arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1806.01138physics.ins-detnucl-ex
收藏

DTAS检测器的表征和性能

Characterization and performance of the DTAS detector

V. Guadilla, J. L. Tain, A. Algora, J. Agramunt, J. Äystö, J. A. Briz, A. Cucoanes, T. Eronen, M. Estienne, M. Fallot, L. M. Fraile, E. Ganioğlu, W. Gelletly, D… 展开作者

V. Guadilla, J. L. Tain, A. Algora, J. Agramunt, J. Äystö, J. A. Briz, A. Cucoanes, T. Eronen, M. Estienne, M. Fallot, L. M. Fraile, E. Ganioğlu, W. Gelletly, D. Gorelov, J. Hakala, A. Jokinen, D. Jordan, A. Kankainen, V. Kolhinen, J. Koponen, M. Lebois, L. Le Meur, T. Martinez, M. Monserrate, A. Montaner-Pizá, I. Moore, E. Nácher, S. E. A. Orrigo, H. Penttilä, I. Pohjalainen, A. Porta, J. Reinikainen, M. Reponen, S. Rice, S. Rinta-Antila, B. Rubio, K. Rytkönen, T. Shiba, V. Sonnenschein, A. A. Sonzogni, E. Valencia, V. Vedia, A. Voss, J. N. Wilson, A. -A. Zakari-Issoufou

摘要:

DTAS是为FAIR的DESPEC实验开发的分段式全吸收γ射线谱仪。它由多达18个NaI(Tl)晶体组成。在这项工作中,我们研究了这种检测器的性能与实验室的来源,也在真实的实验条件下。我们提出了一个程序来离线重建的能量沉积在所有的晶体的光谱仪,这是复杂的NaI(Tl)光产额的非比例性的影响的总和。使用一个系统,以纠正在各个检测器模块的增益的时间变化,基于光脉冲发生器,证明。我们还描述了一个基于事件的方法来评估总结堆积电子失真分段光谱仪。所有这些都允许使用蒙特卡罗模拟对探测器进行仔细表征,这是计算用于分析总吸收γ射线光谱数据的响应函数所需的。特别注意中子与谱仪的相互作用,因为它们是\b{eta} -缓发中子发射核研究中的污染源。

英文摘要:

DTAS is a segmented total absorption γ-ray spectrometer developed for the DESPEC experiment at FAIR. It is composed of up to eighteen NaI(Tl) crystals. In this work we study the performance of this detector with laboratory sources and also under real experimental conditions. We present a procedure to reconstruct offline the sum of the energy deposited in all the crystals of the spectrometer, which is complicated by the effect of NaI(Tl) light-yield non-proportionality. The use of a system to correct for time variations of the gain in individual detector modules, based on a light pulse generator, is demonstrated. We describe also an event-based method to evaluate the summing-pileup electronic distortion in segmented spectrometers. All of this allows a careful characterization of the detector with Monte Carlo simulations that is needed to calculate the response function for the analysis of total absorption γ-ray spectroscopy data. Special attention was paid to the interaction of neutrons with the spectrometer, since they are a source of contamination in studies of \b{eta}-delayed neutron emitting nuclei.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