states is a key concept in high-precision spectroscopy and used in atomic fountain clocks and a number of optical frequency standards. Operation of these standards can involve a number of cyclic switching processes, which may induce cycle synchronous phase excursions of the interrogation signal and thus lead to shifts in the output of the frequency standard. We have built a FPGA-based phase analyzer to investigate these effects and conducted measurements on two frequency standards. For the caesium fountain PTB-CSF2 we were able to exclude phase variations of the…
We experimentally investigate an optical clock based on $^{171}$Yb ($I=1/2$) atoms confined in an optical lattice. We have evaluated all known frequency shifts to the clock transition, including a density-dependent collision shift, with a fractional uncertainty of $3.4 \times 10^{-16}$, limited principally by uncertainty in the blackbody radiation Stark shift. We measured the absolute clock transition frequency relative to the NIST-F1 Cs fountain clock and find the frequency to be 518 295 836 590 865.2(0.7) Hz.
The environmental perturbation on atoms is the key factor restricting the performance of atomic frequency standards, especially in long term scale. In this letter, we demonstrate a real-time noise distinguish operation of atomic clocks. The operation improves the statistical uncertainty by about an order of magnitude of our fountain clock which is deteriorated previously by extra noises. The frequency offset bring by the extra noise is also corrected. The experiment proves the real-time noise distinguish operation can reduce the contribution of ambient noises and improve the uncertainty limit …
The rapid increase in accuracy and stability of optical atomic clocks compared to the caesium atomic clock as primary standard of time and frequency asks for a future re-definition of the second in the International System of Units (SI). The status of the optical clocks based on either single ions in radio-frequency traps or on neutral atoms stored in an optical lattice is described with special emphasis of the current work at the Physikalisch-Technische Bundesanstalt (PTB). Besides the development and operation of different optical clocks with estimated fractional uncertainties in the 10^-18 …
An Al$^+$ clock with $1.6\times10^{-18}$ systematic uncertainty and its frequency ratios
Physikalisch-Technische Bundesanstalt(德国联邦物理技术研究院);Institut für Quantenoptik, Leibniz Universität Hannover(汉诺威莱布尼茨大学量子光学研究所);Laboratoire Kastler Brossel, Sorbonne Université, CNRS, ENS-PSL Research University, Collège de France(索邦大学、法国国家科学研究中心、巴黎文理研究大学、法兰西公学院卡斯特尔-布罗塞尔实验室)
The short-term stability of passive atomic frequency standards, especially in pulsed operation, is often limited by local oscillator noise via intermodulation effects. We present an experimental demonstration of the intermodulation effect on the frequency stability of a continuous atomic fountain clock where, under normal operating conditions, it is usually too small to observe. To achieve this, we deliberately degrade the phase stability of the microwave field interrogating the clock transition. We measure the frequency stability of the locked, commercial-grade local oscillator, for two modul…
We have measured the absolute frequency of the optical lattice clock based on $^{87}$Sr at PTB with an uncertainty of $3.9\times 10^{-16}$ using two caesium fountain clocks. This is close to the accuracy of today's best realizations of the SI second. The absolute frequency of the 5s$^2$ $^1$S$_0$-5s5p $^3$P$_0$ transition in $^{87}$Sr is 429,228,004,229,873.13(17) Hz. Our result is in excellent agreement with recent measurements performed in different laboratories worldwide. We improved the total systematic uncertainty of our Sr frequency standard by a factor of five and reach $3\times 10^{-17…
This paper discusses how to build a time scale with an intermittently-operated optical clock. In particular, it gives suggestions on how long and how often to run an optical clock. It also explores the benefits of having an optical clock in a time scale, by comparing with the current UTC(NIST) performance and the time scale with a continuously-operated Cs fountain.
