用于Mu2e量热计的未掺杂CsI晶体和硅光电倍增器的辐射硬度测试

Radiation hardness test of un-doped CsI crystals and Silicon Photomultipliers for the Mu2e calorimeter

Stefania Baccaro, Alessia Cemmi, Marco Cordelli, Eleonora Diociaiuti, Raffaella Donghia, Simona Giovannella, Stefano Loreti, Stefano Miscetti, Mario Pillon, Ivano Sarra

摘要:

Mu 2 e量能器由1400个未掺杂的CsI晶体耦合到排列在两个环形盘中的大面积UV扩展的硅光电倍增器组成。这种热量计必须提供关于能量、时间和位置的精确信息。它还应该足够快,以处理高速率的背景,它必须在高辐射环境中运行和生存。模拟研究估计,在最热的区域,每个晶体将吸收300 Gy的剂量,并将在3年的运行中暴露于6 x 10^{11} n/cm ^2的中子通量。在意大利的CALLIOPE和FNG ENEA设施中,对未掺杂的CsI晶体进行了辐照试验,辐照剂量高达900 Gy,中子注量率高达9 x 10^{11} n/cm ^2。我们研究了辐照后这些晶体的光产额(LY)和纵向响应均匀性(LRU)的变化。电离剂量不修改LRU,而在900 Gy时观察到LY减少20%。类似地,中子通量引起可接受的LY劣化(<15%)。不同类型的SIPMs(两种不同的阵列模型从滨松和FBK之一)的中子辐照试验也进行了通过测量的泄漏电流和电荷响应的变化,以紫外线发光二极管。我们的结论是,在实验中,我们将需要冷却的SIPMS到0 C减少泄漏电流到一个可接受的水平。

英文摘要:

The Mu2e calorimeter is composed by 1400 un-doped CsI crystals coupled to large area UV extended Silicon Photomultipliers arranged in two annular disks. This calorimeter has to provide precise information on energy, timing and position. It should also be fast enough to handle the high rate background and it must operate and survive in a high radiation environment. Simulation studies estimated that, in the hottest regions, each crystal will absorb a dose of 300 Gy and will be exposed to a neutron fluency of 6 x 10^{11} n/cm^2 in 3 years of running. Test of un-doped CsI crystals irradiated up to 900 Gy and to a neutron fluency up to 9 x 10^{11} n/cm^2 have been performed at CALLIOPE and FNG ENEA facilities in Italy. We present our study on the variation of light yield (LY) and longitudinal response uniformity (LRU) of these crystals after irradiation. The ionization dose does not modify LRU while a 20% reduction in LY is observed at 900 Gy. Similarly, the neutron flux causes an acceptable LY deterioration (<15%). A neutron irradiation test on different types of SIPMs (two different array models from Hamamatsu and one from FBK) have also been carried out by measuring the variation of the leakage current and the charge response to an ultraviolet led. We concluded that, in the experiment, we will need to cool down the SIPMs to 0 C reduce the leakage current to an acceptable level.

机构、总结、中文摘要均由AI分析而来,可能存在错误,请以论文正文为准。

↑