用于商用超导量子干涉仪磁强计的微型绝热退磁制冷机

Tiny adiabatic-demagnetization refrigerator for a commercial superconducting quantum interference device magnetometer

Taku J Sato, Daisuke Okuyama, Hideo Kimura

摘要:

为商用超导量子干涉装置磁强计[Quantum Design公司的磁性测量系统]研制了一种小型绝热退磁制冷机(T-ADR)。整个T-ADR系统安装在直径为$8.5$ mm、长度为$250$ mm的圆柱形空间内,并可插入MPMS的狭窄样品管中。T-ADR中的吸附泵是独立的,因此不需要复杂的气体处理系统。使用单晶Gd $_3$ Ga $_5$ O $_{12}$石榴石($\sim 2$克)作为磁制冷剂,常规可达到的最低温度为$\sim 0.56$ K。磁化异常的检测下限为$\sim 1 \times 10^{-7}$ emu,由测量磁化的波动估计。背景水平为$\sim 5 \times 10^{-5}$ emu,在$H = 100$ Oe下低于2 K,这主要归因于来自磁制冷剂的污染顺磁信号。

英文摘要:

A tiny adiabatic-demagnetization refrigerator (T-ADR) has been developed for a commercial superconducting quantum interference device magnetometer [Magnetic Property Measurement System (MPMS) from Quantum Design]. The whole T-ADR system is fit in a cylindrical space of the diameter $8.5$ mm and the length $250$ mm, and can be inserted into the narrow sample tube of MPMS. A sorption pump is self-contained in T-ADR, and hence no complex gas handling system is necessary. With the single crystalline Gd$_3$Ga$_5$O$_{12}$ garnet ($\sim 2$ grams) used as a magnetic refrigerant, the routinely achievable lowest temperature is $\sim 0.56$ K. The lower detection limit for a magnetization anomaly is $\sim 1 \times 10^{-7}$ emu, estimated from fluctuation of the measured magnetization. The background level is $\sim 5 \times 10^{-5}$ emu below 2 K at $H = 100$ Oe, which is largely attributable to a contaminating paramagnetic signal from the magnetic refrigerant.