We report on a series of 42 measurements of the transition frequency of the 429~THz (5s$^2$)~$^1$S$_0$--(5s5p)~$^3$P$_0$ line in $^{87}$Sr taken over three years from 2017 to 2019. They have been performed at the Physikalisch-Technische Bundesanstalt (PTB) between the laboratory strontium lattice clock and the primary caesium fountain clocks CSF1 and CSF2. The length of each individual measurement run has been extended by use of a hydrogen maser as flywheel to improve the statistical uncertainty given by the Cs clocks. We determine an averaged transition frequency of $429\:228\:004\:229\:873.0…
We report on the first comparison of distant caesium fountain primary frequency standards (PFSs) via an optical fiber link. The 1415 km long optical link connects two PFSs at LNE-SYRTE (Laboratoire National de métrologie et d'Essais - SYstème de Références Temps-Espace) in Paris (France) with two at PTB (Physikalisch-Technische Bundesanstalt) in Braunschweig (Germany). For a long time, these PFSs have been major contributors to accuracy of the International Atomic Time (TAI), with stated accuracies of around $3\times 10^{-16}$. They have also been the references for a number of absolute measur…
The absolute frequency of the 1S0-3P0 clock transition of 87Sr has been measured to be 429 228 004 229 873.65 (37) Hz using lattice-confined atoms, where the fractional uncertainty of 8.6x10-16 represents one of the most accurate measurements of an atomic transition frequency to date. After a detailed study of systematic effects, which reduced the total systematic uncertainty of the Sr lattice clock to 1.5x10-16, the clock frequency is measured against a hydrogen maser which is simultaneously calibrated to the US primary frequency standard, the NIST Cs fountain clock, NIST-F1. The comparison i…
We search for ultralight scalar dark matter candidates that induce oscillations of the fine structure constant, the electron and quark masses, and the quantum chromodynamics energy scale with frequency comparison data between an $^{171}$Yb optical lattice clock and a $^{133}$Cs fountain microwave clock that span 298 days with an uptime of 15.4 $\%$. New limits on the couplings of the scalar dark matter to electrons and gluons in the mass range from $10^{-22}$ eV/$c^{2}$ to $10^{-20}$ eV/$c^{2}$ are set, assuming that each of these couplings is the dominant source of the modulation in the frequ…
We report the direct excitation of the highly forbidden (6s^2) 1S0 - (6s6p) 3P0 optical transition in two odd isotopes of ytterbium. As the excitation laser frequency is scanned, absorption is detected by monitoring the depletion from an atomic cloud at ~70 uK in a magneto-optical trap. The measured frequency in 171Yb (F=1/2) is 518,295,836,593.2 +/- 4.4 kHz. The measured frequency in 173Yb (F=5/2) is 518,294,576,850.0 +/- 4.4 kHz. Measurements are made with a femtosecond-laser frequency comb calibrated by the NIST cesium fountain clock and represent nearly a million-fold reduction in uncertai…
The frequency ratio of the 1S0(F=1/2)-3P0(F=1/2) clock transition in 171Yb and the 1S0(F=9/2)-3P0(F=9/2) clock transition in 87Sr is measured by an optical-optical direct frequency link between two optical lattice clocks. We determined the ratio (ν_{Yb}/ν_{Sr}) to be 1.207 507 039 343 340 4(18) with a fractional uncertainty of 1.5x10^{-15}. The measurement uncertainty of the frequency ratio is smaller than that obtained from absolute frequency measurements using the International Atomic Time (TAI) link. The measured ratio agrees well with that derived from the absolute frequency measurement re…
We propose a method of joint interrogation in a single atom interferometer which overcomes the dead time between consecutive measurements in standard cold atomic fountains. The joint operation enables for a faster averaging of the Dick effect associated with the local oscillator noise in clocks and with vibration noise in cold atom inertial sensors. Such an operation allows achieving the lowest stability limit due to atom shot noise. We demonstrate a multiple joint operation in which up to five clouds of atoms are interrogated simultaneously in a single setup. The essential feature of multiple…
Optical lattice clocks surpass primary Cs microwave clocks in frequency stability and accuracy, and are promising candidates for a redefinition of the second in the International System of Units (SI). However, the robustness of optical lattice clocks has not yet reached a level comparable to that of Cs fountain clocks which contribute to International Atomic Time (TAI) by the nearly continuous operation. In this paper, we report the long-term operation of an $^{171}$Yb optical lattice clock with a coverage of 80.3% for half a year including uptimes of 93.9% for the first 24 days and 92.6% for …
A microwave atomic clock scheme based on Rb and Cs atoms trapped in optical lattice with magic wavelength for clock transition is proposed. The ac Stark shift of clock transition due to trapping laser can be canceled at some specific laser wavelengths. Comparing with in fountain clock, the cavity related shifts, the collision shift, and the Doppler effect are eliminated or suppressed dramatically in atomic clock when the magic optical lattice is exploited. By carefully analyzing various sources of clock uncertainty, we conclude that a microwave atomic clock with an accuracy of better than $2\t…
Progress in realizing the SI second had multiple technological impacts and enabled to further constraint theoretical models in fundamental physics. Caesium microwave fountains, realizing best the second according to its current definition with a relative uncertainty of 2-4x10^(-16), have already been superseded by atomic clocks referenced to an optical transition, both more stable and more accurate. Are we ready for a new definition of the second? Here we present an important step in this direction: our system of five clocks connects with an unprecedented consistency the optical and the microw…
We use an atomic fountain clock to measure quantum scattering phase shifts precisely through a series of narrow, low-field Feshbach resonances at average collision energies below $1\,μ$K. Our low spread in collision energy yields phase variations of order $\pm π/2$ for target atoms in several $F,m_F$ states. We compare them to a theoretical model and establish the accuracy of the measurements and the theoretical uncertainties from the fitted potential. We find overall excellent agreement, with small statistically significant differences that remain unexplained.
