arXivDaily arXiv每日学术速递 周一至周五更新
arXiv 1707.05780physics.ins-detnucl-ex
收藏

高通量环境下紧凑型中子谱仪的改进

Upgrade of compact neutron spectrometer for high flux environments

M. Osipenko, A. Bellucci, V. Ceriale, D. Corsini, G. Gariano, F. Gatti, M. Girolami, S. Minutoli, F. Panza, M. Pillon, M. Ripani, D. M. Trucchi

摘要:

本文介绍了一种新型的用于高通量环境的锂基中子谱仪。新的光谱仪是用电子级的商业单晶化学气相沉积金刚石建造的。这些晶体具有更好的电荷收集以及更高的辐射硬度。欧姆金属接触沉积在金刚石上,抑制在先前的原型中观察到的空间电荷的积聚。在探测器端实现了新的无源信号前置放大,以提高分辨率。该前置放大器基于对高中子通量不敏感的RF Transformer。紧凑的机械设计允许将检测器尺寸减小到直径为1 cm、长为13 cm的管。该谱仪在TRIGA堆热柱和DD中子发生器上进行了测试。测试结果表明,能量分辨率为72 keV(RMS)和符合时间分辨率为68 ps(RMS)。测量数据与Geant4模拟一致,除了较大的能量损失尾部,推测与金属接触和胶膨胀的缺陷。

英文摘要:

In this paper a new version of Li6-based neutron spectrometer for high flux environments is described. The new spectrometer was built with commercial single crystal Chemical Vapour Deposition diamonds of electronic grade. These crystals feature better charge collection as well as higher radiation hardness. Ohmic metal contacts were deposited on the diamonds suppressing build-up of space charge observed in the previous prototypes. New passive preamplification of signal at detector side was implemented to improve the resolution. This preamplification is based on RF transformer not sensitive to high neutron flux. Compact mechanical design allowed to reduce detector size to a tube of 1 cm diameter and 13 cm long. The spectrometer was tested in thermal column of TRIGA reactor and at DD neutron generator. The test results indicate an energy resolution of 72 keV (RMS) and coincidence timing resolution of 68 ps (RMS). The measured data are in agreement with Geant4 simulations except for larger energy loss tail presumably related to imperfections of metal contacts and glue expansion.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