Absolute frequency measurement of the $^{40}$Ca$^{+}$ clock transition using a GNSS link to the SI second
Institut für Quantenoptik und Quanteninformation, Österreichische Akademie der Wissenschaften(奥地利科学院量子光学与量子信息研究所);Universität Innsbruck, Institut für Experimentalphysik(因斯布鲁克大学实验物理研究所);Alpine Quantum Technologies GmbH(阿尔卑斯量子技术有限公司);Singapore University of Technology and Design(新加坡科技设计大学);Qatar Center for Quantum Computing, College of Science and Engineering, Hamad Bin Khalifa University(哈马德·本·哈利法大学科学与工程学院卡塔尔量子计算中心)
Aided by ultra-high resolution spectroscopy, the overall systematic uncertainty of the $^{1}S_{0}$-$^{3}P_{0}$ clock resonance for lattice-confined $^{87}$Sr has been characterized to $9\times10^{-16}$. This uncertainty is at a level similar to the Cs-fountain primary standard, while the potential stability for the lattice clocks exceeds that of Cs. The absolute frequency of the clock transition has been measured to be 429,228,004,229,874.0(1.1) Hz, where the $2.5\times10^{-15}$ fractional uncertainty represents the most accurate measurement of a neutral-atom-based optical transition frequency…
Accurate measurements of different transition frequencies between atomic levels of the electronic and hyperfine structure over time are used to investigate temporal variations of the fine structure constant $α$ and the proton-to-electron mass ratio $μ$. We measure the frequency of the $^2S_{1/2}\rightarrow {^2F_{7/2}}$ electric octupole (E3) transition in $^{171}$Yb$^+$ against two caesium fountain clocks as $f(E3) = 642\,121\,496\,772\,645.36(25)$~Hz with an improved fractional uncertainty of $3.9\times 10^{-16}$. This transition frequency shows a strong sensitivity to changes of $α$. Togethe…
International Optical Clock Comparison Using the European Optical Fiber Network
Istituto Nazionale di Ricerca Metrologica (INRIM)(意大利国家计量研究所);Laboratoire Temps Espace (LNE-OP), Observatoire de Paris, Université PSL, Sorbonne Université, Université de Lille, LNE, CNRS(巴黎天文台时间空间实验室(LNE-OP));Laboratoire de Physique des Lasers (LPL), Université Sorbonne Paris Nord, CNRS(索邦巴黎北大学激光物理实验室(LPL));Physikalisch-Technische Bundesanstalt (PTB)(德国联邦物理技术研究院)
Optical clocks are not only powerful tools for prime fundamental research, but are also deemed for the re-definition of the SI base unit second as they now surpass the performance of caesium atomic clocks in both accuracy and stability by more than an order of magnitude. However, an important obstacle in this transition has so far been the limited reliability of the optical clocks that made a continuous realization of a timescale impractical. In this paper, we demonstrate how this situation can be resolved and that a timescale based on an optical clock can be established that is superior to on…